Classification of sphygmomanometers/blood pressure measuring devices A table of blood pressure measurement devices as recommended by the dabl Educational Trust Advisory Board Table of devices currently available on the market Table of devices that have been discontinued by their manufacturers
Current page: Index of devices discontinued by their manufacturer Current page: table of devices discontinued by their manufacturer

Discontinued Sphygmomanometers

Below is a list of blood pressure monitors which previously appeared on the main device tables but have been moved to this table due to the fact that they are no longer available on the market. We have kept the recommendation column should you wish to check the status of a device you already have but which does not appear on the main recommendation tables.

For a complete list of devices see our Device Index and for discontinued devices see our Discontinued Devices Index.

Device Mode AAMI BHS ESH Circumstance Recommendation Ref
Classification
A&D TM-2420
Model 6
Osc Pass B/B   At rest Recommended 16
A&D TM-2420
Model 7
Osc Pass B/B   At rest Recommended 17
A&D TM-2420 /
TM-2020
Osc Fail D/D   At rest Not Recommended 40
A&D TM-2421 Osc Pass B/A   At rest Recommended 18
  A/B   In children aged 7-8 years sitting Questionable 23
  A/B   In children of all ages sitting Questionable 23
  C/C   In children & different posture Not Recommended 23
Fail     At rest Not Recommended 43
Aus Pass     At rest Questionable 43
Accutracker II (30/23) (SunTech) Aus Pass A/C   At rest Not Recommended 24
Boso-Mediwatch Wrist   C/C   At rest; protocol violation Not Recommended 35 SBPM Wrist
Braun BP 3000 Wrist Pass     At Rest Questionable 74 SBPM Wrist
Pass     At rest Questionable 75
Braun BP 3550 Wrist     Pass At rest. Questionable in high range. Recommended 76 SBPM Wrist
Braun PrecisionSensor BP2000 Wrist Pass     At rest, Manufacturer Assistance Questionable 77 SBPM Wrist
Braun PrecisionSensor BP2550 Wrist Pass B/B   Recruitment violation (AAMI) Recruitment violation, Analysis errors (BHS) Questionable 78, 79 SBPM Wrist
  ?B/B   Abstract, SBP results unclear Questionable 80
Braun PrecisionSensor BP2550 (UG) Wrist     Pass At rest Recommended 81 SBPM Wrist
Braun VitalScan Wrist Pass ??   Abstract; Version not stated Questionable 82 SBPM Wrist
CAS Model 9010 Osc Pass NA NA At rest in adults Questionable 13
Neonates Questionable 13
Colin BP-8800 Osc Fail D/D   Protocol violation Not Recommended 14
Colin Pilot 9200 Tonometry Pass     At rest; intra-arterial Questionable 15
Datex-Engstrom Cardiocap II Osc Fail D/D   Protocol violation Not Recommended 19
Daypress 500 (Advantage Medical Research) Osc Pass A/B   At rest Recommended 2
Del Mar
Pressurometer IV
Aus Fail C/D   At rest Not Recommended 27
Dinamap 1846 SX Osc Pass C/C   At rest; diabetic patients Questionable 19
Dinamap 1846 XT Osc Pass C/C   At rest; varying pressures, Recruitment violation, procedure violation Questionable 66 Clinical Automated
Dinamap 8100 Osc Fail B/D   At rest Not Recommended 32,33
Pass B?   Abstract; AAMI and BHS guidelines attached to ad-hoc protocol Questionable 34
Disetronic
Profilomat
Aus Pass B/A   At rest Recommended 6
Pass B/C   In pregnancy Not Recommended 28
Disetronic Profilomat II Osc Fail C/B   At rest Not Recommended 29
Figi OSCILL-IT Osc Pass C/B   At rest Not Recommended 50
Fortec Dr MI-100 Osc Fail Fail*   At rest Not Recommended 44
Hawksley RZS: UK model Hg Fail C/D   At rest Not Recommended 46,47
Hawksley RZS: US model Hg Fail B/D   At rest Not Recommended 46,47
Healthcheck CX-5 060020 Osc Fail Fail*   At rest Not Recommended 44
IEM
Mobil-O-Graph (version 12)
Osc Pass B/A   At rest Recommended 71
Jotatec P24B Osc     Pass Abstract Only Questionable 72
Medilog ABP Aus Pass     At rest Questionable 20
Microlife BP 3BU1-5 Wrist     Fail At rest Not Recommended 67
Microlife M-L Rigid Cylindrical Cuff (Tested with BP A100) Osc     Fail With M-L Rigid Cylindrical Cuff - Cuff Never Released; Device valid with other cuffs Not Recommended 73
Nissei Analogue Monitor Aus Fail Fail*   At rest Not Recommended 44
Nissei DS-175 Osc Fail D/A   At rest Not Recommended 45
Nissei DS-240 Osc Pass B/A   Abstract; detail missing Questionable 21
Nissei WS-310 Wrist Pass     Abstract; Protocol violation Questionable 36
Novacor DIASYS 200 Aus Pass C/C   At rest Not Recommended 25
Novacor DIASYS Integra Aus Pass B/A   At rest Recommended 3
Osc Pass B/B   At rest Recommended 3
Omron HEM-400C Osc Fail Fail*   At rest Not Recommended 44
Omron HEM-403C Osc Fail C/C   Protocol violation Not Recommended 55
Omron HEM-703CP Osc Pass     Intra-arterial Questionable 58
Omron HEM-705CP Osc Pass B/A   At rest Recommended 59
Pass A/A   At rest; ranges of pressure Recommended 60
  B/B   Abstract;
Children and adolescents
Questionable 62
Omron HEM-706 Osc Pass B/C   At rest Not Recommended 63
Omron HEM-706/711 Osc Pass B/A   At rest; ranges of pressure Recommended 60
Omron HEM-713C Osc Pass B/B   At rest Recommended 61
Omron HEM-722C Osc   A/A   Protocol violation Questionable 64
Pass A/A   Rest/Elderly Recommended 65
Omron HEM-735C Osc Pass B/A   Rest/Elderly Recommended 65
Omron HEM-737 Intellisense Osc Pass B/B   At rest Recommended 51
Omron M4 Osc Pass A/A   Abstract Questionable 57
  C/D   Protocol violation Questionable 56
Omron MIT (HEM-751) Osc on inflation Pass B/A   At rest; adults Recommended 49
Pass B/A   At rest; Pregnancy Recommended 48
  A/A   Pregnancy Recommended 69
  A/A   Pre-Eclampsia Recommended 69
Omron MX2 Osc Pass A/A   Abstract Questionable 57
Omron R3 Wrist       Intra-arterial comparison Questionable 37
Fail D/D   At rest Not Recommended 38
Omron RX Wrist Fail B/B   At rest; Abstract Questionable 39
Pass     Abstract; Protocol violation Questionable 36
Fail B/B   At rest; Min Recruitment for BP>130/80, C/C for Mid/High ranges, Bare SBP B, ? recruitment bias Questionable 42
Omron RX-M (HEM-628) Wrist Fail D/D Fail/Fail At rest;
One observer;
IP retrospective
Not Recommended 5,41
Philips HP5308 Aus Fail Fail*   At rest Not Recommended 44
Philips HP5306/B Osc Fail Fail*   At rest Not Recommended 44
Philips HP5332 Osc Fail C/A   At rest Not Recommended 45
Rossmax Osc on inflation     Fail At rest Not Recommended 70
Save 33, Model 2 Osc Pass B/B   At rest Recommended 9
Schiller BR-102 Aus Pass B/B   At rest Recommended 52
Osc Fail D/B   At rest Not Recommended 53
Spacelabs 90202 Osc Pass B/B   At rest Recommended 11
Systema Dr MI-150 Osc Fail Fail*   At rest Not Recommended 44
Terumo ES-H531 Aus Pass A/A   At rest Recommended 4
Osc Pass B/B   At rest Recommended 4
Tycos
QuietTrak
Aus Pass B/B   At rest Recommended 7
Pass A/A   At rest Recommended 8
  A/A   During exercise Recommended 8
  A/A   Different posture Recommended 8
  A/A   In the elderly Recommended 8
  A/A   In children Recommended 8
  A/A   In pregnancy Recommended 8
Pass B/B   At rest. Abstract Questionable 22
Fail D/D   In pre-eclampsia Not Recommended 30
Fail B/B   In pregnancy Not Recommended 31
Visomat OZ2 Osc Pass C/B   At rest Not Recommended 68  
Welch Allyn 'Vital Signs' Osc Pass A/A   At rest. Pressure ranges Recommended CA7
Fail C/D   At rest. Not Recommended CA8
    Fail Children. Adaptation of protocol along BHS guidelines. Not Recommended CA22

References for Validation of Devices

1 O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the CH-Druck/Pressure ERKA ambulatory blood pressure measuring system determined by the British Hypertension Society protocol. J Hypertens 1993;11(suppl 2):S1-S7.
2 Livi R, Teghini L, Cagnoni S, Scarpelli PT. Simultaneous and Sequential Same-Arm Measurements in the Validation Studies of Automated Blood Pressure Measuring Devices. Am J Hypertens 1996;9:1228-1231
3 Gosse P, Laforge A, Ansoborto P, Julien P, Lemetayer P, Clementy J. Clinical evaluation of the DIASYS integra blood pressure recorder. J Hypertens 1997;15:S18. Abstract
4 Kuwajima I, Nishinaga M, Kanamaru A. The Accuracy and Clinical Performance of a New Compact Ambulatory Blood Pressure Monitoring Device, the ES-H531. Am J Hypertens 1998;11:1328-1333
5 Braam R, Aslan B, Thein T. Accuracy of the Omron RX-M, an automated wrist blood pressure measuring device J Hypertens 2003;[dabl® Educational Trust regrets that full details are missing] (Abstract)
6 O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the Profilomat ambulatory blood pressure measuring system determined by the British Hypertension Society Protocol. J Hypertens 1993;11(suppl 22):S9-S15
7 White WB, Susser W, James G, Marra L, McCabe EJ, Pickering TG, Streeten DHP. Multicentre assessment of the QuietTrak ambulatory blood pressure recorder according to the 1992 AAMI guidelines. Am J Hypertens 1994;7:509-14
8 Modesti PA, Costoli A, Cecioni I, Toccafondi S, Carnemolla A, Serneri GGN. Clinical evaluation of the QuietTrak blood pressure recorder according to the protocol of the British Hypertension Society. Blood Press Monit 1996;1:63-68
9 De Gaudemaris R, White A, Pascal I, Siche J-P, Baguet J-P, Mallion J-M. Évaluation du moniteur de pression artérielle en ambulatoire Save 33 modèle2. Arch Mal Coeur 1999;92:1133-1137
10 Mee F, Atkins N, O’Brien E. Evaluation of the Shiller BR-102 ambulatory blood pressure system according to the protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1999;4:35-43
11 O'Brien E, Atkins N, Mee F, O'Malley K. Evaluation of the Spacelabs 90202 according to the AAMI standard and BHS criteria. J Human Hypertens 1991;5:223-226
12 O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the Spacelabs 90207 determined by to the British Hypertension Society Protocol. J Hypertens 1991;9(suppl 5):S25-S31
13 Alpert BS. Validation of CAS model 9010 automated blood pressure monitor: children/adult and neonatal studies Blood Press Monit 1996;1:69-73
14 Naschitz JE, Gaitini L, Loewenstein L, Kedren D, Zuckermann E, Tamir A, Yeshurun D. In-field validation of automatic blood pressure measuring devices J Hum Hypertens 2000;14:37-42
15 Zorn EA, Wilson MB, Angel JJ, Zanella J, Alpert B. Validation of an automated arterial tonometry monitor using Association for the Advancement of Medical Instrumentation standards Blood Press Monit 1997;2:185-188
16 White WB, Pickering TG, Morganroth J, James GD, McCabe EJ, Moucha O, Hunter H. A multicenter evaluation of the A&D TM-2420 ambulatory blood pressure recorder. Am J Hypertens 1991;4:890-6
17 Palatini P, Penzo M, Canali C, Pessina AC. Validation of the accuracy of the A&D TM2420 Model 7 for ambulatory blood pressure monitoring and effect of microphone replacement on its performance. J Ambul Monitor 1991;4:281-8
18 Imai Y, Sasaki S, Minami N, Munakakata M, Hashimoto J, Sakuma H, Sakuma M, Watanabe N, Imai K, Sekino H, Abe K. The accuracy and performance of the A&D TM 2421, a new ambulatory blood pressure monitoring device based on the cuff-oscillometric method and the Korotkov sound technique. Am J Hypertens 1992;5:719-26
19 Kuo CS, Hwu CM, Kwok CF, Hsiao LC, Weih MJ, Lee SH, Ho LT. Using semi-automated oscillometric blood pressure measurement in diabetic patients and their offspring J Diabetes and complications 2000;14:288-293
20 Manning G, Vijan SG, Millar-Craig MW. Technical and clinical evaluation of the Medilog ABP non-invasive blood pressure monitor. J Ambul Monnit 1994;7:255-264
21 Mee F, Atkins N, O'Brien E. Validation of the Nissei DS-240 ambulatory blood pressure measuring system as determined by the British Hypertension Society protocol. J Human Hypertens 1994;8:295
22 O’Shea JC, O’Neill A, O’Brien C, Murphy MB. Evaluation of the Tycos Quiet Trak ambulatory pressure recorder using British Hypertension Society (BHS) protocol. Am J Hypertens 1994;7:118A. Abstract
23 O’Sullivan JJ, Derrick G, Griggs PE, Wren C. Validation of the Takeda 2421 ambulatory blood pressure monitor in children. J Med Eng & Tech 1998;22:101-105
24 Taylor R, Chidley K, Goodwin J, Broeders M, Kirby B. Accutracker II (version 30/23) ambulatory blood pressure monitor: clinical validation using the British Hypertension Society and Association for the Advancement of Medical Instrumentation standards. J Hypertens 1993;11:1275-1282
25 O’Brien E, Mee F, Atkins N, O’Malley K. Accuracy of the Novacor DIASYS 200 determined by the British Hypertension Society Protocol. J Hypertens 1991;9(suppl 5):S9-S15
26 Germanó G, Muscolo M, Angotti S, Bravo S, Codispoti P, Federico L, Damiani S. Evaluation of a New Ambulatory Blood Pressure Device. Am J Hypertens 1998;11:1486-1491
27 O'Brien E, Mee F, Atkins N, O'Malley K. Short Report: Accuracy of the Del Mar Avionics Pressurometer IV determined by the British Hypertension Society Protocol. J Hypertens 1991;9:567-568
28 Franx A, van der Post JAM, van Montfrans GA, Bruinse HW. Comparison of an auscultatory versus an oscillometric ambulatory blood pressure monitor in normotensive, hypertensive and pre-eclamptic pregnancy. Hypertension in pregnancy 1997;16:187-202
29 Mee F, Atkins N, O’Brien E. Evaluation of the Profilomat II ambulatory blood pressure system according to the protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1999;3:353-361
30 Natarajan P, Shennan AH, Penny J, Halligan AW, de Swiet M, Anthony J. Comparison of auscultatory and oscillometric automated blood pressure monitors in the setting of preeclampsia. Am J Obstet Gynecol 1999;181:1203-1210
31 Penny JA, Shennan AH, Rushbrook J, Halligan A, Taylor DJ, de Swiet M. Validation of the Welch Allyn QuietTrak ambulatory blood pressure monitor in pregnancy. Hypertension in Pregnancy 1996;15:313-321
32 O'Brien E, Mee F, Atkins N, O'Malley K. Short report: Accuracy of the Dinamap portable monitor, model 8100 determined by the British Hypertension Society protocol J Hypertens 1993;11:761-763
33 O'Brien E, De Gaudemaria R, Bobrie G, Agabiti Rosei E, Vaisse B and the participants of the First International Consensus Conference on Blood Pressure Self-Measurement. Devices and validation Blood Pressure Monit 2000;5:93-100
34 Parker SB, Steigerwalt SP. The Dinamap dilemma: Inaccuracy of the commonly used Dinamap 8100 compared to simultaneous mercury manometer measurements in hospitalized patients at different levels of blood pressure Am J Hypertens 2004;17(No. 5, Part 2):52A (Abstract P-57)
35 Rogers P, Burke V, Stroud P, Puddey IB. Comparison of oscillometric blood pressure measurements at the wrist with an upper-arm auscultatory mercury sphygmomanometer Clin & Exper Pharmacol & Pharmacol 1999;26:477-481
36 Altunkan Ş, Yildiz SA. Wrist blood pressure measuring devices: a comparative study of the accuracy with a standard auscultatory method using a mercury sphygmomanometer. [dabl® Educational Trust regrets that full details are missing] (Abstract)
37 Eckert S, Gleichmann U, Zagorski O, Klapp A. Validation of the Omron R3 blood pressure self-measuring device through simultaneous comparative invasive measurements according to protocol 58130 of the German Institute for Validation Blood Press Monit 1997;2:189-192
38 Dieterle T, Battegay E, Bucheli B, Martina B. Accuracy and 'range of uncertainty' of oscillometric blood pressure monitors around the upper arm and wrist Blood Press Monit 1998;3:339-346
39 Shennan AH, Rushbrook J, Power J, Wright J, Poston L. An accurate oscillometric wrist blood pressure monitor: validation of the Omron Rx [HEM-608] J Human Hypertens 1998;12:794 (Abstract)
40 O'Brien E, Mee F, Atkins N, O'Malley K. Short Report: Accuracy of the Takeda TM-2420/TM-2020 determined by the British Hypertension Society Protocol. J Hypertens 1991;9:571-572
41 Braam R, Aslan B, Thein T. Accuracy of the Omron RX-M, an automated blood pressure measuring device, measuring blood pressure at the wrist, according to a modified British Hypertension Society protocol Blood Press Monit 2004;9:25-30
42 Cuckson AC, Moran P, Seed P, Reinders A, Shennan AH. Clinical evaluation of an automated blood pressure wrist device Blood Press Monit 2004;9:31-37
43 Keaveney B, Bird R, Caiazza A, Casadei B, Conway J. Measurement of blood pressure using the auscultatory and oscillometric methods in the same cuff deflation: validation and field trial of the A&D TM2421 monitor. J Human Hypertens 2000;14:573-579
44 O'Brien E, Mee F, Atkins N, O'Malley K. Inaccuracy of seven popular sphygmomanometers for home-measurement of blood pressure. J Hypertens 1990;8:621-634
45 O'Brien E, Mee F, Atkins N, Thomas M. Evaluation of three devices for self-measurement of blood pressure: according to the revised British Hypertension Society Protocol: the Omron HEM-705CP, Phillips HP5332, and Nissei DS-175. Blood Pressure Monitoring 1996;1:55-61
46 O'Brien E, Mee F, Atkins N, O'Malley K. Inaccuracy of the Hawksley random-zero sphygmomanometer. Lancet 1990;336:1465-1468.
47 Brown WCB, Kennedy S, Inglis GC, Murray LS, Lever AF. Mechanisms by which the Hawksley random zero sphygmomanometer underestimates blood pressure and produces a non-random distribution of RZ values. J Hum Hypertens 1997;11:75-93
48 Golara M,Benedict A, Jones C. Randhawa M, Poston L, Shennan AH. Inflationary oscillometry provides accurate measurement of blood pressure in pre-eclampsia Br J Obstet Gynaecol 2002;109:1143-1147
49 Cuckson AC, Jones CR, Randhawa M, Vincent T, Poston L, Shennan AH. Oscillometry detected on inflation can accurately reflect mercury sphygmomanometry J Hypertension 2002;20 (suppl 4):S234 (Abstract)
50 Germanó G, Muscolo M, Angotti S, Bravo S, Codispoti P, Federico L, Damiani S. Evaluation of a New Ambulatory Blood Pressure Device. Am J Hypertens 1998;11:1486-1491
51 Anwar YA, Giacco S, McCabe EJ, Tendler BE, White WB. Evaluation of the efficacy of the Omron HEM-737 Intellisense device for use on adults according to the recommendations of the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1998;3:261-26
52 Mee F, Atkins N, O’Brien E. Evaluation of the Shiller BR-102 ambulatory blood pressure system according to the protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1999;4:35-43
53 Mee F, Atkins N, O’Brien E. Evaluation of the Shiller BR-102 ambulatory blood pressure system according to the protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1999;4:35-43
55 Walma EP, van Dooren C, van der Does E, Prins A, Mulder P, Hoes AW. Accuracy of an oscillometric automatic blood pressure device: the Omron HEM403C. J Human Hypertens 1995;9:169-174.
56 Naschitz JE, Loewenstein L, Lewis R, Keren D, Gaitini L, Tamir A, Yeshurun D. Accuracy of the OMRON M4 automatic blood pressure measuring device. J Hum Hypertens 2000;14:423-427
57 Artigao M, Llavador J, Rubio M, Torres C, Lopez J, Sanchis C, Vidal A, Puras A. Evaluation of two devices for self-measurement of blood pressure according to the British Hypertension Society protocol: The Omron M4 and MX2. Am J Hypertens 1998;16:S276 (Abstract)
58 Rithalia SVS, Edwards D. Comparison of oscillometric and intra-arterial blood pressure and pulse measurement. J Med Engin Tech 1994;18:179-181.
59 O'Brien E, Mee F, Atkins N. An accurate automated device for home blood pressure measurement at last! The Omron HEM-705CP. J Hypertens 1994;12:1317-1318
60 Artigao LM, Llavador JJ, Puras A, Lopex-Abril J, Rubio MM, Torres C, Vidal A, Sanchi C, Division JA, Naharro F, Caldevilla D, Fuentes G. Evaluation and validation of Omron Hem 705 CP and Hem 706/711 monitors for self-measurement of blood pressure. 2000.[dabl® Educational Trust regrets that full details are missing]
61 Mufunda J, Sparks B, Chifomba J, Dakwa C, Matenga JA, Adams JA, et al. Comparison of the Omron HEM-713C automated blood pressure monitor with a standard auscultatory method using a mercury manometer. Central Afr J Med 1996;42:230-232
62 Furusawa EA, Ruiz MFO, Saito MI, Koch VH Evaluation of the Omron 705-CP blood pressure monitor for use in children and adolescents J Hypertens 2004;22 (suppl 1):S140 (Abstract P510)
63 Foster C, McKinlay JM, Cruickshank JM, Coats AJS. Accuracy of the Omron HEM 706 portable monitor for home measurement of blood pressure. J Human Hypertens 1994;8:661-664.
64 Córdoba R, Fuertes MI, Alvarez A, Molina L, Solans R, Melero I. Validation of a self-monitoring blood pressure monitor: The Omron-HM 722C. Aten Primaria 1997;20:247-250
65 Bortolotto LA, Henry O, Hanon O, Sikias P, Mourad J-J, Girerd X. Validation of two devices for self-measurement of blood pressure by elderly patients according to the revised British Hypertension Society protocol: the Omron HEM-722C and HEM-735C. Blood Press Monit 1999;4:21-25
66 Beaubien ER, Card CM, Card SE, Biem HJ, Wilson TW. Accuracy of the Dinamap 1846 XT automated blood pressure monitor J Hum Hypertens 2002;16:647-652
67 Altunkan S, Iliman N, Altunkan E. Validation of the Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure measuring devices in adults according to the International Protocol Blood Pressure Monitoring 2007;12:119-125
68 Dieterle T, Battegay E, Bucheli B, Martina B. Accuracy and 'range of uncertainty' of oscillometric blood pressure monitors around the upper arm and wrist. Blood Press Moni 1998;3:339-346
69 de Greeff A, Beg Z, Gangji Z, Dorney E, Shennan AH. Accuracy of inflationary versus deflationary oscillometry in pregnancy and preeclampsia: OMRON-MIT versus OMRON-M7. Blood Press Monit 2009;14:37-40
70 O'Brien E, Atkins N, Murphy A, Lyons S. Validation of the Rossmax blood pressure measuring monitor according to the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults Blood Press Monit 2003;8:261-265
71 Jones CR, Taylor K, Chowienczyk P, Poston L, Shennan A. A validation of the Mobil O Graph (version 12) ambulatory blood pressure monitor. Blood Press Monit 2000;5:233-238
72 Octavio JA, Leal N, Contreras J, Octavio B, Cressa L, Valecillos I, Castellanos D, Fabiano D. Accuracy of Jotatec®-P24B device for teletransmitted ambulatory blood pressure monitoring J Hypertens 2008;26 (suppl. 1):S493 (Abstract PS34/MON/60)
73 Bonso E, Saladini F, Zanier A, Benetti E, Dorigatti F, Palatini P. Accuracy of a single rigid conical cuff with standard-size bladder coupled to an automatic oscillometric device over a wide range of arm circumferences. Hypertens Res. 2010;33(11):1186-91.
74 Nolly H, Romero M, Nolly A, Silva M, Nolly M. Clinical evaluation of the Braun BP 3000 Easy Click according to the ANSI/AAMI SP 10:2002 Blood Pressure Monitoring 2006;11:91-95
75 Nolly H, Romero M, Nolly A, Silva M, Nolly M. Clinical evaluation of the Braun BP 3000 Easy Click according to the ANSI/AAMI SP 10:2002. Blood Press Monit 2006;11:91-95
76 Iliman N, Altunkan S, Kayaturk N, Altunkan E. Validation of the Braun BP 3550 wrist blood pressure measuring device with a position sensor and an EasyClick cuff according to the International Protocol in adults. Blood Press Monit 2007;12:45-49
77 Wessig K, Hollinger S, Schmalzhaf I, Lenz T. Clinical evaluation of the efficacy of the Braun PrecisionSensor oscillometric wrist blood pressure monitor for use on adults versus auscultation as defined by ANSI/AAMI SP 10-1992 Blood Pres Monit 2000;5:239-245
78 Nolly H, Osso P, Nolly M. Home blood pressure monitoring Intercontinental Cardiology 2001;10:111-114
79 Nolly H, Osso A, Romero M, Nolly M. The reproducibility of home- self measurements using the Braun PrecisionSensor BP 2550 Monitor J Hypertens 2001;19 (suppl 2):S26
80 Sandevik E, Dyrdal A. Evaluation of the Braun BP2550 PrecisionSensor wrist blood pressure monitor in clinical practice [dabl® Educational Trust regrets that full details are missing] (Abstract)
81 Nolly H, Romero M, Nolly A, Osso P, Reinoso O, Nolly M. Home blood pressure measurement: validation of the Braun BP 2550 (UG) monitor according to the ESH International Protocol Blood Press Monit 2004;9:53-58
82 Nolly H, Romero M, Nolly A, Nolly M. Home blood pressure (BP) measurement using B.P. Braun Vitalscan monitor. J Hypertens 200;18(suppl 4):S119 (Abstract)