Classification of sphygmomanometers/blood pressure measuring devices A table of blood pressure measurement devices as recommended by the dabl Educational Trust Advisory Board Table of devices currently available on the market Table of devices that have been discontinued by their manufacturers
Table of currently available blood pressure monitors Devices used in a clinical setting Table of devices used for self/home blood pressure measurement Table of devices used for Ambulatory Blood Pressure Measurement

Sphygmomanometers for Ambulatory Blood Pressure Measurement

The following table is a list of all ABPM devices currently available. Discontinued devices are shown on a separate table. A complete list of all devices is available on our Device Index.

Devices for ABPM

  FTS: Study evaluated prior to publication using the dabl®Educational Fast Track validation Service
Device Mode AAMI BHS ESH 2002 ESH 2010 ISO 81060-2:2013 ISO 81060-2:2018 Circumstance Recommendation Ref
A&D TM-2430 Osc Pass A/A         At rest Recommended 6
  A/A         Children and Adolescents Recommended 27
A&D TM-2440 Osc         Pass   A&D TM-2441 Equivalence Recommended E196
A&D TM-2441 Osc         Pass   At rest Recommended 42
Advanced Biosensor AM5600 Aus Pass           In children; AAMI Method 1 only; BHS grading not appropriate; contradictions Questionable 28
Beneware ABP-021 Osc       Pass     At rest Recommended 176
Cardiette BP One Osc   A/A         At rest Recommended 25
Custo-med
custo screen 300
Osc       Pass     At rest Recommended 36
Custo-med
custo screen 400
Osc       Pass     At rest Recommended 36
Custo-med
custo screen pediatric
Osc       Pass     Children aged 3 to 12 years Recommended 40
EnviteC PhysioQuant Osc     Pass       Results and explanations incomplete, No Plots Questionable 29
HealthStats BPro Wrist Pass   Pass       Study details omitted Not Recommended 26
Hingmed WBP-02A Osc       Pass     At rest Recommended 41
IEM
Mobil-O-Graph NG
Osc   A/A         At rest Recommended 33
Meditech ABPM-04 Osc Pass B/B         At rest Recommended 1
Meditech ABPM-05 Osc Pass B/B         ABPM-04 Equivalence Recommended E14
Meditech card(X)plore Osc Pass B/B         ABPM-04 Equivalence Recommended E36
Microlife WatchBP O3 (3MZ0) Osc     Pass       At rest Recommended 31
Microlife WatchBP O3 (3MZ1) Osc     Pass       At rest Recommended 32
Nissei DS-250 Osc     Pass       Abstract; 'Passes' incorrectly reported Questionable 17
Aus     Pass      
PAR Medizintechnik PHYSIO-PORT Osc       Pass     PAR Medizintechnik Tonoport VI Equivalence Recommended E162
PAR Medizintechnik Tonoport VI Osc       Pass     At rest Recommended 39
Schiller BR-102 Plus Osc     Pass       At rest Recommended 30
Aus     Pass       Recommended 30
Seinex SE-25M Osc Pass A/B         Analysis violations (AAMI & BHS); BHS 'pass' incorrectly reported Questionable 16
Somnomedics Somnotouch-NIBP Wrist PTT       Pass     At rest Recommended 38
Spacelabs 90207 Osc Pass B/B         At rest Recommended 2
Pass B/B         In pregnancy Recommended 3
Pass A/B         Elderly standing & sitting SBP < 161 mmHg Recommended 4
Pass A/C         In pregnancy Not Recommended 7
Pass B/C         In pregnancy Not Recommended 8
Fail D/D         In pre-eclampsia Not Recommended 9
Pass C/C         In pre-eclampsia Not Recommended 10
Pass/ Fail C/D         In children Not Recommended 11
Pass A/D         Elderly supine over all pressures Not Recommended 12
Pass C/B         During haemodialysis Not Recommended 13
Spacelabs 90217 Osc Pass A/A         At rest Recommended 5
Spacelabs 90227 OnTrak Osc   A/A   Pass     At rest Recommended 37
Suntech Medical AGILIS Osc     Pass       At rest Recommended 21
Suntech Medical OSCAR 2 Osc     Pass       At rest Recommended 18
  A/A         At rest Recommended 24
TensioMed Tensioday Osc Pass A/A         At rest; grades of pressure; varying positions Recommended 15
Tiba Medical Ambulo 2400 Osc Pass A/A         Protocol violations Questionable 35
Tonoport V Osc     Fail       At rest Not Recommended 20
    Pass       At rest Questionable  22

References for Validation of Devices for ABPM

1 Barna I, Keszei A, Dunai A. Evaluation of Meditech ABPM-04 ambulatory blood pressure measuring device according to the British Hypertension Society protocol. Blood Press Monit. 1998;3:363-368
2 O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the Spacelabs 90207 determined by to the British Hypertension Society Protocol. J Hypertens 1991;9(suppl 5):S25-S31
3 Shennan AH, Kissane J, de Swiet M. Validation of the Spacelabs 90207 ambulatory blood pressure monitoring in pregnancy. Br J Obstet & Gynae 1993;100:904-8
4 Iqbal P, Fotherby MD, Potter JF. Validation of the Spacelabs 90207 automatic non-invasive blood pressure monitor in elderly subjects. Blood Press Monit 1996;1:367-73
5 Baumgart P, Kamp J. Accuracy of the Spacelabs Medical 90217 ambulatory blood pressure monitor. Blood Press Monit 1998;3:303-307
6 Palatini P, Frigo G, Bertolo O, Roman E, Da Corta R, Winnicki M. Validation of the A&D TM-2430 device for ambulatory blood pressure monitoring and evaluation of performance according to subjects’ characteristics. Blood Press Monit 1998;3:255-260
7 O'Brien E, Mee F, Atkins N, Halligan A, O'Malley K. Accuracy of the Spacelabs 90207 ambulatory blood pressure measuring system in normotensive pregnant women determined by the British Hypertension Society Protocol. J Hypertens 1993;11(suppl 5):S282-S283
8 Franx A, van der Post JAM, van Montfrans GA, Bruinse HW. Comparison of an auscultatory versus an oscillometric ambulatory blood pressure monitor in normotensive, hypertensive and preeclamptic pregnancy. Hypertension in pregnancy 1997;16:187-202
9 Natarajan P, Shennan AH, Penny J, Halligan AW, de Swiet M, Anthony J. Comparison of auscultatory and oscillometric automated blood pressure monitors in the setting of preeclampsia. Am J Obstet Gynecol 1999;181:1203-1210
10 Shennan A, Haligan A, Gupta M, Taylor D, de Swiet M. Oscillometric blood pressure measurements in severe preeclampsia: validation of the Spacelabs 90207. Br J obstet & Gynaecol 1996;103:171-3
11 Belsha CW, Wells TG, Rice HB, Neaville WA, Berry PL. Accuracy of the Spacelabs 90207 ambulatory blood pressure monitor in children and adolescents. Blood Press Monit 1996;1:127-133
12 Iqbal P, Fotherby MD, Potter JF. Validation of the Spacelabs 90207 automatic non-invasive blood pressure monitor in elderly subjects. Blood Press Monit 1996;1:367-73
13 Peixoto AJ, Gray TA, Crowley ST. Validation of the Spacelabs 90207 ambulatory blood pressure device for hemodialysis patients. Blood Press Monit 1999;4:217-221
15 Nemeth Z, Moczar K, Deak G. Evaluation of the Tensioday ambulatory blood pressure monitor according to the the protocols of the British Hypertension society and the Association for the Advancement of Medical Instrumentation. Blood Press Monit 2002;7:191-197
16 Tannahill G, Gilmore C, Silke B. Validation of the upper arm algorithm for blood pressure determination incorporated in the Seinex digital devices using the AAMI and British Hypertension Society protocols Northern Ireland Medicine Today 2002;33:26-27
17 Altunkan Ş, Iliman N, Altunkan E. Validation of the Nissei DS-250 ambulatory blood pressure monitoring device according to the International Protocol J Hypertens 2003;21 (suppl 4):S22(P1.5) (Abstract)
18 Jones, SC, Bilous M, Winship S, Finn P, Goodwin J. Validation of the OSCAR 2 oscillometric 24-hour ambulatory blood pressure monitor according to the International Protocol for the validation of blood pressure measuring devices. Blood Press Monit 2004;9:219-223
  Rogoza AN. Accuracy of blood pressure estimation by automated monitors Functional diagnostics 2003;1:56-64. (In Russian)
20 O'Brien E, Atkins N, Murphy A, Lyons S. Validation of the Tonoport V ambulatory blood pressure monitor according to the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults Blood Press Monit 2003;8:255-260
21 Mallion JJM, Pierre H, Neuder Y, Ormezzano O, Baguet JP Validation of the AGILIS ambulatory blood pressure monitor according to the European society of Hypertension International Protocol for Validation of Blood Pressure Measuring devices in Adults Blood Press Monit 2005;10:97-101
22 Haensel A, Utech K, Langewitz W Validation of the TONOPORT V blood pressure measuring monitor in adults Journal of Human Hypertension 2005; 19 :745-749
24 Goodwin J, Bilous M, Winship S, Finn P, Jones SC. Validation of the Oscar 2 oscillometric 24-h ambulatory blood pressure monitor according to the British Hypertension Society protocol Blood Pressure Monitoring 2007;12:113-117
25 Friz HP, Sega R, Facchetti R, Primit L, Beltrame L, Bombelli M. Accuracy evaluation of the 'Cardiette BP one' ambulatory blood pressure monitor Blood Press Monit 2008;13:107-110
26 Nair D, Tan SY, Gan HW, Lim SF, Tan J, Zhu M, Gao H, Chua NH, Peh WL, Mak KH. The use of ambulatory tonometric radial arterial wave capture to measure ambulatory blood pressure: the validation of a novel wrist-bound device in adults Journal of Human Hypertension 2008;22:220-222 (Research Letter)
27 Yip GW, So HK, Li AM, Tomlinson B, Wong SN, Sung RY. Validation of A&D TM-2430 upper-arm blood pressure monitor for ambulatory blood pressure monitoring in children and adolescents, according to the British Hypertension Society protocol. Blood Press Monit 2012;17(2):76-79 doi: 10.1097/MBP.0b013e328351d4a4
28 Jones DP, Richey PA, Alpert BS. Validation of the AM5600 ambulatory blood pressure monitor in children and adolescents. Blood Press Monit 2008;13:349-351
29 Langewitz W, Tanner S. Validation of the PhysioQuant blood pressure measuring monitor in adults. Blood Press Monit 2009;14:220-221
30 Denchev SV, Simova II, Matveev MG. Evaluation of the Schiller BR-102 plus noninvasive ambulatory blood pressure monitor according to the International Protocol introduced by the Working Group on Blood Pressure Monitoring of the European Society of Hypertension Blood Press Monit 2007;12:329-333
31 Stergiou GS, Tzamouranis D, Nasothimiou EG, Karpettas N, Protogerou A. Are there really differences between home and daytime ambulatory blood pressure? Comparison using a novel dual-mode ambulatory and home monitor. J Hum Hypertens 2010;24(3):207-12.
32 Ragazzo F, Saladini F, Palatini P. Validation of the Microlife WatchBP O3 device for clinic, home, and ambulatory blood pressure measurement, according to the International Protocol. Blood Press Monit 2010;15(1):59-62.
33 Weiß W, Tölle M, Zidek W, van der Giet M. Validation of the Mobil-O-Graph: 24 h-blood pressure measurement device. Blood Press Monit 2010;15:225-228 doi: 10.1097/MBP.0b013e328338892f
  Bartosh LF, Dorogova IV, Kuznetsova TN, Krylova AV. Validation of the BPLab® 24-hour blood pressure monitor in pregnancy in accordance with the ESH-2001 international protocol. Arterial Hypertension 2006;12(3):3-6 (In Russian)
35 Alpert BS. Validation of the Tiba Medical Ambulo 2400 ambulatory blood pressure monitor to the ISO Standard and BHS protocol. Blood Press Monit 2010;15(5):275-277 doi: 10.1097/MBP.0b013e32833c8b39
36 Bramlage P, Deutsch C, Krüger R, Wolf A, Müller P, Zwingers T, Beime B, Mengden T. Validation of the custo screen 400 ambulatory blood pressure-monitoring device according to the European Society of Hypertension International Protocol revision 2010. Vascular Health and Risk Management 2014;10:303-309 doi: 10.2147/VHRM.S63602
37 A de Greeff, AH Shennan. Validation of Spacelabs 90227 OnTrak upper arm blood pressure monitor, for clinical use, according to the European Society of Hypertension International Protocol 2010 and the British Hypertension Society Protocol. Available from: ESH-IP 2010 and BHS Validation of Spascelabs 90227 OnTrak
38 Bilo G., Zorzi C., Ochoa Munera J.E., Torasco C., Giuli V. and Parati G. Validation of the Somnotouch-NIBP non-invasive continuous blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010. Blood Pressure Monitoring 2015 Oct; 20(5):291-94 doi: 10.1097/MBP.0000000000000124
39 M. Abou-Dakn, C. Döhmen and S. Wenzel Validation of the TONOPORT VI ambulatory blood pressure monitor in adults according to the European Society of Hypertension International Protocol revision 2010. Journal of Human Hypertension advance online publication, 14 July 2016; doi:10.1038/jhh.2016.40
40 Beime B., Deutsch C., Krüger R., Wolf A., Müller P., Hammel G., Bramlage P. Validation of the custo screen pediatric blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010. European Journal of Paediatrics 2017 Feb 24, Article 2874. doi: 10.1007/s00431-017-2874-3
41 Fania C. MD, Vezzù L. RN, Lazzaretto I. RN, Palatini P. MD. Validation of the Hingmed WBP-02A upper arm oscillometric monitor for ambulatory blood pressure measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010. Available from: ESH-IP 2010 Validation of Hingmed WBP-02A.pdf
42 Kario K., Hoshide S., Saito K., Sato K., Hamasaki H., Suwa H., Tomitani N. Validation of the TM-2441 ambulatory blood pressure measurement device according to the ISO 81060-2: 2013 standard. Blood Press Monit. 2019 Feb;24(1):38-41. doi: 10.1097/MBP.0000000000000357
176 Validation of the Beneware model ABP-021 ambulatory blood pressure monitor according to the revised 2010 European Society of hypertension international protocol Telmo Pereira and João Guimarães. Blood Pressure Monitoring 23(4):p 210-213, August 2018. | DOI: 10.1097/MBP.0000000000000330