Accuracy Criteria
Importance of Validation
There is a large market for blood pressure measuring
devices, not only in clinical medicine, but also with the public
where the demand for self-measurement of blood pressure is growing
rapidly. For the consumer, whether medical or lay, device accuracy
should be of prime importance in selecting a blood pressure measuring
device. However, the majority of devices available have not been
evaluated independently for accuracy according to the three most widely
used protocols the British Hypertension Society (BHS) Protocol and
the Association for the Advancement of Medical Instrumentation (AAMI)
Standard, and more recently the International Protocol of the European
Society of Hypertension (IP).
The purpose of the
dabl®Educational
Trust
website is to produce regular reviews of blood pressure
measuring devices to guide the would-be-purchaser through a complex
market. As the majority of blood pressure measuring devices has
not been independently validated, the devices listed on the website
are representative, therefore, of only a fraction of the many
devices available. It is hoped that manufacturers will wish to
list devices for which validation is planned and that in time
the devices listed will be representative of the market.
Declaration of Blood Pressure Measuring Device Equivalence
If a manufacturer has two or more devices covered by the same validation study a Declaration of Blood Pressure Measuring Device Equivalence must be completed by a Director of the manufacturing company. This signed and sealed document will be listed as a reference in the validation tables. Click here for a copy.
Criteria for device recommendation
The following criteria have been used to designate
devices according to accuracy:
'Recommended'- a device that
achieves a BHS grade B or A for both systolic and diastolic blood
pressures, and/or passes the ESH International Protocol criteria
for both systolic and diastolic pressures.
'Not recommended'- devices that cannot be
recommended fall into two categories: firstly, there are devices
which clearly fail to achieve the above accuracy criteria and are
categorically not recommended for clinical use when accurate measurement
is required, and secondly, there are devices for which the evidence
needed to make a recommendation is uncertain or unsatisfactory,
and such devices are designated as 'questionable'. Devices in this
category cannot, however, be recommended for clinical use when accurate
measurement is required on the basis of the evidence available,
though this designation might later change if further evidence allowed
for the removal of the 'questionable' designation. These categories
are listed in the tables as:
-
Not recommended - on evidence - a device that
achieves a BHS grade C or D for either systolic or diastolic pressure,
and/or fails the ESH International Protocol criteria for either
systolic or diastolic pressures.
-
Not recommended - questionable evidence - a device
for which there is doubt about the strength of evidence, as may
occur in the following circumstances: (i) when the validation
results are presented in abstract form only without sufficient
detail being available to appraise the methodology, it may be
best to withhold an opinion until the full results have been published,
or at least provided to a would-be purchaser by the manufacturer;
(ii) when the conditions of the protocols have not been fully
adhered to (listed as 'protocol violation'); (iii) when a device
fulfils the criteria of the AAMI protocol alone for both systolic
and diastolic pressures; (iv) when a device fulfils the AAMI criteria
for intra-arterial validation, it may be best to await a validation
against indirect blood pressure measurement before recommending
the device for general clinical use (the BHS and ESH International
Protocol do not advocate validation using direct intra-arterial
measurement).
Approval by CE, ISO or any other body are not criteria considered for device recommendation.
Protocol Adherence
Adherence to protocols is critical. The criteria within modern protocols are derived from a wealth of data and have been rigorously tested. Deviations from this weaken the validity of the results. In particular, recruitment of extra subjects tends to be common. This is usually due to extra subjects being tested while striving to attain quotas for the highest and lowest ranges. However, their inclusion affects the blood pressure distribution upon which the outcome criteria are based. The correct approach is to use only the exact numbers suggested in the order they are recruited. Independence of the validation is paramount and any assistance by the manufacturers in the validation procedure automatically compromises the results.
There is no valid reason for protocol deviation and, scientifically, little tolerance can be made for this. The tables list the outcomes as stated in the papers (a pass means a pass for both SBP and DBP ) and is followed by columns leading to whether or not the device can be recommended for blood pressure measurement. Quite simply, if a device passes a correclty performed published validation, that assesses individual measurements, then it is recommended.
Accuracy criteria tables
The following tables summarise the accuracy criteria for the International Protocol of the European Society for Hypertension (IP), the British Hypertension Society (BHS) Revised Protocol of 1993 and their original protocol of 1990, and the Association for the Advancement of Medical Instrumentation (AAMI)
Standard.
These tables are for quick reference purposes only and should not be used without knowledge of the full protocols.
ESH International Protocol 2010 |
Sequential Measurements |
IP Grading Part 1 - Minimum Requirements |
N |
≤ 5 mmHg |
≤ 10 mmHg |
≤ 15 mmHg |
Two of |
73 |
87 |
96 |
All of |
65 |
81 |
93 |
IP Grading Part 2 |
N |
2/3 ≤ 5 mmHg |
|
0/3 ≤ 5 mmHg |
At least |
24 |
|
|
At most |
|
|
3 |
Recruitment Ranges |
SBP |
DBP |
mmHg |
Subjects |
mmHg |
Subjects |
90*..129 |
10 - 12 |
40*..79 |
10 - 12 |
130..160 |
10 - 12 |
80..100 |
10 - 12 |
161..180* |
10 - 12 |
101..130* |
10 - 12 |
* |
Altogether, up to 4 recruitment pressures are permitted to be outside these limits. |
Test Measurement Distribution |
SBP |
DBP |
mmHg |
Measurements |
mmHg |
Measurements |
≤ 100 |
≥ 1 |
≤ 50 |
≥ 1 |
< 130 |
22 - 44* |
< 80 |
22 - 44* |
130..160 |
22 - 44* |
80..100 |
22 - 44* |
> 160 |
22 - 44* |
> 100 |
22 - 44* |
≥ 170 |
≥ 1 |
≥ 120 |
≥ 1 |
* |
The difference between the maximum and minimum of these 3 counts for SBP and, separately, for DBP, must be less than or equal to 19. |
ESH International Protocol 2002 |
Sequential Measurements |
IP Grading Phase 1 - Minimum Requirements |
N |
≤ 5 mmHg |
≤ 10 mmHg |
≤ 15 mmHg |
At least one of |
25 |
35 |
40 |
IP Grading Phase 2.1 - Minimum Requirements |
N |
≤ 5 mmHg |
≤ 10 mmHg |
≤ 15 mmHg |
Two of |
65 |
80 |
95 |
All of |
60 |
75 |
90 |
IP Grading Phase 2.2 |
N |
2/3 ≤ 5 mmHg |
|
0/3 ≤ 5 mmHg |
At least |
22 |
|
|
At most |
|
|
3 |
Ranges |
SBP |
DBP |
mmHg |
Subjects Phase 1,2 |
mmHg |
Subjects Phase 1,2 |
90..129 |
5, 11 |
40..79 |
5, 11 |
130..160 |
5, 11 |
80..100 |
5, 11 |
161..180 |
5, 11 |
101..130 |
5, 11 |
BHS Revised Protocol (1993) |
Sequential Measurements |
Grading - Minimum Requirements |
Grade |
≤ 5 mmHg |
≤ 10 mmHg |
≤ 15 mmHg |
A |
60 |
85 |
95 |
B |
50 |
75 |
90 |
C |
40 |
65 |
85 |
D |
0 |
0 |
0 |
Ranges |
SBP |
DBP |
mmHg |
Minumum |
mmHg |
Minumum |
< 90 |
8 |
< 60 |
8 |
90..129 |
20 |
60..79 |
20 |
130..160 |
20 |
80..100 |
20 |
161..180 |
20 |
101..110 |
20 |
> 180 |
8 |
> 110 |
8 |
BHS Protocol (Original 1990) |
Simultaneous or Bracketed Sequential Measurements |
Grading - Minimum Requirements |
Grade |
≤ 5 mmHg |
≤ 10 mmHg |
≤ 15 mmHg |
A |
80 |
90 |
95 |
B |
65 |
85 |
95 |
C |
45 |
75 |
90 |
D |
0 |
0 |
0 |
Ranges |
SBP |
DBP |
mmHg |
Minumum |
mmHg |
Minumum |
100..140 |
15% of BPs |
60..80 |
20% of BPs |
140..180 |
15% of BPs |
80..100 |
20% of BPs |
180..220 |
15% of BPs |
100..120 |
20% of BPs |
220..240 |
15% of BPs |
|
|
AAMI Protocol |
Simultaneous Measurements |
Criterion 1 Grading (at least 255 pairs of individual measurements) |
Grade |
Mean |
|
SD |
Pass |
≤ 5mmHg |
and |
≤ 8mmHg |
Fail |
> 5mmHg |
or |
> 8mmHg |
Criterion 1 Grading (85 subject mean pressures)
(Linear interpolation of SD limits or Means between those listed)
|
Grade |
Mean (mmHg) |
|
SD (mmHg) |
Pass |
5 |
and |
≤ 4.81 |
or |
4.5 |
and |
≤ 5.24 |
or |
4 |
and |
≤ 5.64 |
or |
3.5 |
and |
≤ 5.97 |
or |
3 |
and |
≤ 6.25 |
or |
2.5 |
and |
≤ 6.47 |
or |
2 |
and |
≤ 6.65 |
or |
1.5 |
and |
≤ 6.78 |
or |
1 |
and |
≤ 6.87 |
or |
0.5 |
and |
≤ 6.93 |
or |
0 |
and |
≤ 6.97 |
Fail |
> 5 |
or |
> limit |
Ranges |
SBP |
DBP |
mmHg |
Requirement |
mmHg |
Requirement |
≤ 100 |
≥ 5% |
≤ 60 |
≥ 5% |
≥ 140 |
≥ 20% |
≥ 85 |
≥ 20% |
≥ 160 |
≥ 5% |
≥ 100 |
≥ 5% |
|