Sphygmomanometers for Self-measurement of Blood Pressure (SBPM)
| UPPER ARM DEVICES | WRIST DEVICES | COMMUNITY DEVICES |
The following tables are lists of currently available Upper Arm and Wrist devices for Self-measurement of Blood Pressure and Devices for Measuring Blood Pressure in the Community. Discontinued devices are shown on a separate table. A complete list of all devices is available on our Device Index.
Upper Arm Devices for Self-measurement of Blood Pressure
Device |
Mode |
AAMI |
BHS |
ESH 2002 |
ESH 2010 |
ISO 81060-2:2013 |
ISO 81060-2:2018 |
Circumstance |
Recommendation Ref |
A&D UA-611Plus |
Osc |
|
A/A |
|
|
|
|
UA-1020 Equivalence |
Recommended E194 |
A&D UA-631 (UA-779 Life Source) |
Osc |
|
|
Pass |
|
|
|
At rest, Recruitment violations |
Recommended 4 |
A&D UA-651 |
Osc |
|
|
|
Pass |
|
|
Study (96) extended to ESH-IP 2010 requirements |
Recommended 97 |
|
|
|
Pass |
|
|
With a wide range cuff study extended to ESH-IP 2010 requirements |
Recommended 106 |
A&D UA-651Plus |
Osc |
|
A/A |
|
|
|
|
UA-1020 Equivalence |
Recommended E193 |
A&D UA-651SLPlus |
Osc |
|
A/A |
|
|
|
|
UA-1020 Equivalence |
Recommended E192 |
A&D UA-654MR |
Osc |
|
|
|
Pass |
|
|
UA-651 Equivalence |
Recommended E191 |
A&D UA-704 |
Osc |
|
A/A |
|
|
|
|
Study details omitted |
Questionable 8 |
A&D UA-705 |
Osc |
|
A/A |
|
|
|
|
At rest |
Recommended 18 |
A&D UA-767 |
Osc |
Pass |
A/A |
|
|
|
|
At rest; not high BP |
Recommended 3 |
A&D UA-767F |
Osc |
|
|
|
Pass |
|
|
UA-651 Equivalence |
Recommended E119 |
A&D UA-767S |
Osc |
|
|
|
Pass |
|
|
UA-651 Equivalence |
Recommended E120 |
A&D UA-774
(UA-767 Plus) |
Osc |
|
A/A |
|
|
|
|
At rest; tables incomplete |
Recommended 11 |
|
A/A |
|
|
|
|
At rest; Recruitment violations; Simultaneous readings |
Questionable 2 |
A&D UA-778 |
Osc |
Pass |
B/A |
|
|
|
|
Children Only, Ad hoc protocol adaptation. (No general validation) |
Recommended 53 |
A&D UA-787 |
Osc |
|
|
Pass |
|
|
|
|
Recommended 6 |
A&D UA-853 |
Osc |
|
|
Pass |
|
|
|
Minor recruitment modification |
Recommended 46 |
A&D UA-1020 |
Osc |
|
A/A |
|
|
|
|
D-ring 22cm to 32cm cuff |
Recommended 84 |
|
A/A |
|
|
|
|
Cylindrical 17cm to 32cm cuff |
Recommended 84 |
|
A/A |
|
|
|
|
D-ring 16cm to 24cm cuff |
Recommended 125 |
|
A/A |
|
|
|
|
D-ring 22cm to 32cm cuff |
Recommended 125 |
|
A/A |
|
|
|
|
D-ring 23cm to 37cm cuff |
Recommended 125 |
|
A/A |
|
|
|
|
D-ring 31cm to 45cm cuff |
Recommended 125 |
|
A/A |
|
|
|
|
Cylindrical 17cm to 32cm cuff |
Recommended 125 |
|
A/A |
|
|
|
|
Cylindrical 23cm to 27cm cuff |
Recommended 125 |
A&D UA-1200BLE |
Osc |
|
|
|
|
Pass |
|
At rest |
Recommended 126 |
Alvita Advanced (MA801f) |
Osc |
|
|
|
Pass |
|
|
Rossmax CF175 Equivalence |
Recommended E132 |
Alvita Arm (CG155f) |
Osc |
|
|
|
Pass |
|
|
Rossmax CF175 Equivalence |
Recommended E133 |
Alvita/Kinetik Wellbeing TMB-1970 |
Osc |
|
|
Pass |
|
|
|
Transtek TMB-986 Equivalence |
Recommended E303 |
Alvita/Kinetik Wellbeing TMB-1491-S |
Osc |
|
|
|
Pass |
|
|
Transtek TMB-1491 Equivalence |
Recommended E305 |
Alvita/Kinetik Wellbeing TMB-2083-N |
Osc |
|
|
|
Pass |
|
|
Transtek TMB-1491 Equivalence |
Recommended E306 |
Andon KD-391 |
Osc (Semi-Auto) |
|
|
Pass |
|
|
|
At rest |
Recommended 56 |
Andon KD-556 |
Osc |
|
|
Pass |
|
|
|
Andon KD-5915 Equivalence |
Recommended E42 |
Andon KD-575 |
Osc |
|
|
Fail |
|
|
|
At rest |
Not Recommended 59 |
Andon KD-595 |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 118 |
Andon KD-5031 |
Osc |
|
|
Pass |
|
|
|
Andon KD-5915 Equivalence |
Recommended E43 |
|
|
|
Pass |
|
|
At rest |
Recommended 117 |
Andon KD-5851 |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 102 |
Andon KD-5913 |
Osc |
|
|
Pass |
|
|
|
Andon KD-5915 Equivalence |
Recommended E45 |
Andon KD-5915 |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 57 |
Andon KD-5917 |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 95 |
Andon KD-5963 |
Osc |
|
|
Pass |
|
|
|
Andon KD-5915 Equivalence |
Recommended E44 |
Andon KD-5965 |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 104 |
Andon MD-534 |
Osc |
|
|
Pass |
|
|
|
Andon KD-5915 Equivalence |
Recommended E46 |
AVITA BPM63S |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 103 |
AVITA BPM64 |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 121 |
Beurer BM27 |
Osc |
|
|
|
Pass |
|
|
BEURER BM28 Equivalence |
Recommended E284 |
Beurer BM28 |
Osc |
|
|
|
Pass |
|
|
At Rest |
Recommended E130 |
Beurer BM35 |
Osc |
|
|
Pass |
|
|
|
HL-868BA Equivalence |
Recommended E67 |
Beurer BM35/1 |
Osc |
|
|
|
Pass |
|
|
SEJOY BP-1307 Equivalence |
Recommended E219 |
Beurer BM44 |
Osc |
|
|
Pass |
|
|
|
Andon KD-5915 Equivalence |
Recommended E68 |
Beurer BM45 |
Osc |
|
|
|
Pass |
|
|
Grandway G.Lab MD2680 Equivalence |
Recommended E228 |
Beurer BM47 |
Osc |
|
|
Pass |
|
|
|
Andon KD-5915 Equivalence |
Recommended E105 |
Beurer BM51 |
Osc |
|
|
|
Pass |
|
|
AVITA BPM64 Equivalence |
Recommended
E216 |
Beurer BM53 |
Osc |
|
|
|
Pass |
|
|
Beurer BM28 Equivalence |
Recommended
E364 |
Beurer BM54 |
Osc |
|
A/A |
|
Pass |
Pass |
|
G.LAB MD2680 Equivalence |
Recommended E214 |
Beurer BM55/BM66 |
Osc |
|
|
Pass |
|
|
|
Andon KD-5915 Equivalence |
Recommended E131 |
Beurer BM58 |
Osc |
|
|
Pass |
|
|
|
Andon KD-5915 Equivalence |
Recommended E78 |
Beurer BM60 |
Osc |
|
|
Pass |
|
|
|
DS-400 Equivalence |
Recommended E69 |
Beurer BM65 |
Osc |
|
|
Pass |
|
|
|
Andon KD-5915 Equivalence |
Recommended E79 |
Beurer BM67 |
Osc |
|
A/A |
|
Pass |
Pass |
|
G.LAB MD2680 Equivalence |
Recommended E213 |
Beurer BM81 |
Osc |
|
|
|
Pass |
|
|
Grandway MD2301 Equivalence |
Recommended E220 |
Beurer BM85 |
Osc |
|
|
|
Pass |
|
|
iHealth BP3 Equivalence |
Recommended E126 |
Beurer BM95 |
Osc |
|
|
|
Pass |
|
|
SEJOY BP-1307 Equivalence |
Recommended E173 |
Beurer SBM43 |
Osc |
|
|
Pass |
|
|
|
Andon KD-5915 Equivalence |
Recommended E80 |
Beurer SBM45 |
Osc |
|
|
Pass |
|
|
|
Andon KD-5915 Equivalence |
Recommended E81 |
BIOS BD204 |
Osc |
|
|
Pass |
|
|
|
LD-578 Equivalence |
Recommended E74 |
BIOS BD209 |
Osc |
|
|
Pass |
|
|
|
LD-578 Equivalence |
Recommended E66 |
Blipcare BP 800 |
Osc |
|
|
|
Pass |
|
|
Transtek TMB-1491 Equivalence |
Recommended E207 |
Braun BP4600 |
Osc |
|
|
Pass |
|
|
|
HL-868BA Equivalence |
Recommended E35 |
Braun BP4900 |
Osc |
|
|
Pass |
|
|
|
HL-868BA Equivalence |
Recommended E40 |
Braun BP5900 |
Osc |
|
|
Pass |
|
|
|
HL-868BA Equivalence |
Recommended E41 |
Braun BP6000 |
Osc |
|
|
Pass |
|
|
|
TMB-986 Equivalence |
Recommended E95 |
Braun BP6100 |
Osc |
|
|
Pass |
|
|
|
TMB-986 Equivalence |
Recommended E96 |
Braun BUA6150 |
Osc |
|
|
Pass |
|
|
|
TMB-986 Equivalence |
Recommended E182 |
Braun BUA6350 |
Osc |
|
|
Pass |
|
|
|
TMB-986 Equivalence |
Recommended E208 |
Braun BP6200 |
Osc |
|
|
Pass |
|
|
|
TMB-986 Equivalence |
Recommended E97 |
Braun BUA5000
(BUA5000LA,BUA5000LAD1) |
Osc |
|
|
Pass |
|
|
|
Microlife BP A100 Plus Equivalence |
Recommended E127 |
Braun BUA7200 |
Osc |
|
|
|
Pass |
|
|
AVITA BPM64 Equivalence |
Recommended E177 |
CA-MI Srl CHIESI 0100000328 (DIESIS PLUS) |
Osc |
|
|
|
Pass |
|
|
Nissei DSK-1011 Equivalence |
Recommended E139 |
CA-MI Srl CHIESI 0100000329 (DIESIS) |
Osc |
|
|
|
Pass |
|
|
Nissei DSK-1011 Equivalence |
Recommended E138 |
Citizen CH-432B |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 40 |
Citizen CH-452 |
Osc |
|
|
Pass |
|
|
|
CH-432B Equivalence |
Recommended E73 |
Citizen CHU304 |
Osc |
|
|
Pass |
|
|
|
Andon KD-5915 Equivalence |
Recommended E123 |
Citizen CHU305 |
Osc |
|
|
|
Pass |
|
|
Andon KD-5917 Equivalence |
Recommended E124 |
Citizen CHUD514 |
Osc |
|
|
|
Pass |
|
|
Andon KD-595 Equivalence |
Recommended E179 |
Citizen CHUD517 |
Osc |
|
|
|
Pass |
|
|
Andon KD-5917 Equivalence |
Recommended E180 |
Colson MAM BP3AA1-2 |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 16 |
Foracare Fora D 30 |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 72 |
Foracare Fora D 40b |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 75 |
Foracare Fora P30 Plus |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 82 |
Foracare Fora P80 Diamond Cuff BP
(Fora Test N'Go BP) |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 94 |
G.LAB MD2680 |
Osc |
|
A/A |
|
Pass |
Pass |
|
At rest |
Recommended 119 |
G.LAB MD4180 |
Osc |
|
|
|
Pass |
|
|
Grandway MD2301 Equivalence |
Recommended E153 |
Pass |
A/A |
|
|
|
|
Grandway MD2301 Equivalence |
Recommended E158 |
Globalcare GCE603 |
Osc |
|
|
|
Pass |
|
|
At rest; Standard cuff |
Recommended 128 |
Grandway MD2301 |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 108 |
Pass |
A/A |
|
|
|
|
At rest |
Recommended 110 |
Greater Goods 0040 |
Osc |
|
|
|
Pass |
|
|
Transtek TMB-1491 Equivalence |
Recommended E221 |
Hanvon FF680 |
Aus |
|
|
|
|
|
Pass |
At rest |
Recommended 243 |
HARTMANN Tensoval Duo Control |
Aus-Osc |
Pass |
A/A |
Pass |
|
|
|
At rest |
Recommended 45 |
HARTMANN Tensoval Duo Control II |
Aus-Osc |
Pass |
A/A |
|
Pass* |
|
|
*ESH-IP 2010 in Aus mode only |
Recommended 87 |
HARTMANN Veroval Duo Control |
Aus-Osc |
Pass |
A/A |
|
Pass* |
|
|
*ESH-IP 2010 in Aus mode only
HARTMANN Tensoval Duo Control II Equivalence |
Recommended E181 |
HARTMANN Veroval BPU22 |
Osc |
|
|
|
Pass |
|
|
Globalcare - GCE603 Equivalence |
Recommended E358 |
HARTMANN Veroval BPU26 |
Osc |
|
|
|
Pass |
|
|
Globalcare - GCE603 Equivalence |
Recommended E359 |
HARTMANN Veroval ECG and BP Monitor |
Osc |
|
|
|
Pass |
|
|
SEJOY BP-1307 Equivalence |
Recommended E164 |
HARTMANN Veroval upper arm BP monitor |
Osc |
|
|
|
Pass |
|
|
BPM63S Equivalence |
Recommended E159 |
Health and Life
(Health & Life)
HL-868BA |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 44 |
Health and Life
(Health & Life)
HL-868ED |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 127 |
Health and Life
(Health & Life)
HL-888HA |
Osc |
|
Pass |
|
|
|
|
Abstract; Protocol adjustments |
Questionable 21 |
|
A/A |
|
|
|
|
At rest |
Recommended 26 |
Honsun LD-30 |
Osc |
|
|
Pass |
|
|
|
LD-578 Equivalence |
Recommended E66 |
Honsun LD-578 |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 47 |
IEM
Stabil-O-Graph |
Osc |
|
A/A |
Pass |
|
|
|
At rest |
Recommended 27 |
iHealth BP3 (KD-931) |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 70 |
|
|
Pass |
|
|
|
KD-5915 Equivalence |
Recommended E60 |
iHealth BP5 |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 85 |
Inventum BDA432 |
Osc |
|
|
|
Pass |
|
|
Transtek TMB-1491 Equivalence |
Recommended E155 |
Inventum BDA632 |
Osc |
|
A/A |
|
Pass |
|
|
Pangao PG-800B11 Equivalence |
Recommended E157 |
ICO MT-20 |
Osc |
|
|
|
Pass |
|
|
Andon KD-5965 Equivalence |
Recommended E326 |
ICO MT-30 |
Osc |
|
|
|
Pass |
|
|
Andon KD-5851 Equivalence |
Recommended E327 |
JAMR B23 |
Osc |
|
|
|
|
|
Pass (+1:2020 |
At rest |
Recommended 251 |
JAMR F1701T |
Osc |
|
|
|
|
|
Pass |
At rest |
Recommended 178 |
JOYTECH DBP-1333b |
Osc |
|
|
|
|
|
Pass (+1:2020) |
At rest |
Recommended 194 |
JOYTECH DBP-6279B |
Osc |
|
|
|
|
|
Pass |
At rest |
Recommended 131 |
Kinetik Wellbeing WBP1 |
Osc |
|
|
|
Pass |
|
|
Transtek TMB-1491 Equivalence |
Recommended E198 |
Kinetik Wellbeing TMB-1970 |
Osc |
|
|
|
Pass |
|
|
Transtek TMB-986 Equivalence |
Recommended E225 |
Kinetik Wellbeing TMB-2088 |
Osc |
|
|
Pass |
|
|
|
TRANSTEK TMB-986 Equivalence |
Recommended E227 |
Kinetik Wellbeing TMB-2080 |
Osc |
|
|
|
Pass |
|
|
TRANSTEK TMB-1491 Equivalence |
Recommended E229 |
Kingyield BP101H |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 60 |
Lloyds Pharmacy BP11 |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 58 |
Medel Check (92187) |
Osc |
|
|
Pass |
|
|
|
KD-5915 Equivalence |
Recommended E64 |
Medel Check (95124) |
Osc |
|
|
|
Pass |
|
|
AViTA BPM63S Equivalence |
Recommended E148 |
Medel Control |
Osc |
|
|
|
Pass |
|
|
AVITA BPM63S Equivalence |
Recommended E147 |
Medel Elite (92125) |
Osc |
|
|
Pass |
|
|
|
KD-5915 Equivalence |
Recommended E65 |
Medel Idea (91913 IDEA) |
Osc |
Pass |
A/A |
Pass |
|
|
|
SAA-102 Equivalence |
Recommended E33 |
Medipro MediCare 100f |
Osc |
|
|
|
Pass |
|
|
At rest (Note: Not IBP Medipro 100f) |
Recommended 68 |
Microlife BP 3AC1-1 |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 15 |
Pass |
A/A |
|
|
|
|
Pregnancy; Abstract |
Questionable 20 |
Pass |
B/A |
|
|
|
|
Pre-eclampsia; Abstract |
Questionable 20 |
Microlife BP 3AC1-1 PC |
Osc |
|
|
Pass |
|
|
|
BP 3AC1-1 Equivalence |
Recommended E1 |
|
|
Pass |
|
|
|
ESRD population |
Recommended 28 |
Microlife BP 3AC1-2 |
Osc |
|
|
Pass |
|
|
|
BP 3AC1-1 Equivalence |
Recommended E2 |
Microlife BP 3AG1 |
Osc |
|
A/A |
|
|
|
|
BP 3BT0-A Equivalence |
Recommended E3 |
Microlife BP 3BT0-1 |
Osc |
|
A/A |
|
|
|
|
BP 3BT0-A Equivalence |
Recommended E4 |
Microlife BP 3BT0-A |
Osc |
|
A/A |
|
|
|
|
Small recruitment violation |
Recommended 7 |
Pass |
A/B |
|
|
|
|
Normotensive Pregnancy |
Recommended 12 |
Pass |
B/B |
|
|
|
|
Non-proteinuric HBP |
Pass |
A/B |
|
|
|
|
Pre-Eclampsia |
Microlife BP 3BT0-A(2) |
Osc |
|
A/A |
|
|
|
|
BP 3BT0-A Equivalence |
Recommended E5 |
Microlife BP 3BT0-AP |
Osc |
|
A/A |
|
|
|
|
BP 3BT0-A Equivalence |
Recommended E6 |
Microlife BP A1 Easy (BP3GR1-1P) |
Osc |
|
A/A |
Pass |
|
|
|
BP A100 Plus and BP 3BT0-A Equivalences |
Recommended E106 |
Pass |
A/B |
|
|
|
|
Normotensive Pregnancy |
Pass |
B/B |
|
|
|
|
Non-proteinuric HBP |
Pass |
A/B |
|
|
|
|
Pre-Eclampsia |
Microlife BP A2 Basic (BP3GQ1-3P) |
Osc |
|
A/A |
Pass |
|
|
|
BP A100 Plus and BP 3BT0-A Equivalences |
Recommended E107 |
Pass |
A/B |
|
|
|
|
Normotensive Pregnancy |
Pass |
B/B |
|
|
|
|
Non-proteinuric HBP |
Pass |
A/B |
|
|
|
|
Pre-Eclampsia |
Microlife BP A2 Classic (3UG1-2E) |
Osc |
|
A/A |
|
|
|
|
Microlife BP 3BT0-A Equivalence |
Recommended E150 |
|
|
Pass |
|
|
|
Microlife BP A100 Equivalence |
Microlife BP A3 PC |
Osc |
|
|
|
Pass |
|
|
General population |
Recommended 99 |
Microlife BP A6 PC (BP3GU1-8Y) |
Osc |
|
A/A |
Pass |
|
|
|
BP A100 Plus and BP 3BT0-A Equivalences |
Recommended E108 |
Pass |
A/B |
|
|
|
|
Normotensive Pregnancy |
Pass |
B/B |
|
|
|
|
Non-proteinuric HBP |
Pass |
A/B |
|
|
|
|
Pre-Eclampsia |
Microlife BP A100 |
Osc |
|
|
Pass |
|
|
|
BP A100 Plus Equivalence |
Recommended E7 |
|
|
Pass |
|
|
|
With M-L Soft Cuff |
Recommended FTS, 52 |
|
|
Pass |
|
|
|
With M-L Rigid Conical Cuff |
Recommended 37 |
Microlife BP A100 Plus |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 17 |
|
|
Pass |
|
|
|
At rest |
Recommended 29 |
Microlife BP A150 Afib |
Osc |
|
|
Pass |
|
|
|
Microlife BP A100 Equivalence |
Recommended E149 |
|
|
Pass |
|
|
|
Microlife BP A100 Plus Equivalence |
Microlife Exact BP (BP3MD1-3) |
Osc |
|
|
Pass |
|
|
|
BP A100 Equivalence |
Recommended E21 |
Microlife RM 100 |
Osc |
|
A/A |
|
|
|
|
BP 3BT0-A Equivalence |
Recommended E8 |
Microlife WatchBP Home (BP3MX1-1) |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 25 |
|
A/A |
|
|
|
|
Pregnancy |
Recommended 66 |
|
B/A |
|
|
|
|
Pre-Eclampsia |
Recommended 66 |
Microlife WatchBP Home A (BP3MX1-3) |
Osc |
|
|
Pass |
|
|
|
Watch BP Home Equivalence |
Recommended E27 |
Microlife WatchBP Home N (BP3MX1-4) |
Osc |
|
|
Pass |
|
|
|
Watch BP Home Equivalence |
Recommended E70 |
Microlife WatchBP Home S (BP3MX1-5) |
Osc |
|
|
Pass |
|
|
|
Watch BP Home Equivalence |
Recommended E28 |
Microlife WatchBP O3 (3MZ0) |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 61 |
Microlife WatchBP O3 (3MZ1) |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 62 |
Nissei DS-400 |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 48 |
Nissei DS-500 |
Osc |
|
|
Pass |
|
|
|
DS-400 Equivalence |
Recommended E38 |
Nissei DS-A10 |
Osc |
|
|
|
Pass |
|
|
DSK-1011 Equivalence |
Recommended E121 |
Nissei DS-B10 |
Osc |
|
|
|
Pass |
|
|
DSK-1011 Equivalence |
Recommended E122 |
Nissei DS-G10 |
Osc |
|
|
|
Pass |
|
|
DSK-1011 Equivalence |
Recommended E128 |
Nissei DS-N10 |
Osc |
|
|
|
Pass |
|
|
DSK-1031 Equivalence |
Recommended E129 |
Nissei DS-S10 |
Osc |
|
|
|
Pass |
|
|
DSK-1031 Equivalence |
Recommended E130 |
Nissei DSK-1011 |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 76 |
Nissei DSK-1031 |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 77 |
Nissei DSK-1051 |
Osc |
|
|
|
Pass |
|
|
DSK-1031 Equivalence |
Recommended E98 |
Omron 705IT
(HEM-759-E) |
Osc |
|
|
Pass |
|
|
|
|
Recommended 9 |
Pass |
A/A |
|
|
|
|
Minor recruitment modification |
Recommended 14 |
|
|
Pass |
|
|
|
Children and adolescents. Reasonable adaptation of protocol. |
Recommended 22 |
Omron 708-BT
(HEM-7081-ITE) |
Osc |
Pass |
A/A |
|
|
|
|
Omron M7 (HEM-780-E) Equivalence |
Recommended E103 |
Omron BP760
(HEM-7220-Z) |
Osc |
|
|
Pass |
|
|
|
Omron M6 Comfort (HEM-7000-E) Equivalence
New sensor |
Recommended E87 |
Omron BP785
(HEM-7222-Z) |
Osc |
|
|
Pass |
|
|
|
Omron M6 Comfort (HEM-7000-E) Equivalence
New sensor |
Recommended E88 |
Omron EVOLV
(HEM-7600T-E) |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 112 |
|
|
|
Pass* |
|
|
Pregnancy
* Modified ESH-IP 2010 |
Recommended 123 |
Omron HEM-705CP-II
(HEM-759-E2) |
Osc |
|
|
Pass |
|
|
|
705IT (HEM-759-E) Equivalence |
Recommended E12 |
Omron HEM-746C
(HEM-746C-SH) |
Osc |
|
|
Pass |
|
|
|
Omron M2 Compact (HEM-7102-E) Equivalence |
Recommended E54 |
Omron HEM-780REL |
Osc |
|
|
Pass |
|
|
|
At rest; "East Wrap" cuff. Poor study and paper |
Questionable 32 |
Omron HEM-1020 |
Osc |
|
|
Pass |
|
|
|
Supplementary Results |
Recommended 71 |
Omron HEM-7051
(HEM-7051-SH) |
Osc |
|
|
Pass |
|
|
|
Omron M3 Intellisense (HEM-7051-E) Equivalence |
Recommended E48 |
Omron HEM-7080IC |
Osc |
Pass |
A/A |
Pass |
|
|
|
705IT (HEM-759-E) Equivalences |
Recommended E20 |
Omron HEM-7101
(HEM-7101-SH) |
Osc |
|
|
Pass |
|
|
|
Omron M2 Compact (HEM-7102-E) Equivalence |
Recommended E53 |
Omron HEM-7130 (Japanese model) |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 88 |
Omron HEM-7250-IT |
Osc |
|
|
|
Pass |
|
|
Omron HEM-7251G Equivalence |
Recommended E102 |
Omron HEM-7251G |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 83 |
Omron HEM-7252G-HP |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 98 |
Omron HEM-7320F |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 89 |
Omron HEM-7500F |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 90 |
Omron HEM-Solar
(HEM-4500-SOLE) |
Osc |
|
|
Pass |
|
|
|
Omron M1 Plus (HEM-4011C-E) Equivalence |
Recommended E26 |
Omron IA2
(HEM-7011-C1) |
Osc |
Pass |
A/A |
Pass |
|
|
|
705IT (HEM-759-E) Equivalence |
Recommended E13 |
Omron i-C10
(HEM-7070-E) |
Osc |
|
|
Pass |
|
|
|
Omron M6 Comfort (HEM-7000-E) Equivalence |
Recommended E18 |
Pass |
A/A |
|
|
|
|
Omron M7 (HEM-780-E) Equivalence |
Recommended E19 |
Omron i-Q132 SpotArm
(HEM-1010-E) |
Osc |
|
|
Pass |
|
|
|
Supplementary Results |
Recommended 71 |
Omron i-Q142 SpotArm
(HEM-1040-E) |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 81 |
Omron M1
(HEM-4030-E) |
Osc |
|
|
Pass |
|
|
|
Omron M1 Plus (HEM-4011C-E) Equivalence |
Recommended E59 |
Omron M1 Classic
(HEM-442-E) |
Osc |
|
|
Pass |
|
|
|
Omron M1 Plus (HEM-4011C-E) Equivalence |
Recommended E10 |
Omron M1 Compact
(HEM-4022-E) |
Osc |
|
|
Pass |
|
|
|
Omron M1 Plus (HEM-4011C-E) Equivalence |
Recommended E11 |
Omron M1 Plus
(HEM-4011C-E) |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 29 |
Omron M2
(HEM-7117-E) |
Osc |
|
|
Pass |
|
|
|
Omron M3 Intellisense (HEM-7051-E) Equivalence |
Recommended E49 |
Omron M2
(HEM-7117-E8) |
Osc |
|
|
Pass |
|
|
|
Omron M3 Intellisense (HEM-7051-E) Equivalence
New sensor |
Recommended E83 |
Omron M2
(HEM-7119-E) |
Osc |
|
|
Pass |
|
|
|
Omron M3 Intellisense (HEM-7051-E) Equivalence
New sensor |
Recommended E90 |
Omron M2
(HEM-7121-E) |
Osc |
|
|
|
Pass |
|
|
Omron HEM-7130 (Japanese model) Equivalence |
Recommended E112 |
Omron M2 Basic
(HEM-7116-E) |
Osc |
|
|
Pass |
|
|
|
Omron M3 Intellisense (HEM-7051-E) Equivalence |
Recommended E50 |
Omron M2 Basic
(HEM-7116-E2) |
Osc |
|
|
Pass |
|
|
|
Omron M3 Intellisense (HEM-7051-E) Equivalence
New sensor |
Recommended E89 |
Omron M2 Basic
(HEM-7116-E8) |
Osc |
|
|
Pass |
|
|
|
Omron M3 Intellisense (HEM-7051-E) Equivalence
New sensor |
Recommended E82 |
Omron M2 Basic
(HEM-7120-E) |
Osc |
|
|
|
Pass |
|
|
Omron HEM-7130 (Japanese model) Equivalence |
Recommended E111 |
Omron M2 Compact
(HEM-7102-E) |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 31 |
Omron M3
(HEM-7131-E) |
Osc |
|
|
|
Pass |
|
|
Omron M6 AC (HEM-7322-E) Equivalence |
Recommended E113 |
Omron M3
(HEM-7154-E) |
Osc |
|
|
|
Pass |
|
|
Omron M6 AC (HEM-7322-E) Equivalence |
Recommended E197 |
Omron M3
(HEM-7200-E) |
Osc |
|
|
Pass |
|
|
|
Omron M3 Intellisense (HEM-7051-E) Equivalence |
Recommended E51 |
|
|
|
Pass |
|
|
Supplementary Results |
Recommended 65 |
Omron M3
(HEM-7200-E2) |
Osc |
|
|
Pass |
|
|
|
Omron M3 Intellisense (HEM-7051-E) Equivalence
New sensor |
Recommended E91 |
Omron M3
(HEM-7200-E8) |
Osc |
|
|
Pass |
|
|
|
Omron M3 Intellisense (HEM-7051-E) Equivalence
New sensor |
Recommended E84 |
Omron M3 Comfort
(HEM-7134-E) |
Osc |
|
|
|
Pass |
|
|
Omron M6 Comfort(HEM-7321-E) Equivalence |
Recommended E140 |
|
|
|
Pass* |
|
|
Pregnancy
* Modified ESH-IP 2010 |
Recommended 122 |
Omron M3 Comfort
(HEM-7155-E) |
Osc |
|
|
|
Pass |
|
|
Omron M6 Comfort(HEM-7321-E) Equivalence |
Recommended E199 |
Omron M3 Intellisense
(HEM-7051-E) |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 31 |
Omron M3 IT
(HEM-7131U-E) |
Osc |
|
|
|
Pass |
|
|
Omron M6 AC (HEM-7322-E) Equivalence |
Recommended E114 |
Omron M3W
(HEM-7202-E) |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 78 |
Omron M3 Comfort
(HEM-7155-E) |
Osc |
|
|
|
Pass |
|
|
Omron M6 Comfort(HEM-7321-E) Equivalence |
Recommended E199 |
Omron M4 Intelli IT
(HEM-7155T-EBK) |
Osc |
Pass |
A/A |
Pass |
|
|
|
705IT (HEM-759-E) Equivalence |
Recommended E200 |
Omron M5-I
(HEM-757-E) |
Osc |
|
|
Pass |
|
|
|
|
Recommended 9 |
|
|
Pass |
|
|
|
Elderly Population |
Recommended 33 |
Omron M6
(HEM-7001-E) |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 19 |
|
|
Pass |
|
|
|
General Population |
Recommended 34 |
|
|
Pass |
|
|
|
Obese Population |
Recommended 34 |
|
|
Pass |
|
|
|
Elderly Population |
Recommended 23 |
Pass |
A/A |
|
|
|
|
705IT (HEM-759-E) Equivalence |
Recommended E23 |
Omron M6
(HEM-7211-E) |
Osc |
Pass |
A/A |
Pass |
|
|
|
705IT (HEM-759-E) Equivalences |
Recommended E58 |
|
|
|
Pass |
|
|
Supplementary Results |
Recommended 65 |
Omron M6
(HEM-7211-E8) |
Osc |
|
|
Pass |
|
|
|
705IT (HEM-759-E) Equivalence
New sensor |
Recommended E85 |
Omron M6 AC
(HEM-7322-E) |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 93 |
Omron M6 Comfort
(HEM-7000-E) |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 29 |
Pass |
A/A |
|
|
|
|
Omron M7 (HEM-780-E) Equivalence |
Recommended E17 |
Omron M6 Comfort
(HEM-7221-E) |
Osc |
|
|
Pass |
|
|
|
Omron M6 Comfort (HEM-7000-E) Equivalence |
Recommended E56 |
Pass |
A/A |
|
|
|
|
Omron M7 (HEM-780-E) Equivalence |
Recommended E57 |
Omron M6 Comfort
(HEM-7221-E8) |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 80 |
|
|
Pass |
|
|
|
Omron M6 Comfort (HEM-7000-E) Equivalence
New sensor |
Recommended E86 |
Omron M6 Comfort
(HEM-7223-E) |
Osc |
|
|
|
Pass |
|
|
Omron M6 Comfort (HEM-7221-E8) Equivalence |
Recommended E92 |
Omron M6 Comfort
(HEM-7321-E) |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 92 |
Omron M6 Comfort
(HEM-7360-E) |
Osc |
|
|
|
Pass |
|
|
Omron M6 Comfort(HEM-7321-E) Equivalence |
Recommended E201 |
Omron M6 Comfort IT
(HEM-7322U-E) |
Osc |
|
|
|
Pass |
|
|
Omron M6 Comfort (HEM-7321-E) Equivalence |
Recommended E115 |
|
|
|
Pass |
|
|
In a diabetic population |
Recommended 114 |
Omron M6W
(HEM-7213-E) |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 79 |
Omron M7
(HEM-780-E) |
Osc |
Pass |
A/A |
|
|
|
|
General population |
Recommended 38 |
|
|
Pass |
|
|
|
Population requiring Large Cuff |
Recommended 35 |
|
A/A |
|
|
|
|
Pregnancy |
Recommended 50 |
|
B/B |
|
|
|
|
Pre-Eclampsia |
Recommended 50 |
Omron M7 Intelli IT
(HEM-7322T-E) |
Osc |
|
|
|
Pass |
|
|
Omron M6 Comfort(HEM-7321-E) Equivalence |
Recommended E154 |
Omron M7 Intelli IT
(HEM-7361T-EBK) |
Osc |
|
|
|
Pass |
|
|
Omron M6 Comfort(HEM-7321-E) Equivalence |
Recommended E202 |
Omron M10-IT
(HEM-7080IT-E) |
Osc |
|
|
Pass |
|
|
|
Omron M6 Comfort (HEM-7000-E) Equivalence |
Recommended E16 |
Pass |
A/A |
|
|
|
|
Omron M7 (HEM-780-E) Equivalence |
Recommended E9 |
Omron MX2 Basic
(HEM-742-E2) |
Osc |
|
|
Pass |
|
|
|
Omron M2 Compact (HEM-7102-E) Equivalence |
Recommended E55 |
Omron MX3 Plus
(HEM-742-E) |
Osc |
|
|
Pass |
|
|
|
Subject to corrected tables being published |
Recommended 13, FTS |
Omron SEM-1
(HEM-7051-C12) |
Osc |
|
|
Pass |
|
|
|
Omron M3 Intellisense (HEM-7051-E) Equivalence |
Recommended E52 |
Omron X3 Comfort
(HEM-7155-EO) |
Osc |
|
|
|
Pass |
|
|
Omron M6 Comfort(HEM-7321-E) Equivalence |
Recommended E203 |
Omron X4 Smart
(HEM-7155T-ESL) |
Osc |
|
|
|
Pass |
|
|
Omron M6 Comfort(HEM-7321-E) Equivalence |
Recommended E206 |
Omron X6 Comfort
(HEM-7360-EO) |
Osc |
|
|
|
Pass |
|
|
Omron M6 Comfort(HEM-7321-E) Equivalence |
Recommended E204 |
Omron X7 Smart
(HEM-7361T-ESL) |
Osc |
|
|
|
Pass |
|
|
Omron M6 Comfort(HEM-7321-E) Equivalence |
Recommended E205 |
Oregon Scientific BPU 330 |
Osc |
|
|
Pass |
|
|
|
At rest, Plots incorrect |
Recommended 49 |
Oregon Scientific BPW810 |
|
|
|
Pass |
|
|
|
At rest |
Questionable 24 |
Panasonic EW3106 |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 69, FTS |
Panasonic EW3109 |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 69, FTS |
Philips DL8760 |
Osc |
|
|
|
Pass |
|
|
At rest
General Population |
Recommended 120 |
Pangao PG-800B11 |
Osc |
|
A/A |
|
Pass |
|
|
General Population |
Recommended 101 |
Pic Solution (Pikdare, formely Artsana)
Pic Indolor (Pikdare, formely Artsana)
CARDIOafib |
Osc |
|
|
|
Pass |
|
|
Rossmax CF175 Equivalence |
Recommended E163 |
Pic Solution (Artsana)
Pic Indolor (Artsana)
CARDIO maxi |
Osc |
|
|
|
Pass |
|
|
Transtek TMB-1491 Equivalence |
Recommended E144 (v. 2019) |
Pic Solution (Artsana)
Pic Indolor (Artsana)
CARDIOSimple |
Osc |
|
|
|
Pass |
|
|
Andon KD-5917 Equivalence |
Recommended E146 |
Pic Solution (Artsana)
Pic Indolor (Artsana)
Classic Check (100) |
Osc |
|
|
Pass |
|
|
|
My Check Equivalence |
Recommended E30 |
Pic Solution (Artsana)
Pic Indolor (Artsana)
Classic Check (200) |
Osc |
|
|
Pass |
|
|
|
My Check Equivalence |
Recommended E72 |
Pic Solution (Pikdare, formely Artsana)
Pic Indolor (Pikdare, formely Artsana)
clearRAPID |
Osc |
|
|
|
Pass |
|
|
Transtek TMB-1491 Equivalence |
Recommended E143 (v. 2019) |
Pic Solution (Artsana)
Pic Indolor (Artsana)
Comfort Check |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 54 |
Pic Solution (Artsana)
Pic Indolor (Artsana)
CS 410 |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 41 |
Pic Solution (Artsana)
Pic Indolor (Artsana)
CSI 610 |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 39 |
Pic Solution (Artsana)
Pic Indolor (Artsana)
Daily Check |
Osc |
|
|
Pass |
|
|
|
My Check Equivalence |
Recommended E75 |
Pic Solution (Artsana)
Pic Indolor (Artsana)
Digit Extra |
Osc |
|
|
Pass |
|
|
|
KD-5915 Equivalence |
Recommended E93 |
Pic Solution (Artsana)
Pic Indolor (Artsana)
Digit Smart |
Osc |
|
|
Pass |
|
|
|
KD-5915 Equivalence |
Recommended E94 |
Pic Solution (Pikdare, formely Artsana)
Pic Indolor (Pikdare, formely Artsana)
easyRAPID |
Osc |
|
|
|
Pass |
|
|
Transtek TMB-1491 Equivalence |
Recommended E141 (v. 2019) |
Pic Solution (Pikdare, formely Artsana)
Pic Indolor (Pikdare, formely Artsana)
helpRAPID |
Osc |
|
|
|
Pass |
|
|
Transtek Artsana helpRAPID (Transtek TMB-1586-A) Equivalence |
Recommended E160 (v. 2019) |
Pic Solution (Artsana)
Pic Indolor (Artsana)
Help Check |
Osc |
|
|
Pass |
|
|
|
My Check Equivalence |
Recommended E76 |
Pic Solution (Pikdare, formely Artsana)
Pic Indolor (Pikdare, formely Artsana)
liteRAPID ARM (REF 020225333000000) |
Osc |
|
|
|
Pass |
|
|
Transtek TMB-1776 Equivalence |
Recommended E222 |
Pic Solution (Artsana)
Pic Indolor (Artsana)
Mamy Check |
Osc |
|
|
Pass |
|
|
|
My Check Equivalence |
Recommended E77 |
Pic Solution (Pikdare, formely Artsana)
Pic Indolor (Pikdare, formely Artsana)
MaxiRAPID (REF 02022535000000) |
Osc |
|
|
|
Pass |
|
|
Transtek TMB-1491 Equivalence |
Recommended E223 |
Pic Solution (Artsana)
Pic Indolor (Artsana)
My Check |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 54 |
Pic Solution (Pikdare, formely Artsana)
Pic Indolor (Pikdare, formely Artsana)
mobileRAPID |
Osc |
|
|
|
Pass |
|
|
Transtek LS808-B Equivalence |
Recommended E165 |
Pic Solution (Artsana)
Pic Indolor (Artsana)
One Check |
Osc |
|
|
Pass |
|
|
|
My Check Equivalence |
Recommended E31 |
Pic Solution (Pikdare) OneRapid |
Osc |
|
|
|
|
|
Pass (2020-01) |
Transtek TMB-2296-BT Equivalence |
Recommended E384 |
Pic Solution (Artsana)
Pic Indolor (Artsana)
Personal Check |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 54 |
Polygreen K-Jump
KP-7670 |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 105 |
Polygreen K-Jump
KP-7770 |
Osc |
|
|
Pass |
|
|
|
K-Jump KP7500D Equivalence |
Recommended E71 |
Qardio Qardioarm |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 109 |
Rossmax CF175 |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 100 |
Rossmax ME 701 series |
Osc |
|
A/A |
Pass |
|
|
|
At rest |
Recommended 30 |
SAFETY PRONTEX INTEGRA (NISSEI DS-B33-03) |
Osc |
|
|
|
Pass |
|
|
Nissei DSK-1011 Equivalence |
Recommended E167 |
SAFETY PRONTEX SYNTESI (NISSEI DS-B23-03) |
Osc |
|
|
|
Pass |
|
|
Nissei DSK-1011 Equivalence |
Recommended E166 |
Sanity Family MD3150 |
Osc |
|
|
|
Pass |
|
|
G.Lab MD2680 Equivalence |
Recommended E402 |
SEJOY BP-1209 |
Osc |
|
|
|
Pass |
|
|
SEJOY BP-1307 Equivalence |
Recommended E175 |
SEJOY BP-1307 |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 116 |
Seinex SE-9400 |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 15 |
Sensacare SAA-102 (Standard) |
Osc |
Pass |
A/A |
Pass |
|
|
|
Three studies. BHS under reported. |
Recommended 36 |
Shenzhen Pump Medical System Co. Ltd. (BPUMP) BPUMP BF1112 |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 115 |
SmartHeart 01-509 |
Osc |
|
|
|
Pass |
|
|
Transtek LS808-B Equivalence |
Recommended E171 |
Spengler K-Jump
KP7500D |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 29 |
Spengler K-Jump
Pro M |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 42 |
Taidoc U-Right TD-3124 |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 91 |
Thermor Bios BD215 |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 86 |
Transtek LS808-B |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 113 |
Transtek TMB-986 |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 67 |
Transtek TMB-1491 |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 107 |
Transtek TMB-1776 |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 129 |
Transtek TMB-2266 |
Osc |
|
|
|
|
|
Pass (+1:2020) |
At rest |
Recommended 231 |
Transtek TMB-2296-BT |
Osc |
|
|
|
|
|
Pass (2020-01) |
At rest |
Recommended 164 |
Transtek Artsana helpRAPID
(Transtek TMB-1586-A) |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 111 |
UEBE Visomat Comfort 20/40 |
Osc |
|
|
Pass |
|
|
|
At Rest |
Recommended 43 |
UEBE Visomat Comfort E (Ref 24016) |
Osc |
|
|
|
Pass |
|
|
Double Comfort (Osc Mode) Equivalence |
Recommended E104 |
UEBE Visomat Comfort Eco |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 51 |
UEBE Visomat Comfort Eco (Ref 24026) |
Osc |
|
|
|
Pass |
|
|
Double Comfort (Osc Mode) Equivalence |
Recommended E135 |
UEBE Visomat Comfort Form |
Osc |
|
|
Pass |
|
|
|
At rest |
Recommended 55 |
UEBE Visomat Double Comfort |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 64, 74 |
Aus |
|
|
|
Pass |
|
|
At rest |
Recommended 63, 73 |
Welch-Allyn transtelephonic home monitor |
Osc |
Pass |
? |
|
|
|
|
Parkinson's disease, 30 sub * 2 measurements |
Questionable 5 |
YuWell YE660D |
Osc |
|
|
|
|
|
Pass (+1:2020) |
At rest |
Recommended 205 |
YuWell YE690A |
Osc |
|
|
|
Pass |
|
|
At rest |
Recommended 124 |
* Note: Grading criteria had not been established though BHS protocol was in operation.5
References for Validation of Upper Arm Devices for Self-measurement of Blood Pressure
1 |
Dieterle T, Battegay E, Bucheli B, Martina B. Accuracy and 'range of uncertainty' of oscillometric blood pressure monitors around the upper arm and wrist. Blood Press Monit 1998;3:339-46 |
2 |
Kobalava ZD, Kotovskaia IV, Rusakova OS, Babaeva LA. Validation of UA-767 Plus device for self measurement of blood pressure. Clin Pharmacol Ther 2003;12 (2):70-2 |
3 |
Rogoza AN, Pavlova TS, Sergeeva MV. Validation of A&D UA-767 device for the self-measurement of blood pressure. Blood Press Monit 2000;5(4):227-31 |
4 |
Longo D, Bertolo O, Toffanin G, Frezza P, Palatini P. Validation of the A&D UA-631 (UA-779 Life Source) device for self-measurement of blood pressure and relationship between its performance and large artery compliance. Blood Press Monit 2002;7:243-8 |
5 |
Herbst T, Anis-Anwar Y, Giacco S, White W. Evaluation of the overall system precision of the Welch-Allyn transtelephonic home blood pressure monitor in adults with Parkinson's disease. Blood Press Monit 2002;7:285-8 |
6 |
Longo D, Toffanin G, Garbelotto R, Zaetta V, Businaro L, Palatini P. Performance of the UA-787 oscillometric blood pressure monitor according to the European Society of Hypertension Protocol. Blood Press Monit 2003;8:91-5 |
7 |
Cuckson AC, Reinders A, Shabeeh H, Shennan AH. Validation of the Microlife BP 3BTO-A oscillometric blood pressure monitoring device according to a modified British Hypertension Society protocol. Blood Press Monit 2002;7(6):319-24 |
8 |
Koblava ZD, Kotovskaya YV, Rodionov E Validation of semi-automatic device UA-704 for self-measurement of blood pressure. Blood Press Monit 2005;10:223-5 |
9 |
El Assad M, Topouchian J, Asmar R. Evaluation of two devices for self-measurement of blood pressure according to the international protocol: the Omron M5-I and the Omron 705IT. Blood Press Monit 2003;8(3):127-33 |
11 |
Verdecchia P, Angeli F, Poeta F, Reboldi GP, Borgioni C, Pittavini L, Porcellati C. Validation of the A&D UA774 (UA-767Plus) device for self measurement of blood pressure. Blood Press Monit 2004;9:225-9 |
12 |
Reinders A, Cuckson AC, Lee JTM, Shennan AH. An accurate automated blood pressure device for use in pregnancy and pre-eclampsia: the Microlife 3BTO-A. BJOG 2005;112(7):915-20 |
13 |
Coleman A, Freeman P, Steel S, Shennan A Validation of the Omron MX3 Plus oscillometric blood pressure monitoring device according to the European Society of Hypertension international protocol. Blood Press Monit 2005;10:165-8 |
14 |
Coleman A, Freeman P, Steel S, Shennan A. Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol. Blood Press Monit 2006;11:27-32 |
15 |
Topouchian JA, El Assad MA, Orobinskaia LV, El Feghali RN, Asmar RG. Validation of two devices for self-measurement of brachial blood pressure according to the International Protocol of the European Society of Hypertension: the Seinex SE-9400 and the Microlife BP 3AC1-1. Blood Press Monit 2005;10:325-31 |
16 |
Pereira T, Maldonado J. Validação de aparelho de medição automática da pressão arterial Colson MAM BP3AA1-2 segundo o Protocolo da Sociedade Europeia de Hipertensão. Rev Port Cardiol 2005;24(11):1341-51 |
17 |
Stergiou GS, Giovas PP, Neofytou MS, Adamopoulos DN. Validation of the Microlife BP A 100 Plus device for self-home blood pressure measurement according to the International. Protocol Blood Press Monit 2006;11:157-60 |
18 |
Verdecchia P, Angeli F, Gattobigio R, Angeli E, Pede S, Pittavini L. Validation of the A&D UA-705 device for self-measurement of blood pressure according to the British Hypertension Society protocol. Blood Press Monit 2006;11:223-7 |
19 |
Topouchian JA, El Assaad MA, Orobinskaia LV, El Feghali RN, Asmar RG. Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT). Blood Press Monit 2006;11:165-71 |
20 |
de Greeff A, Reggiori F, Anthony J, Shennan A. The Microlife 3AC1: An accurate blood pressure measurement device in pregnancy and pre-eclampsia [abstract]. J Hypertens 2006;24(suppl 4):S279 Abstract P14.475 |
21 |
Tholl U, Assouline E. Validation of the automatic arm type non-invasive blood pressure monitor HL-888HA [abstract].J Hypertens 2006;24(suppl 4):S285 Abstract P14.496 |
22 |
Stergiou GS, Yiannes NG, Rarra VC. Validation of the Omron 705 IT oscillometric device for home blood pressure measurement in children and adolescents: The Arsakion School Study. Blood Press Monit 2006,11:229-34 |
23 |
Altunkan S, Iliman N, Altunkan E. Validation of the Omron M6 (HEM-7001-E) upper arm blood pressure measuring device according to the International Protocol in elderly patients. Blood Press Monit 2008;13:117-22 |
24 |
Li L, Wang Hui W, Jin Lan W. Validation of the Oregon Scientific BPW810 blood pressure monitor according to the European Society of Hypertension Protocol. Blood Press Monit 2006;11:343-7 |
25 |
Stergiou GS, Giovas PP, Charilaos P. Gkinos CP, Patouras JD. Validation of the Microlife WatchBP Home device for self home blood pressure measurement according to the International Protocol. Blood Press Monit 2007;12:185-8 |
26 |
de Greeff A, Shennan AH. The Health and Life device (HL888HA): accuracy assessment in an adult population according to the British Hypertension Society protocol. Blood Press Monit 2007;12:107-11 |
27 |
Westhoff TH, Schmidt S, Zidek W, van der Giet M. Validation of the Stabil-O-Graph blood pressure self-measurement device. Journal of Human Hypertension 2007;22:233-5 (Research Letter) |
28 |
Thompson AM, Eguchi K, Reznik ME, Shah SS, Pickering TG. Validation of an oscillometric home blood pressure monitor in an end-stage renal disease population and the effect of arterial stiffness on its accuracy. Blood Press Monit 2007;12:227-32 |
29 |
Belghazi J, El Feghali RN, Moussalem T, Rejdych M, Asmar RG. Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension. Vascular Health and Risk Management 2007;3(4):389-400 |
30 |
de Greeff A, Shennan AH. The Rossmax (ME 701 series) upper arm device: accuracy assessment in an adult population according to the International and the British Hypertension Society protocols. Blood Press Monit 2008;13:43-8 |
31 |
Asmar R, Khabouth J, Topouchian J, El Feghali R, Mattar J. Validation of three automatic devices for self-measurement of blood pressure according to the International Protocol: The Omron M3 Intellisense (HEM-7051-E), the Omron M2 Compact (HEM 7102-E), and the Omron R3-I Plus (HEM 6022-E). Blood Press Monit 2010;15(1):49-54 |
32 |
Viera AJ, Hinderliter AL. Validation of the HEM-780REL with easy wrap cuff for self-measurement of blood pressure according to the European Society of Hypertension International Protocol. Blood Press Monit 2007;12:335-8 |
33 |
Omboni S, Riva I, Giglio I, Caldara G, Groppelli A, Parati G. Validation of the Omron M5-I, R5-I and HEM-907 automated blood pressure monitors in elderly individuals according to the International Protocol of the European Society of Hypertension. Blood Press Monit 2007;12:233-42 |
34 |
Altunkan S, Iliman N, Kayaturk N, Altunkan E. Validation of the Omron M6 (HEM-7001-E) upper-arm blood pressure measuring device according to the International Protocol in adults and obese adults. Blood Press Monit 2007;12:219-25 |
35 |
El Feghali RN, Topouchian JA, Pannier BM, El Assaad HA, Asmar RG. Validation of the Omron M7 (HEM-780-E) blood pressure measuring device in a population requiring large cuff use according to the International Protocol of the European Society of Hypertension. Blood Press Monit 2007;12:173-8 |
36 |
Zaetta V, Daniele L, Perkovic D, Prattico F, Barisa M, Perfetti P, Gabrieli A, Buonocore F, Winnicki M. Validation of the SAA-102 home blood pressure monitor according to the protocols of the European Society of Hypertension, the Association for the Advancement of Medical Instrumentation and the British Society of Hypertension. Blood Press Monit 2007;12:363-8 |
37 |
Bonso E, Saladini F, Zanier A, Benetti E, Dorigatti F, Palatini P. Accuracy of a single rigid conical cuff with standard-size bladder coupled to an automatic oscillometric device over a wide range of arm circumferences. Hypertens Res. 2010;33(11):1186-91. |
38 |
Coleman A, Steel S, Freeman P, de Greeff A, Shennan A. Validation of the Omron M7 (HEM-780-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol. Blood Press Monit 2008;13:49-54 |
39 |
Pini C, Pastori M, Baccheschi J, Omboni S, Parati G. Validation of the Artsana CSI 610 automated blood pressure monitor in adults according to the International Protocol of the European Society of Hypertension. Blood Press Monit 2007;12:179-84 |
40 |
Cotte UV, Faltenbacher VH, von Willich W, Bogner JR. Trial of validation of two devices for self-measurement of blood pressure according to the European Society of Hypertension International Protocol: the Citizen CH-432B and the Citizen CH-656C. Blood Press Monit 2008;13:55-62 |
41 |
Pini C, Natalizi A, Gerosa PF, Frigerio M, Omboni S, Parati G. Validation of the Artsana CS 410 automated blood pressure monitor in adults according to the International Protocol of the European Society of Hypertension. Blood Press Monit 2008;13:177-82 |
42 |
de Greeff A, Ahmed N, Sheppard E, Andrew H. Shennan AH. A self-measurement device suitable for hospital use: validation of the Spengler Pro M according to the International Protocol of the European Society of Hypertension. Blood Press Monit 2008;13:183-6 |
43 |
Stergiou GS, Tzamouranis D, Nasothimiou EG, Protogerou AD. Can an electronic device with a single cuff be accurate in a wide range of arm size? Validation of the Visomat Comfort 20/40 device for home blood pressure monitoring. Journal of Human Hypertension 2008;23:Advanced Online Publication |
44 |
Chen HE, Cui Y, Sheng CS, Li LH, Li Y, Wang JG. Validation of the Health & Life HL868BA blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol. Blood Press Monit 2008;13:305-8 |
45 |
de Greeff A, Arora J, Hervey S, Liu B, Shennan AH. Accuracy assessment of the Tensoval duo control according to the British and European Hypertension Societies' standards. Blood Press Monit 2008;13:111-6 |
46 |
Bonso E, Ragazzo F, Palatini P. Validation of A&D UA-85X device for blood pressure measurement. Blood Press Monit 2008;13:339-341 |
47 |
Zhang Y, Wang J, Huang QF, Sheng CS, Li Y, Wang JG. Validation of the HONSUN LD-578 blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol. Blood Press Monit 2009;14:128-31 |
48 |
Duhig KE, de Greeff A, van der Westhuizen A, Baker E, Shennan AH. Validation of the Nissei DS-400 in a low-resource setting. Blood Press Monit 2009;14:132-5 |
49 |
Li L, Zhang XY, Yan CH, Liang QX. Validation of the Oregon Scientific BPU 330 for self-monitoring of blood pressure according to the International Protocol. Vascular Health and Risk Management 2008;4(5):1121-5 |
50 |
de Greeff A, Beg Z, Gangji Z, Dorney E, Shennan AH. Accuracy of inflationary versus deflationary oscillometry in pregnancy and preeclampsia: OMRON-MIT versus OMRON-M7. Blood Press Monit 2009;14:37-40 |
51 |
Perkovic D, Guarnieri C, Palatini P. Validation of the visomat comfort eco blood pressure measuring device according to the International Protocol. Blood Press Monit 2009;14:178-80 |
52 |
Bonso E, Dorigatti F, Palatini P. Accuracy of the BP A100 blood pressure measuring device coupled with a single cuff with standard-size bladder over a wide range of arm circumferences. Blood Press Monit 2009;14:216-9 |
53 |
Narogan MV, Narogan MI, Syutkina EV. Validation of A&D UA-778 blood pressure monitor in children. Blood Press Monit 2009;14:228-31 |
54 |
Germano G, Psimenos A, Sarullo F, Venditti A, Pecchioli V, Asmar R. Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol: The Pic Indolor Personal Check, Comfort Check, My Check and Travel Check. Blood Pressure 2009;18 (Suppl 1):15-23 |
55 |
Bonso E, Dorigatti F, Saladini F, Palatini P. Validation of the Visomat Comfort form device for home blood pressure measurement according to the International Protocol. Blood Press Monit 2010;15(1):63-6 |
56 |
Wan Y, Pan F, Liu Y, Liang Y, Yang Z, Zhao H, Xu Y. Validation of the Andon KD-391 semiautomated blood pressure monitor in adults according to the International Protocol. Blood Press Monit 2009;14:181-4 |
57 |
Huang QF, Wang J, Sheng CS, Zhang NN, Li Y, Wang JG. Validation of the ANDON KD-5915 blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol. Blood Press Monit 2010;15(4):232-4 |
58 |
Coleman A, Steel S, deGreeff A, Shennan A. Validation of the Lloyds Pharmacy BP11 oscillometric blood pressure monitor according to the International Protocol of the European Society of Hypertension. Blood Press Monit 2010;15(3):163-6 |
59 |
Pan F, Wan Y, Chen W, Li N, Tong W, Xu Y. Validation of the Andon KD-575 automated blood pressure monitor according to the European Society of Hypertension International protocol in adults. Blood Press Monit 2010;15(3):167-70. |
60 |
Zeng WF, Huang QF, Sheng GS, Li Y, Wang JG. Validation of the Kingyield BP101H blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol. Blood Press Monit 2010;15(6):329-31. |
61 |
Stergiou GS, Tzamouranis D, Nasothimiou EG, Karpettas N, Protogerou A. Are there really differences between home and daytime ambulatory blood pressure? Comparison using a novel dual-mode ambulatory and home monitor. J Hum Hypertens 2010;24(3):207-12. |
62 |
Ragazzo F, Saladini F, Palatini P. Validation of the Microlife WatchBP O3 device for clinic, home, and ambulatory blood pressure measurement, according to the International Protocol. Blood Press Monit 2010;15(1):59-62. |
63 |
Masiero S, Fania C, Palatini P. Validation of the UEBE Visomat Double Comfort upper arm blood pressure monitor, in auscultation mode, for clinic use and self-measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit 2011;16(4):208-10. |
64 |
Masiero S, Fania C, Palatini P. Validation of the UEBE Visomat Double Comfort upper arm blood pressure monitor, in oscillometric mode, for clinic use and self-measurement in a general population according to the European Society of Hypertension International Protocol, revision 2010. Blood Press Monit 2011;16(5):262-4 |
65 |
Topouchian J, Agnoletti D, Blacher J, Youssef A, Ibanez I, Khabouth J, Khawaja S, Beaino L, Asmar R. Validation of four automatic devices for selfmeasurement of blood pressure according to the international protocol of the European Society of Hypertension. Vascular Health and Risk Management 2011:7;709-17 doi: 10.2147/VHRM.S27193 |
66 |
Chung Y, de Greeff A, Shennan A. Validation and Compliance of a Home Monitoring Device in Pregnancy: Microlife WatchBP Home. Hypertens Pregnancy 2009;28(3):348-59. |
67 |
Liu WJ, Li SG, Song Z, Gong W. Validation of the TRANSTEK blood pressure monitor TMB-986 for home blood pressure monitoring according to the International Protocol. Blood Press Monit 2010;15(5):278-80 doi: 10.1097/MBP.0b013e32833e43ca |
68 |
Yi J, Wan Y, Pan F, Yu X, Zhao H, Shang F, Xu Y. Validation of the Medipro MediCare 100f upper arm blood pressure monitor, for self-measurement, according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit 2011;16(4):197-9. doi: 10.1097/MBP.0b013e328346d715 |
69 |
Bonso E, Dorigatti F, Palatini P. Validation of Panasonic EW3106 and EW3109 devices for blood pressure measurement according to the International Protocol. Blood Press Monit 2010;15(1):55-8. doi: 10.1097/MBP.0b013e32833531b3 |
70 |
Chen C, Shang F, Wang J, Chen J, Ji N, Wan Y. Validation of the iHealth BP3 upper-arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit 2012;17(6):253-6. doi: 10.1097/MBP.0b013e32835b9ecb |
71 |
Takahashi H, Yoshika M, Yokoi T. Validation of home blood pressure-monitoring devices, Omron HEM-1020 and Omron i-Q132 (HEM-1010-E), according to the European Society of Hypertension International Protocol. Blood Press Monit 2011;16(4):203-7. doi: 10.1097/MBP.0b013e328348b688 |
72 |
Palatini P. Validation of the Foracare Fora D30 upper arm blood pressure monitor, for self measurement, according to the European Society of Hypertension International Protocol 2002 [Internet]. Dublin: dablEducational Trust; 2010 Apr 05. 6 p. Available from: http://www.dableducational.org/Publications/2010/ESH-IP 2002 Validation of Foracare Fora D30.pdf |
73 |
Masiero S, Fania C, Palatini P. Validation of the UEBE Visomat Double Comfort upper arm blood pressure monitor, in auscultation mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2011 Apr 27. 4 p. Available from: http://www.dableducational.org/Publications/2011/ESH-IP 2010 Validation of UEBE Visomat_Double_Comfort (Auscultatory Mode).pdf. |
74 |
Masiero S, Fania C, Palatini P. Validation of the UEBE Visomat Double Comfort upper arm blood pressure monitor, in oscillometry mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2011 Apr 27. 4 p. Available from: http://www.dableducational.org/Publications/2011/ESH-IP 2010 Validation of UEBE Visomat_Double_Comfort (Oscillometric Mode).pdf. |
75 |
Ding D. Validation of the Foracare Fora D40b upper arm blood pressure monitor, for self measurement, according to the European Society of Hypertension International Protocol Revision 2010 [Internet]. Dublin: dablEducational Trust; 2011 Jul 29. 7 p. Available from: http://www.dableducational.org/Publications/2011/ESH-IP 2010 Validation of Foracare Fora D40b.pdf |
76 |
de Greeff A, Shennan AH. Validation of the Nissei DSK-1011 upper arm blood pressure monitor, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2011 Sep 01. 9 p. Available from: http://www.dableducational.org/Publications/2011/ESH-IP 2010 Validation of Nissei DSK-1011.pdf. |
77 |
de Greeff A, Shennan AH. Validation of the Nissei DSK-1031 upper arm oscillometric blood pressure monitor intended for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2011 Oct 28. 3 p. Available from: http://www.dableducational.org/Publications/2011/ESH-IP 2010 Validation of Nissei DSK-1031.pdf. |
78 |
Takahashi H. Validation of the Omron M3W upper arm blood pressure monitor, in oscillometry mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2012 Mar 01. 4 p. Available from: http://www.dableducational.org/Publications/2012/ESH-IP 2010 Validation of Omron M3W (HEM-7202-E).pdf |
79 |
Takahashi H. Validation of the Omron M6W upper arm blood pressure monitor, in oscillometry mode, for self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2012 Mar 01. 4 p. Available from: http://www.dableducational.org/Publications/2012/ESH-IP 2010 Validation of Omron M6W (HEM-7213-E).pdf |
80 |
Asmar R. Validation of the OMRON M6 COMFORT (HEM-7221-E8) Upper Arm blood pressure monitor, for self/home measurement, according to the revised (2010) European Society of Hypertension International Protocol. [Internet]. Dublin: dablEducational Trust; 2012 Jun 06. 3 p. Available from: http://www.dableducational.org/Publications/2012/ESH-IP 2010 Validation of Omron M6 Comfort (HEM-7221-E8).pdf |
81 |
Takahashi H. Validation of the Omron i-Q142(HEM-1040-E), an upper arm blood pressure monitor, in oscillometry mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2012 Jun 07. 4 p. Available from: http://www.dableducational.org/Publications/2012/ESH-IP 2010 Validation of Omron i-Q142 (HEM-1040-E).pdf |
82 |
Ding D. Validation of the FORA P30 Plus Upper Arm Blood Pressure Monitor, for Clinic Use and Self/Home Measurement, According to the European Society of Hypertension International Protocol Revision 2010 [Internet]. Dublin: dablEducational Trust; 2012 Oct 19. 6 p. Available from: http://www.dableducational.org/Publications/2012/ESH-IP 2010 Validation of Foracare Fora P30 Plus.pdf |
83 |
Takahashi H, Yokoi T, Yoshika M. Validation of the OMRON HEM-7251G upper arm blood pressure monitor, in oscillometry mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2013 Feb 01. 4 p. Available from: http://www.dableducational.org/Publications/2013/ESH-IP 2010 Validation of Omron HEM-7251G.pdf |
84 |
Zeng WF, Kang YY, Liu M, Li Y, Wang JG. Validation of the A&D UA-1020 upper-arm blood pressure monitor for home blood pressure monitoring according to the British Hypertension Society Protocol. Blood Press Monit 2013;18(3):177-81. doi: 10.1097/MBP.0b013e32836141ca |
85 |
Shang F, Zhu Y, Zhu Z, Liu L, Wan Y. Validation of the iHealth BP5 wireless upper arm blood pressure monitor for self-measurement according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit 2013;18(5):278-81. doi: 10.1097/MBP.0b013e3283638f04 |
86 |
Palatini P. Validation of the Thermor Bios BP215 upper arm blood pressure monitor, for clinic use and self measurement, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2013 Oct 09. 11 p. Available from: http://www.dableducational.org/Publications/2013/ESH-IP 2010 Validation of Thermor Bios BD215.pdf |
87 |
de Greeff A, Shennan AH. Validation of the Tensoval Duo Control II blood pressure monitor for clinic use and self-measurement according to the British Hypertension Society protocol and the European Society of Hypertension International Protocol Revision 2010. Blood Press Monit 2013;18(3):161-6. doi: 10.1097/MBP.0b013e328360fb52 |
88 |
Takahashi H, Yokoi T, Yoshika M. Validation of the OMRON HEM-7130 upper arm blood pressure monitor, in oscillometry mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2013 Nov 14. 4 p. Available from: http://www.dableducational.org/Publications/2013/ESH-IP 2010 Validation of Omron HEM-7130.pdf |
89 |
Takahashi H, Yokoi T, Yoshika M. Validation of the OMRON HEM-7320F upper arm blood pressure monitor, in oscillometry mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2013 Nov 14. 4 p. Available from: http://www.dableducational.org/Publications/2013/ESH-IP 2010 Validation of Omron HEM-7500F.pdf |
90 |
Takahashi H, Yokoi T, Yoshika M. Validation of the OMRON HEM-7500F upper arm blood pressure monitor, in oscillometry mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2013 Nov 14. 4 p. Available from: http://www.dableducational.org/Publications/2013/ESH-IP 2010 Validation of Omron HEM-7500F.pdf |
91 |
Ding D. Validation of the Taidoc U-Right TD-3124 upper arm blood pressure monitor, for self measurement, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2013 Dec 03. 11 p. Available from: http://www.dableducational.org/Publications/2013/ESH-IP 2010 Validation of Taidoc U-Right TD-3124.pdf |
92 |
Takahashi H, Yokoi T, Yoshika M. Validation of the OMRON M6 Comfort (HEM-7321-E) upper arm blood pressure monitor, in oscillometry mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2014 Jan 22. 4 p. Available from: http://www.dableducational.org/Publications/2014/ESH-IP 2010 Validation of Omron M6 Comfort (HEM-7321-E).pdf |
93 |
Takahashi H, Yokoi T, Yoshika M. Validation of the OMRON M6 AC (HEM-7322-E) upper arm blood pressure monitor, in oscillometry mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2014 Jan 22. 4 p. Available from: http://www.dableducational.org/Publications/2014/ESH-IP 2010 Validation of Omron M6 AC (HEM-7322-E).pdf |
94 |
Ding D. Validation of the Foracare Diamond Cuff BP (Test N'Go BP) upper arm blood pressure monitor, for self measurement, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2014 May 02. 9 p. Available from: http://www.dableducational.org/Publications/2014/ESH-IP 2010 Validation of Foracare Diamond Cuff BP.pdf |
95 |
Guo WG1, Li BL, He Y, Xue YS, Wang HY, Zheng QS, Xiang DC. Validation of the Andon KD-5917 automatic upper arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit. Blood Press Monit 2014;19(4):242-5. doi: 10.1097/MBP.0000000000000048 |
96 |
Benetti E1, Fania C, Palatini P. Validation of the A&D BP UA-651 device for home blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit 2014;19(1):50-3. doi: 10.1097/MBP.0000000000000020 |
97 |
Benetti E1, Fania C, Palatini P. Validation of the A&D UA-651 upper arm blood pressure monitor, for self measurement, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2014 Jul 15. 8 p. Available from: http://www.dableducational.org/Publications/2014/ESH-IP 2010 Validation of AnD UA-651.pdf. |
98 |
Takahashi H, Yokoi T, Yoshika M. Validation of the Omron HEM-7252G-HP upper arm blood pressure monitor, in oscillometry mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2015 Jan 16. 4 p. Available from: http://www.dableducational.org/Publications/2015/ESH-IP 2010 Validation of Omron HEM-7252G-HP.pdf
|
99 |
Bing S, Hou H, Zhang W, Wan Y. Validation of the Microlife BP A3 PC upper arm blood pressure monitor, for clinical use and self-measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dabl Educational Trust; 2015 March 26. 9 p. Available from: http://www.dableducational.org/Publications/2015/ESH-IP 2010 Validation of Microlife BPA3PC.pdf |
100 |
Lu Zhang, Yuan-Yuan Kang, Wei-Fang Zeng, Yan Li and Ji-Guang Wang. Validation of the Rossmax CF175 upper arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Blood Pressure Monitoring 2015, 20:104-107, DOI: 10.1097/MBP.0000000000000089 |
101 |
Bing S., Zhang C., Wang L., Li L., Wan Y. Validation of the Pangao PG-800B11 blood pressure monitor according to the European Society of Hypertension and the British Hypertension Society protocols. Blood Press Monit.: 2014 December; 19(6):366-369 doi: 10.1097/MBP.0000000000000078. |
102 |
Lili W., Yinghui J., Chengdong W., Lei C., Dalin D., and Haiyan Z. Validation of the Andon KD-5851 upper arm blood pressure monitor, for self-measurement, according to the European Society of Hypertension International Protocol revision 2010. Blood Pressure Monitoring 2015 Aug;20(4):228-31. doi:10.1097/MBP.0000000000000114. |
103 |
Kang Y-Y, Zeng W-F, Liu M, Li Y, and Wang J-G. Validation of the AVITA BPM63S upper arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Blood Pressure Monitoring 2014 Feb;19(1):46-9. doi:10.1097/MBP.0000000000000014. |
104 |
Huang J., Li Z., Li G. and Liu Z. Validation of the Andon KD-5965 upper-arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Blood Pressure Monitoring 2015 Oct; 20(5):283-85. doi:10.1097/MBP.0000000000000129. |
105 |
Topouchian J., Agnoletti D., Blacher J., Youssef A., Chahine M.N., Ibanez I., Assemani N., Asmar R. Validation of four devices: Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670 for home blood pressure measurement according to the European Society of Hypertension International Protocol. Vascular Health and Risk Management, 16 January 2014;10:33-44. doi:10.2147/VHRM.S53968 eCollection 2014. |
106 |
Benetti E., Fania C. and Palatini P. Validation of the A&D BP UA-651 device with a wide-range cuff for home blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010 Blood Press Monit. 2015 Jun;20(3):164-7. doi: 10.1097/MBP.0000000000000100. |
107 |
Tian H., Zeng S., Zhong X., Gong W. and Liu W. Validation of Transtek blood pressure monitor TMB-1491 for self-measurement according to the European Society of Hypertension International Protocol reversion 2010. Blood Press Monit. 2015 May |
108 |
Chen, Wan; Zeng, Zhao-Lin; Bing, Sen; Li, Lin-Yi; Wang, Rui; Wan, Yi Validation of the Grandway MD2301 digital automatic blood pressure monitor according to the European Society of Hypertension International Protocol. Blood Pressure Monitoring April 18, 2016; doi:10.1097/MBP.0000000000000191 |
109 |
TOPOUCHIAN J, ZELVEYAN P, HAKOBYAN Z, MELKONYAN A, ASMAR R. Validation of the QARDIO QARDIOARM upper arm blood pressure monitor, in oscillometry mode, for self measurement in persons fulfilling the population as described in this paper, according to the European Society of Hypertension International Protocol revision 2010. [Internet]. Dublin: dablEducational Trust; 2016 Jul 8. 4 p. Available from: http://www.dableducational.org/Publications/2016/ESH-IP 2010 Validation of QARDIO QARDIOARM.pdf |
110 |
Jinhua Huanga, Yun Wanga, Zhaoying Liua and Yuling Wang Simultaneous validation of the Grandway MD2301 digital automatic blood pressure monitor by the British Hypertension Society and the Association for the Advancement of Medical Instrumentation/the International Organization for Standardization protocols. Blood Pressure Monitoring. Oct 2016. doi: 10.1097/MBP.0000000000000218 |
111 |
Zhonghua Liu, Xianyue Liu, Hengyi Zhou Validation of Artsana helpRAPID according to European Society of Hypertension International Protocol revision 2010 in adults [Internet]. Dublin: dabl Educational Trust; 2016 November 22. Available from: http://www.dableducational.org/Publications/2016/ESH-IP 2010 Validation of Artsana helpRAPID.pdf |
112 |
Hakuo Takahashi Validation of the Omron EVOLV (HEM-7600T-E) upper arm blood pressure monitor, in oscillometry mode, for self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dabl Educational Trust; 2016 November 28. Available from: http://www.dableducational.org/Publications/2016/ESH-IP 2010 Validation of HEM-7600T-E.pdf |
113 |
Liu, Zhong Hua; Liu, Xian Yue; Wu, Wen Jun Validation of Transtek LS808-B for self/home measurement according to the European Society of Hypertension International Protocol revision 2010. Blood Pressure Monitoring. 2016 Dec; 21(6):352-55. doi: 10.1097/MBP.0000000000000212 |
114 |
AZAKI A., DIAB R., HARB A., CHAHINE M.N., ASMAR R. Validation of the OMRON M6 COMFORT IT (HEM-7322U-E) Upper Arm blood pressure monitor for self/home measurement, according to the revised (2010) European Society of Hypertension International Protocol [Internet]. Dublin: dablEducational Trust; 2017 March 28. 5 p. Available from: http://www.dableducational.org/Publications/2017/ESH-IP 2010 Validation of M6 Comfort IT (in diabetic) May 2016.pdf |
115 |
Chen Q, Kang YY, Li Y, Wang JG. Validation of the BPUMP BF1112 upper-arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Blood Pressure Monitoring. 2017 Apr; 22(2):105-108. doi: 10.1097/MBP.0000000000000236 |
116 |
Lei L, Chen Y, Chen Q, Li Y, Wang JG. Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. [Internet]. Dublin: dabl Educational Trust; 2017 May 15. Available from: http://www.dableducational.org/Publications/2017/ESH-IP 2010 Validation of SEJOY BP-1307.pdf
|
117 |
Wua N, Zhangb X, Wangb W and Zhangb H. Validation of the Andon KD5031 for clinical use and self-measurement according to the European Society of Hypertension International Protocol. Blood Pressure Monitoring. 2016 Oct; 21(5):310-312(3). doi: 10.1097/MBP.0000000000000204 |
118 |
Wua N, Zhangb X, Wangb W and Zhangb H. Validation of the Andon KD595 for clinical use and self-measurement according to the European Society of Hypertension International Protocol. Blood Pressure Monitoring. 2016 Apr;21(2):124-7. doi: 10.1097/MBP.0000000000000164 |
119 |
Guo YF, Zhang XQ, Fu JL, Zhang X, Shan HZ, Shi N and Yan LQ. Validation of the G.LAB MD2680 digital automatic blood pressure monitor according to international protocols. Blood Pressure Monitoring. 2017 Aug;22(4):221-225. doi: 10.1097/MBP.0000000000000254 |
120 |
Chanine M.N., Bou Harb S., El Rahman Saad A., Sarkis P., Azaki A., Allouch A. and Asmar R. Validation of the PHILIPS DL8760, oscillometric, upper arm blood pressure monitor, for self-measurement in general population, according to the European Society of Hypertension International Protocol revision 2010. [Internet]. Dublin: dabl Educational Trust; 2017 July 31. Available from: http://www.dableducational.org/Publications/2017/ESH-IP 2010 Validation of PHILIPS DL8760.pdf
|
121 |
Kang Y.Y., Chen Q., Liu C.Y., Li Y. and Wang J.G. Validation of the AVITA BPM64 upper-arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Blood Pressure Monitoring. 2017 Oct 18. doi: 10.1097/MBP.0000000000000296
|
122 |
TOPOUCHIAN J., ZELVEIAN P. and ASMAR R. Validation of the automatic blood pressure measurements device, the OMRON M3 COMFORT (HEM -7134-E)® in Pregnancy according to the Modified European Society of Hypertension International Protocol (ESH-IP). [Internet]. Dublin: dabl Educational Trust; 2018 Jan 08. Available from: http://www.dableducational.org/Publications/2017/ESH-IP 2010 Validation of OMRON M3 COMFORT (HEM-7134-E) v2.pdf
|
123 |
TOPOUCHIAN J., ZELVEIAN P. and ASMAR R. Validation of the automatic blood pressure measurements device, the OMRON EVOLV (HEM- 7600 T-E)® in Pregnancy according to the Modified European Society of Hypertension International Protocol (ESH-IP). [Internet]. Dublin: dabl Educational Trust; 2018 Jan 08. Available from: http://www.dableducational.org/Publications/2017/ESH-IP 2010 Validation of OMRON EVOLV (HEM-7600 T-E) v2.pdf
|
124 |
Chen Q., Lei L., Li Y. and Wang JG. Validation of the YuWell YE690A upper-arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit. 2017 Oct;22(5):295-297. doi: 10.1097/MBP.0000000000000278
|
125 |
Zeng, Wei-Fang; Kang, Yuan-Yuan; Liu, Ming; Li, Yan; Wang, Ji-Guang. Validation of the A&D UA-1020 upper-arm blood pressure monitor with six different-shaped or different-sized cuffs according to the British Hypertension Society protocol. Blood Pressure Monitoring: October 2013 - Volume 18 - Issue 5 - p 272-277. doi: 10.1097/MBP.0b013e32836464a3
|
126 |
Kario K., Saito K., Sato K., Hamasaki H., Suwa H., Okura A. and Hoshide S. Validation of the A&D BP UA-1200BLE device for home blood pressure measurement according to the ISO 81060-2:2013 standard. Blood Press Monit. 2018 Dec;23(6):312-314. doi: 10.1097/MBP.0000000000000347
|
127 |
Zhang W., Lei L., Li Y., Wang JG. Validation of the HL868ED upper-arm blood pressure monitor for clinical use and self-measurement according to the European Society of Hypertension International Protocol revision 2010.
Blood Press Monit. 2019 Oct 25. doi: 10.1097/MBP.0000000000000411
|
128 |
Cheng Song, Yang Yu, Bao-Chuan Lu and Xi-Ling Yan Validation of the Globalcare GCE603 automated blood pressure monitor for self-measurement according to the European Society of Hypertension International Protocol revision 2010.
Blood Pressure Monitoring: October 2020 - Volume 25 - Issue 5 - p 291-294. doi: 10.1097/MBP.0000000000000461
|
129 |
Zhonghua Liu and Liyi Chen. Validation of Transtek TMB-1776 according to European Society of Hypertension International Protocol revision 2010 in adults.
Blood Pressure Monitoring: December 2019 - Volume 24 - Issue 6 - p 319-322. doi: 10.1097/MBP.0000000000000406
|
130 |
Validation of the blood pressure measurement device Beurer BM 28 according to the European Society of Hypertension International Protocol revision 2010. Deutsch, Cornelia; Bramlage, Carsten; Botta, Beate; Kruger, Ralf; Forstner, Klaus; Bramlage, Peter; Beime, Beate
Blood Pressure Monitoring: August 2021 - Volume 26 - Issue 4 - p 292-298 doi: 10.1097/MBP.0000000000000528
|
131 |
Validation of the JOYTECH DBP-6279B blood pressure monitor in adults and adolescents according to the AAMI/ESH/ISO universal standard(ISO 81060-2:2018 + Amd.1:2020)
Baojing Duana, Le Heb, Jianling Zhangc, Lixiu Fangd and Gang Lia.
Blood Pressure Monitoring ():10.1097/MBP.0000000000000662, July 12, 2023. | DOI: DOI: 10.1097/MBP.0000000000000662
|
164 |
Bin Peng, Jia Hu, Xinda Wang, Zijian Xie, Xiaoqin Du, Chaoya Li, Jiahui Liang. Validation of the Transtek TMB-2296-BT blood 1 pressure monitor in adults according to the ISO 81060-2:2018 + Amd.1:2020. dablEducational Trust; September 2023
Available from: http://www.dableducational.org/Publications/Validation of the Transtek TMB-2296-BT for Home BPM.
|
178 |
Validation of the JAMR F1701T (arm type) pressure monitor according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/ISO 81060-2:2018 protocol. Li, Jing; Han, Li-Hong; Feng, Ru; Wu, Li-Hao; Cai, Jie; Zeng, Tian-Ying; Liu, Hui; Zeng, Ya-Wen; Zhou, Li; Zhan, Zhong-Qun
Blood Pressure Monitoring 28(3):p 163-166, June 2023. |
DOI: 10.1097/MBP.0000000000000643
|
194 |
Validation of the DBP-1333b upper-arm blood pressure monitor according to the AAMI/ESH/ISO universal standard (ISO 81060-2:2018+Amd.1:2020) Guocun Hou, Yulu Wu, Jiaying Wang and Jia Zhi
Blood Pressure Monitoring 2024 Jan 9. |
DOI: 10.1097/MBP.0000000000000688.
|
205 |
Validation of the YuWell YE660D oscillometric upper-arm blood pressure monitor for clinic and home in general population according to the AAMI/ESH/ISO universal standard (ISO 81060-2:2018+Amd.1:2020) Hui-Jie Zhang, Yue Peng, Juan Zhang, Jin Zhang, Li-Ni Teng, Shu-Juan Zhang, De-Jun Zhou and Ming-Zhi Long. Blood Pressure Monitoring 2023, 28:276–279. |
Validation of the Transtek TMB-2266 for Clinic and Self BPM.
|
231 |
Validation of the Transtek TMB-2266 oscillometric upper-arm blood pressure monitor for clinic and self use in general population according to the AAMI/ESH/ISO universal standard (ISO 81060-2:2018+Amd.1:2020) Zi jian Xie, Li Zhu, Xiao Zheng, Chao ya Li, Jia Hu, Xinda Wang, Hua Deng, Xiaoqin Du, Jiahui Liang, Bin Peng |
Validation of the Transtek TMB-2266 for Clinic and Self BPM.
|
243 |
Validation of the FF680 upper-arm blood pressure monitor according to the AAMI/ESH/ISO universal standard (ISO 81060-2:2018) for use in clinical and self/home blood pressure monitoring among adults. Hao Chen
Validation of the Hanvon FF680 for Clinic and Self BPM.
|
Wrist Devices for Self-measurement of Blood Pressure
Recommendations for all wrist devices are based on their being positioned strictly according to manufacturer instructions.
For a complete list of devices see our Device Index and for discontinued devices see our Discontinued Devices Index.
Device |
Mode |
AAMI |
BHS |
ESH 2002 |
ESH 2010 |
ISO 81060-2:2013 |
ISO 81060-2:2018 |
Circumstance |
Recommendation Ref |
A&D UB-1100BLE |
Wrist |
|
|
|
Pass |
|
|
A&D UB-543 Equivalence |
Recommended E195 |
A&D UB-511 |
Wrist |
|
Pass |
|
|
|
|
At rest |
Questionable 7 |
A&D UB-522 |
Wrist |
|
|
|
Pass |
|
|
A&D UB-543 Equivalence |
Recommended E169 |
A&D UB-525 |
Wrist |
|
|
|
Pass |
|
|
A&D UB-543 Equivalence |
Recommended E170 |
A&D UB-533 |
Wrist |
|
|
|
Pass |
|
|
A&D UB-543 Equivalence |
Recommended E168 |
A&D UB-543 |
Wrist |
|
|
|
Pass |
|
|
At rest |
Recommended 32 |
Alvita/Kinetik Wellbeing TMB1117-S |
Wrist |
|
|
Pass |
|
|
|
Transtek TMB-988 Equivalence |
Recommended E302 |
Alvita Wrist (S150) |
Wrist |
|
|
|
Pass |
|
|
Rossmax S150 Equivalence |
Recommended E134 |
AVITA BPM15S |
Wrist |
|
|
|
Pass |
|
|
At rest |
Recommended 31 |
AVITA BPM17 |
Wrist |
|
|
|
Pass |
|
|
At rest |
Recommended 35 |
Beurer BC51 |
Wrist |
|
|
|
Pass |
|
|
AVITA BPM17 Equivalence |
Recommended E218 |
Beurer BC54 |
Wrist |
|
|
Pass |
|
|
|
Transtek TMB-988 Equivalence |
Recommended E215 |
Braun BP 2000 (BBP2000) |
Wrist |
Pass |
A/A |
Pass |
|
|
|
BP W100 Equivalence; Supine position |
Recommended E109 |
Braun BP 2200 (BBP2200) |
Wrist |
Pass |
A/A |
Pass |
|
|
|
BP W100 Equivalence; Supine position |
Recommended E110 |
Braun BP VitalScan Plus 1650 |
Wrist |
|
|
Pass |
|
|
|
At rest |
Recommended 11 |
Braun BP VitalScan Plus 1700 |
Wrist |
|
|
Pass |
|
|
|
BP1650 Equivalence |
Recommended E24 |
Braun BP VitalScan Plus 1750 |
Wrist |
|
|
Pass |
|
|
|
BP1650 Equivalence |
Recommended E25 |
Braun BPW4100 |
Wrist |
|
|
|
Pass |
|
|
BP210 Equivalence |
Recommended E117 |
Braun BPW4300 |
Wrist |
|
|
|
Pass |
|
|
BP210 Equivalence |
Recommended E118 |
Braun BPW4500 |
Wrist |
|
|
|
Pass |
|
|
AVITA BPM17 Equivalence |
Recommended E176 |
Citizen CH-656C |
Wrist |
|
|
Pass |
|
|
|
At rest |
Recommended 10 |
G.LAB MD2231 |
Wrist |
|
A/A |
|
Pass |
Pass |
|
At rest |
Recommended 34 |
G.LAB (Grandway) MD2200 |
Wrist |
|
A/A |
|
Pass |
Pass |
|
At rest |
Recommended 33 |
HARTMANN Veroval BPW22 |
Wrist |
|
|
|
Pass |
|
|
BPM15S Equivalence |
Recommended E172 |
HARTMANN Veroval BPW26 |
Wrist |
|
|
|
Pass |
|
|
AViTA BPM15S Equivalence |
Recommended E368 |
HARTMANN Veroval wrist BP monitor |
Wrist |
|
|
|
Pass |
|
|
BPM15S Equivalence |
Recommended E161 |
Health and Life
(Health & Life)
HL168JD |
Wrist |
|
|
Pass |
|
|
|
At rest |
Recommended 16 |
Honsun SCIAN LD-735 |
Wrist |
|
|
|
Pass |
|
|
General population;
Excluded individuals with arrhythmias and peripheral arterial disease |
Recommended 38 |
iHealth BP7 (KD-972) |
Wrist |
|
|
|
Pass |
|
|
At rest |
Recommended 28 |
Inventum BDP619 |
Wrist |
|
A/A |
|
Pass |
|
|
Pangao PG-800A11 Equivalence |
Recommended E156 |
JOYTECH DBP-2242 |
Wrist |
|
|
|
|
|
Pass(+1:2020) |
At rest |
Recommended 93 |
JOYTECH DBP-8278B |
Wrist |
|
|
|
|
|
Pass(+1:2020) |
At rest |
Recommended 94 |
Kingyield BP210 |
Wrist |
|
|
|
Pass |
|
|
Study (22) extended to ESH-IP 2010 requirements |
Recommended 24 |
Kinetik Wellbeing WBP3 |
Wrist |
|
|
Pass |
|
|
|
Transtek TMB-988 Equivalence |
Recommended E226 |
Medel Replay (91914REPLAY) |
Wrist |
Pass |
B/B |
Pass |
|
|
|
SAW-102 Equivalence |
Recommended E34 |
Microlife BP W100 |
Wrist |
Pass |
A/A |
Pass |
|
|
|
Supine position |
Recommended 14 |
Microlife BP W200-1 |
Wrist |
|
|
Pass |
|
|
|
MAM averaging and supine position |
Recommended 12 |
Nissei WSK-1011 |
Wrist |
|
|
|
Pass |
|
|
At rest |
Recommended 21 |
Nissei WSK-1021 |
Wrist |
|
|
|
Pass |
|
|
WSK-1011 Equivalence |
Recommended E99 |
Omron 637IT
(HEM-637-E) |
Wrist |
Pass |
B/A |
Fail |
|
|
|
Abstract; BHS 'pass' incorrectly reported |
Questionable 3 |
|
|
|
|
|
|
AD Hoc protocol,
+ position sensor, std. protocol projections inconclusive |
Questionable 1 |
|
|
Pass |
|
|
|
Adults |
Recommended 4 |
|
|
Pass |
|
|
|
Obese Adults |
Recommended 4 |
|
|
Pass |
|
|
|
Elderly |
Recommended 5 |
Omron NightView
(HEM-9601T) |
Wrist |
|
|
|
|
Pass |
|
At rest
Additional comparison study with upper-arm monitor |
Recommended 39 40 |
Omron R3 Intellisense
(HEM-6021-E) |
Wrist |
|
|
Pass |
|
|
|
R3-I Plus (HEM-6022-E)
Equivalence |
Recommended E47 |
Omron R3-I Plus
(HEM-6022-E) |
Wrist |
|
|
Pass |
|
|
|
At rest |
Recommended 18 |
Omron R5-I
(HEM-630-E) |
Wrist |
|
|
Pass |
|
|
|
Elderly Population |
Recommended 9 |
Omron R6
(HEM-6000-E) |
Wrist |
|
|
Pass |
|
|
|
R7 (HEM-637-IT) Equivalence |
Recommended E22 |
Omron R6
(HEM-6000-E7) |
Wrist |
|
|
Pass |
|
|
|
R7 (HEM-637-IT) Equivalence
New sensor |
Recommended E39 |
Omron R6
(HEM-6052-E) |
Wrist |
|
|
Pass |
|
|
|
R7 (HEM-637-IT) Equivalence |
Recommended E62 |
Omron R6
(HEM-6052-E7) |
Wrist |
|
|
Pass |
|
|
|
R7 (HEM-637-IT) Equivalence
New sensor |
Recommended E63 |
Omron R7
(HEM-637-E7) |
Wrist |
|
|
Pass |
|
|
|
R7 (HEM-637-IT) Equivalence
New sensor |
Recommended E61 |
Omron R7
(HEM-637-IT) |
Wrist |
|
|
Pass |
|
|
|
At rest |
Recommended 6 |
Omron RS1
(HEM-6120-E) |
Wrist |
|
|
|
Pass |
|
|
RS3
(HEM-6130-E) Equivalence |
Recommended E100 |
Omron RS1
(HEM-6160-E) |
Wrist |
Pass |
|
|
Pass |
|
|
RS4
(HEM-6181-E) Equivalence |
Recommended E183 |
Omron RS2
(HEM-6121-E) |
Wrist |
|
|
|
Pass |
|
|
RS3
(HEM-6130-E) Equivalence |
Recommended E101 |
Omron RS2
(HEM-6161-E) |
Wrist |
Pass |
|
|
Pass |
|
|
RS4
(HEM-6181-E) Equivalence |
Recommended E184 |
Omron RS3
(HEM-6130-E) |
Wrist |
|
|
|
Pass |
|
|
At rest |
Recommended 26 |
Omron RS3 Intelli IT
(HEM-6161T-E) |
Wrist |
Pass |
|
|
Pass |
|
|
RS4
(HEM-6181-E) Equivalence |
Recommended E185 |
Omron RS4
(HEM-6181/HEM-6181-E) |
Wrist |
|
|
|
Pass |
Pass |
|
|
Recommended 37 |
Omron RS6
(HEM-6221-E) |
Wrist |
|
|
|
Pass |
|
|
At rest |
Recommended 27 |
Omron RS7 Intelli IT
(HEM-6232T/HEM-6232T-E) |
Wrist |
|
|
|
Pass |
|
|
At rest
Patients with arm circumference ≥ 32 cm |
Recommended 36 |
|
|
|
Pass |
Pass |
|
|
Recommended 37 |
Omron RS8
(HEM-6310F-E) |
Wrist |
|
|
|
Pass |
|
|
At rest |
Recommended 29 |
Omron RX3
(HEM-640-E) |
Wrist |
|
|
Pass |
|
|
|
Abstract only |
Questionable 2 |
Pangao PG-800A11 |
Wrist |
|
A/A |
|
Pass |
|
|
General Population |
Recommended 30 |
Pic Solution (Pikdare, formely Artsana)
Pic Indolor (Pikdare, formely Artsana)
liteRAPID WRIST (REF 02022534000000) |
Wrist |
|
|
Pass |
|
|
|
Transtek TMB-988 Equivalence |
Recommended E224 |
Pic Solution (Pikdare, formely Artsana)
Pic Indolor (Pikdare, formely Artsana)
mobileRAPID WRIST |
Wrist |
|
|
|
Pass |
|
|
AViTA BPM17 Equivalence |
Recommended E178 |
Pic Solution (Artsana)
Pic Indolor (Artsana)
Self Check |
Wrist |
|
|
Pass |
|
|
|
Travel Check Equivalence |
Recommended E32 |
Pic Solution (Pikdare, formely Artsana)
Pic Indolor (Pikdare, formely Artsana)
smartRAPID |
Wrist |
|
|
Pass |
|
|
|
Transtek TMB-988 Equivalence |
Recommended E142 (v. 2019) |
Pic Solution (Artsana)
Pic Indolor (Artsana)
Travel Check |
Wrist |
|
|
Pass |
|
|
|
At rest |
Recommended 17 |
Rossmax S150 |
Wrist |
|
|
|
Pass |
|
|
Study (23) protocol violations corrected |
Recommended 25 |
Samsung SBM-100A |
Wrist |
|
|
Fail |
|
|
|
At rest |
Not Recommended 8 |
Sensacare SAW-102 (Standard) |
Wrist |
Pass |
B/B |
Pass |
|
|
|
At rest |
Recommended 13 |
Transtek TMB-988 |
Wrist |
|
|
Pass |
|
|
|
At rest |
Recommended 20 |
UEBE Visomat Handy |
Wrist |
|
|
Pass |
|
|
|
At rest |
Recommended 19 |
Visicor HM40 |
Wrist |
|
|
Pass |
|
|
|
Supine position |
Recommended 15 |
References for Validation of Wrist Devices for Self-measurement of Blood Pressure
1 |
Yarrows SA. Comparison of the Omron HEM-637 wrist monitor to the auscultation method with the wrist position sensor on or disabled. Am J Hypertens 2004;17(1):54-8. Erratum in. Am J Hypertens 2004;17(4):383. |
2 |
El-Assaad MA, Orobinskaya LV, Topouchian JA, Asmar RG. Validation of the Omron® RX-3 Blood Pressure Measuring Device at the wrist level according to the International Protocol of the European Society of Hypertension [abstract]. Am J Hypertens 2005;18(S4):29A Abstract P58 |
3 |
Altunkan Ş, Iliman N, Altunkan E Accuracy of the new wrist blood pressure monitor (Omron 637 IT), for blood pressure measurement [abstract]. J Hypertens 2003;21(Suppl 4):S22 Abstract P1.6 |
4 |
Altunkan Ş, Öztas K, Altunkan E. Validation of the Omron 637IT wrist blood pressure measuring device with a position sensor according to the International Protocol in adults and obese adults. Blood Press Monit 2006;11:79-85 |
5 |
Altunkan Ş, Altunkan E. Validation of the Omron 637IT wrist blood pressure measuring device with a position sensor according to the International Protocol in the elderly. Blood Press Monit 2006;11:97-102 |
6 |
Topouchian JA, El Assaad MA, Orobinskaia LV, El Feghali RN, Asmar RG. Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT). Blood Press Monit 2006;11:165-71 |
7 |
Angeli F, Sardone M, Angeli E, Repaci S, Gattobigio R, Verdecchia P. Validation of the A&D wrist-cuff UB-511 (UB-512) device for self-measurement of blood pressure. Blood Press Monit 2006;11:349-54 |
8 |
Altunkan Ş, Iliman N, Altunkan E. Validation of the Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure measuring devices in adults according to the International Protocol. Blood Press Monit 2007;12:119-25 |
9 |
Omboni S, Riva I, Giglio I, Caldara G, Groppelli A, Parati G. Validation of the Omron M5-I, R5-I and HEM-907 automated blood pressure monitors in elderly individuals according to the International Protocol of the European Society of Hypertension. Blood Press Monit 2007;12:233-42 |
10 |
Cotte UV, Faltenbacher VH, von Willich W, Bogner JR. Trial of validation of two devices for self-measurement of blood pressure according to the European Society of Hypertension International Protocol: the Citizen CH-432B and the Citizen CH-656C. Blood Press Monit 2008;13:55-62 |
11 |
Nolly H, Osso P, Nolly A, Silva M, Nolly M, Romero M. Self-measurement of blood pressure: validation of the Braun BP VitalScan Plus 1650 monitor according to the European Society of Hypertension International Protocol. Blood Press Monit 2008;13:101-6 |
12 |
Palatini P, Dorigatti F, Bonso E, Ragazzo F. Validation of the Microlife BP W200-1 wrist device for blood pressure measurement. Blood Press Monit 2008;13:295-8 |
13 |
Zaetta V, Longo D, Perkovic D, Perfetti P, Gabrieli A, Prattic� F, Winnicki M. Validation of the SAW-102 wrist home blood pressure monitor according to the protocols of the British Hypertension Society, the Association for the Advancement of Medical Instrumentation, and the European Society of Hypertension. Blood Press Monit 2009;14:32-6 |
14 |
Palatini P, Dorigatti F, Bonso E, Ragazzo F. Validation of Microlife BP W100 wrist device assessed according to the European Society of Hypertension and the British Hypertension Society protocols. Blood Press Monit 2009;14:41-4 |
15 |
Dorigatti F, Bonso E, Saladini F, Palatini P. Validation of the visocor HM40 wrist blood pressure measuring device according to the International Protocol. Blood Press Monit 2009;14:83-6 |
16 |
Sheng CS, Huang QF, Wang J, Zhang Y, Li Y, Wang JG. Validation of the Health & Life HL168JD wrist blood pressure monitor for home blood pressure monitoring according to the International Protocol. Blood Press Monit 2009;14:125-7 |
17 |
Germano G, Psimenos A, Sarullo F, Venditti A, Pecchioli V, Asmar R. Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol: The Pic Indolor Personal Check, Comfort Check, My Check and Travel Check. Blood Pressure 2009;18 (Suppl 1):15-23 |
18 |
Asmar R, Khabouth J, Topouchian J, El Feghali R, Mattar J. Validation of three automatic devices for self-measurement of blood pressure according to the International Protocol: The Omron M3 Intellisense (HEM-7051-E), the Omron M2 Compact (HEM 7102-E), and the Omron R3-I Plus (HEM 6022-E). Blood Press Monit 2010;15(1):49-54 doi: 10.1097/MBP.0b013e3283354b11 |
19 |
Saladini F, Benetti E, Palatini P. Accuracy of the visomat handy wrist blood pressure measuring device according to the International Protocol. Blood Press Monit 2010;15(5):281-4 doi: 10.1097/MBP.0b013e32833e50f2 |
20 |
Tian HY, Liu WJ, Li SG, Song Z, Gong W. Validation of the Transtek TMB-988 wrist blood pressure monitor for home blood pressure monitoring according to the International Protocol. Blood Press Monit 2010;15(6):326-8 doi: 10.1097/MBP.0b013e32833f56fb |
21 |
de Greeff A, Shennan AH. Validation of the Nissei WSK-1011 wrist blood pressure monitor, for self measurement, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2011 Sep 09 9 p. Available from: http://www.dableducational.org/Publications/2011/ESH-IP 2010 Validation of Nissei WSK-1011.pdf. |
22 |
Zeng WF, Huang QF, Sheng CS, Li Y, Wang JG. Validation of the Kingyield BP210 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension-International Protocol. Blood Press Monit. Blood Press Monit 2012;17(5):42-4 doi: 10.1097/MBP.0b013e32834f8301 |
23 |
Zeng WF, Huang QF, Sheng CS, Li Y, Wang JG. Validation of the Rossmax S150 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit 2012;17(5):204-6 doi: 10.1097/MBP.0b013e328359c426 |
24 |
Zeng WF, Huang QF, Sheng CS, Li Y, Wang JG. Validation of the Kingyield BP210 wrist blood pressure monitor, for self measurement, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2013 Jan 07. 7 p. Available from: http://www.dableducational.org/Publications/2013/ESH-IP 2010 Validation of Kingyield BP210.pdf |
25 |
Zeng WF, Huang QF, Sheng CS, Li Y, Wang JG. Validation of the Rossmax S150 wrist blood pressure monitor, for self measurement, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2013 Apr 08. 8 p. Available from: http://www.dableducational.org/Publications/2013/ESH-IP 2010 Validation of Rossmax S150.pdf |
26 |
Takahashi H, Yokoi T, Yoshika M. Validation of the OMRON RS3 (HEM-6130-E) wrist blood pressure monitor, in oscillometry mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2013 Feb 01. 4 p. Available from: http://www.dableducational.org/Publications/2013/ESH-IP 2010 Validation of Omron RS3 (HEM-6130-E).pdf |
27 |
Takahashi H, Yokoi T, Yoshika M. Validation of the OMRON RS6 (HEM-6221-E) wrist blood pressure monitor, in oscillometry mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2013 Feb 01.4 p. Available from: http://www.dableducational.org/Publications/2013/ESH-IP 2010 Validation of Omron RS6 (HEM-6221-E).pdf |
28 |
Wang Q, Zhao H, Chen W, Li N, Wan Y. Validation of the iHealth BP7 wrist blood pressure monitor, for self-measurement, according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit. 2014;19(1):54-7. doi: 10.1097/MBP.0000000000000017. |
29 |
Takahashi H, Yokoi T, Yoshika M. Validation of the OMRON RS8 (HEM-6310F-E) wrist blood pressure monitor, in oscillometry mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2014 Feb 27.4 p. Available from: http://www.dableducational.org/Publications/2014/ESH-IP 2010 Validation of Omron RS8 (HEM-6310F-E).pdf |
30 |
Xie P., Wang Y., Xu X., Huang, F., Pan J. Validation of the Pangao PG-800A11 wrist device assessed according to the European Society of Hypertension and the British Hypertension Society protocols. Blood Press Monit. 2015 April; 20(2):108-111 doi: 10.1097/MBP.0000000000000095. |
31 |
Kang Y-Y, Zeng W-F, Zhang L, Li Y, and Wang J-G. Validation of the AVITA BPM15S wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit. 2014 Jun; 19(3):183-6. doi: 10.1097/MBP.0000000000000041. |
32 |
Fania C., Benetti E. and Palatini P. Validation of the A&D BP UB-543 wrist device for home blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010.Blood Press Monit. 2015 Aug; 20(4):237-40. doi: 10.1097/MBP.0000000000000117. |
33 |
Liu ZY, Zhang QH, Ye XL, Liu DP, Cheng K, Zhang CH, Wan Y. Validation of the G.LAB MD2200 wrist blood pressure monitor according to the European Society of Hypertension, the British Hypertension Society, and the International Organization for Standardization Protocols. Blood Press Monit. 2017 Feb 7. doi: 10.1097/MBP.0000000000000234. |
34 |
Jiao Y1, Guan Q, Wu L, Wang C and Cao L. Validation of the G.LAB MD2231 digital automatic wrist blood pressure monitor according to multiple protocols. Blood Press Monit. 2017 Aug;22(4):226-229. doi: 10.1097/MBP.0000000000000256 |
35 |
Kang YY, Chen Q, Liu CY, Li Y, Wang JG. Validation of the AVITA BPM17 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit. 2017 Aug;22(4):230-233. doi: 10.1097/MBP.0000000000000261 |
36 |
Roland A., Topouchian J., Tasic N., Dragisic D., Tasic D. and Filipovic M. Validation of the OMRON RS7® (HEM-6232T-E) in Patients with arm circumference ≥ 32 cm according to the European Society of Hypertension International Protocol (ESH-IP). V 1.2. Foundation - Medical Research Institutes, Place de Saint-Gervais 1, Boite postale 2049, 1211 Gen�ve 1, Suisse. http://www.dableducational.org/pdfs/FTS/Validation of Omron RS7 in Obese Patients 06 Sept 2018.pdf |
37 |
Saito K., Hishiki Y., Takahashi H. Validation of two automatic devices, Omron HEM-6232T and HEM-6181, for self-measurement of blood pressure at the wrist according to the ANSI/AAMI/ISO 81060-2:2013 protocol and the European Society of Hypertension International Protocol revision 2010. Vascular Health and Risk Management. 2019 March 6; Volume 2019:15 Pages 47-55. doi:10.2147/VHRM.S188089 |
38 |
Kang, Yuan-Yuan; Chen, Qi; Li, Yan; Wang, Ji-Guang. Validation of the SCIAN LD-735 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Blood Pressure Monitoring: August 2016 - Volume 21 - Issue 4 - p 255-258 http://www.dableducational.org/pdfs/FTS/FT1403 Honsun SCIAN LD-735.pdf doi:10.1097/MBP.0000000000000192 |
39 |
Kuwabara M, Harada K, Hishiki Y, Ohkubo T, Kario K, Imai Y. Validation of a wrist-type home nocturnal blood pressure monitor in the sitting and supine position according to the ANSI/AAMI/ISO81060-2:2013 guidelines: Omron HEM-9601T. J Clin Hypertens. 2020;00:1-9. doi:10.1111/jch.13864 |
40 |
Kario K, Tomitani N, Iwashita C, Shiga T, Kanegae H. The first simultaneous self-monitoring comparison of a supine algorithm-equipped wrist nocturnal home BP monitoring device with an upperarm device. J Clin Hypertens. 2021;00:1-9. doi:10.1111/jch.14179 |
93 |
Validation of the validity of the new wrist-type fully automatic blood pressure monitor: DBP-2242 according to ISO 81060-2:2018+AMD 1:2020 Yuehong Jina, Xiaoming Beib, and Ying Jina. Blood Press Monit. 2024 Jul 16. doi: 10.1097/MBP.0000000000000716.
Available from
|
94 |
Validation of the DBP-8278B Wrist-type Fully Automatic Blood Pressure Monitor in the general population according to the ISO 81060-2:2018/AMD 1:2020 protocol. Qi Zhou. Blood Pressure Monitoring:10.1097/MBP.0000000000000717.
Available from
|
Devices for Measuring Blood Pressure in the Community
For a complete list of devices see our Device Index and for discontinued devices see our Discontinued Devices Index.
Device |
Mode |
AAMI |
BHS |
ESH 2002 |
ESH 2010 |
ISO 81060-2:2013 |
ISO 81060-2:2018 |
Circumstance |
Recommendation Ref |
A&D TM-2655 |
Osc |
Pass |
A/A |
|
|
|
|
Methods poorly explained |
Questionable 3 |
Keito K6 |
Wrist |
|
|
|
|
|
|
Standing; Fail Ad Hoc protocol |
Questionable 2 |
Pharma-Smart PS-2000 |
Cuff |
Pass |
A/A |
|
|
|
|
At rest; BHS protocol violations |
Questionable 1 |
References for Validation of Devices for Measuring Blood Pressure in the Community
1 |
Alpert BS. Validation of the Pharma-Smart PS-2000 public use blood pressure monitor Blood Press Monit 2004;9:19-23 |
2 |
Staal EM, Nygard OK, Omvik P, Gerdts E. Blood pressure measurements by the Keito machine. Evaluation versus office blood pressure by physicians Blood Press Monit 2004;9:167-72 |
3 |
Kobalava ZD, Kotovskaya YV, Babaeva LA, Moiseev VS. Validation of the TM-2655 oscillometric device for blood pressure measurement Blood Press Monit 2006;11:87-90 doi: 10.1097/01.mbp.0000200484.49540.12. |
|