Below is a list of blood pressure monitors which
previously appeared on the main device tables but have been moved to this table due to
the fact that they are no longer available on the market. We have kept the
recommendation column should you wish to check the status of a device you already have
but which does not appear on the main recommendation tables.
1 |
O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the CH-Druck/Pressure ERKA ambulatory blood pressure measuring system determined by the British Hypertension Society protocol. J Hypertens 1993;11(suppl 2):S1-S7. |
2 |
Livi R, Teghini L, Cagnoni S, Scarpelli PT. Simultaneous and Sequential Same-Arm Measurements in the Validation Studies of Automated Blood Pressure Measuring Devices. Am J Hypertens 1996;9:1228-1231 |
3 |
Gosse P, Laforge A, Ansoborto P, Julien P, Lemetayer P, Clementy J. Clinical evaluation of the DIASYS integra blood pressure recorder. J Hypertens 1997;15:S18. Abstract |
4 |
Kuwajima I, Nishinaga M, Kanamaru A. The Accuracy and Clinical Performance of a New Compact Ambulatory Blood Pressure Monitoring Device, the ES-H531. Am J Hypertens 1998;11:1328-1333 |
5 |
Braam R, Aslan B, Thein T. Accuracy of the Omron RX-M, an automated wrist blood pressure measuring device J Hypertens 2003;[dabl® Educational Trust regrets that full details are missing] (Abstract) |
6 |
O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the Profilomat ambulatory blood pressure measuring system determined by the British Hypertension Society Protocol. J Hypertens 1993;11(suppl 22):S9-S15 |
7 |
White WB, Susser W, James G, Marra L, McCabe EJ, Pickering TG, Streeten DHP. Multicentre assessment of the QuietTrak ambulatory blood pressure recorder according to the 1992 AAMI guidelines.
Am J Hypertens 1994;7:509-14
|
8 |
Modesti PA, Costoli A, Cecioni I, Toccafondi S, Carnemolla A, Serneri GGN. Clinical evaluation of the QuietTrak blood pressure recorder according to the protocol of the British Hypertension Society. Blood Press Monit 1996;1:63-68 |
9 |
De Gaudemaris R, White A, Pascal I, Siche J-P, Baguet J-P, Mallion J-M. Évaluation du moniteur de pression artérielle en ambulatoire Save 33 modèle2. Arch Mal Coeur 1999;92:1133-1137 |
10 |
Mee F, Atkins N, O’Brien E. Evaluation of the Shiller BR-102 ambulatory blood pressure system according to the protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1999;4:35-43 |
11 |
O'Brien E, Atkins N, Mee F, O'Malley K. Evaluation of the Spacelabs 90202 according to the AAMI standard and BHS criteria. J Human Hypertens 1991;5:223-226 |
12 |
O'Brien E, Mee F, Atkins N, O'Malley K. Accuracy of the Spacelabs 90207 determined by to the British Hypertension Society Protocol. J Hypertens 1991;9(suppl 5):S25-S31 |
13 |
Alpert BS. Validation of CAS model 9010 automated blood pressure monitor: children/adult and neonatal studies Blood Press Monit 1996;1:69-73 |
14 |
Naschitz JE, Gaitini L, Loewenstein L, Kedren D, Zuckermann E, Tamir A, Yeshurun D. In-field validation of automatic blood pressure measuring devices J Hum Hypertens 2000;14:37-42 |
15 |
Zorn EA, Wilson MB, Angel JJ, Zanella J, Alpert B. Validation of an automated arterial tonometry monitor using Association for the Advancement of Medical Instrumentation standards Blood Press Monit 1997;2:185-188 |
16 |
White WB, Pickering TG, Morganroth J, James GD, McCabe EJ, Moucha O, Hunter H. A multicenter evaluation of the A&D TM-2420 ambulatory blood pressure recorder. Am J Hypertens 1991;4:890-6 |
17 |
Palatini P, Penzo M, Canali C, Pessina AC. Validation of the accuracy of the A&D TM2420 Model 7 for ambulatory blood pressure monitoring and effect of microphone replacement on its performance. J Ambul Monitor 1991;4:281-8 |
18 |
Imai Y, Sasaki S, Minami N, Munakakata M, Hashimoto J, Sakuma H, Sakuma M, Watanabe N, Imai K, Sekino H, Abe K. The accuracy and performance of the A&D TM 2421, a new ambulatory blood pressure monitoring device based on the cuff-oscillometric method and the Korotkov sound technique. Am J Hypertens 1992;5:719-26 |
19 |
Kuo CS, Hwu CM, Kwok CF, Hsiao LC, Weih MJ, Lee SH, Ho LT. Using semi-automated oscillometric blood pressure measurement in diabetic patients and their offspring J Diabetes and complications 2000;14:288-293 |
20 |
Manning G, Vijan SG, Millar-Craig MW. Technical and clinical evaluation of the Medilog ABP non-invasive blood pressure monitor. J Ambul Monnit 1994;7:255-264 |
21 |
Mee F, Atkins N, O'Brien E. Validation of the Nissei DS-240 ambulatory blood pressure measuring system as determined by the British Hypertension Society protocol. J Human Hypertens 1994;8:295 |
22 |
O’Shea JC, O’Neill A, O’Brien C, Murphy MB. Evaluation of the Tycos Quiet Trak ambulatory pressure recorder using British Hypertension Society (BHS) protocol. Am J Hypertens 1994;7:118A. Abstract |
23 |
O’Sullivan JJ, Derrick G, Griggs PE, Wren C. Validation of the Takeda 2421 ambulatory blood pressure monitor in children. J Med Eng & Tech 1998;22:101-105 |
24 |
Taylor R, Chidley K, Goodwin J, Broeders M, Kirby B. Accutracker II (version 30/23) ambulatory blood pressure monitor: clinical validation using the British Hypertension Society and Association for the Advancement of Medical Instrumentation standards. J Hypertens 1993;11:1275-1282 |
25 |
O’Brien E, Mee F, Atkins N, O’Malley K. Accuracy of the Novacor DIASYS 200 determined by the British Hypertension Society Protocol. J Hypertens 1991;9(suppl 5):S9-S15 |
26 |
Germanó G, Muscolo M, Angotti S, Bravo S, Codispoti P, Federico L, Damiani S. Evaluation of a New Ambulatory Blood Pressure Device. Am J Hypertens 1998;11:1486-1491 |
27 |
O'Brien E, Mee F, Atkins N, O'Malley K. Short Report: Accuracy of the Del Mar Avionics Pressurometer IV determined by the British Hypertension Society Protocol. J Hypertens 1991;9:567-568 |
28 |
Franx A, van der Post JAM, van Montfrans GA, Bruinse HW. Comparison of an auscultatory versus an oscillometric ambulatory blood pressure monitor in normotensive, hypertensive and pre-eclamptic pregnancy. Hypertension in pregnancy 1997;16:187-202 |
29 |
Mee F, Atkins N, O’Brien E. Evaluation of the Profilomat II ambulatory blood pressure system according to the protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1999;3:353-361 |
30 |
Natarajan P, Shennan AH, Penny J, Halligan AW, de Swiet M, Anthony J. Comparison of auscultatory and oscillometric automated blood pressure monitors in the setting of preeclampsia. Am J Obstet Gynecol 1999;181:1203-1210 |
31 |
Penny JA, Shennan AH, Rushbrook J, Halligan A, Taylor DJ, de Swiet M. Validation of the Welch Allyn QuietTrak ambulatory blood pressure monitor in pregnancy. Hypertension in Pregnancy 1996;15:313-321 |
32 |
O'Brien E, Mee F, Atkins N, O'Malley K. Short report: Accuracy of the Dinamap portable monitor, model 8100 determined by the British Hypertension Society protocol J Hypertens 1993;11:761-763 |
33 |
O'Brien E, De Gaudemaria R, Bobrie G, Agabiti Rosei E, Vaisse B and the participants of the First International Consensus Conference on Blood Pressure Self-Measurement. Devices and validation Blood Pressure Monit 2000;5:93-100 |
34 |
Parker SB, Steigerwalt SP. The Dinamap dilemma: Inaccuracy of the commonly used Dinamap 8100 compared to simultaneous mercury manometer measurements in hospitalized patients at different levels of blood pressure Am J Hypertens 2004;17(No. 5, Part 2):52A (Abstract P-57) |
35 |
Rogers P, Burke V, Stroud P, Puddey IB. Comparison of oscillometric blood pressure measurements at the wrist with an upper-arm auscultatory mercury sphygmomanometer Clin & Exper Pharmacol & Pharmacol 1999;26:477-481 |
36 |
Altunkan Ş, Yildiz SA. Wrist blood pressure measuring devices: a comparative study of the accuracy with a standard auscultatory method using a mercury sphygmomanometer. [dabl® Educational Trust regrets that full details are missing] (Abstract) |
37 |
Eckert S, Gleichmann U, Zagorski O, Klapp A. Validation of the Omron R3 blood pressure self-measuring device through simultaneous comparative invasive measurements according to protocol 58130 of the German Institute for Validation Blood Press Monit 1997;2:189-192 |
38 |
Dieterle T, Battegay E, Bucheli B, Martina B. Accuracy and 'range of uncertainty' of oscillometric blood pressure monitors around the upper arm and wrist Blood Press Monit 1998;3:339-346 |
39 |
Shennan AH, Rushbrook J, Power J, Wright J, Poston L. An accurate oscillometric wrist blood pressure monitor: validation of the Omron Rx [HEM-608] J Human Hypertens 1998;12:794 (Abstract) |
40 |
O'Brien E, Mee F, Atkins N, O'Malley K. Short Report: Accuracy of the Takeda TM-2420/TM-2020 determined by the British Hypertension Society Protocol. J Hypertens 1991;9:571-572 |
41 |
Braam R, Aslan B, Thein T. Accuracy of the Omron RX-M, an automated blood pressure measuring device, measuring blood pressure at the wrist, according to a modified British Hypertension Society protocol Blood Press Monit 2004;9:25-30 |
42 |
Cuckson AC, Moran P, Seed P, Reinders A, Shennan AH. Clinical evaluation of an automated blood pressure wrist device Blood Press Monit 2004;9:31-37 |
43 |
Keaveney B, Bird R, Caiazza A, Casadei B, Conway J. Measurement of blood pressure using the auscultatory and oscillometric methods in the same cuff deflation: validation and field trial of the A&D TM2421 monitor. J Human Hypertens 2000;14:573-579 |
44 |
O'Brien E, Mee F, Atkins N, O'Malley K. Inaccuracy of seven popular sphygmomanometers for home-measurement of blood pressure. J Hypertens 1990;8:621-634 |
45 |
O'Brien E, Mee F, Atkins N, Thomas M. Evaluation of three devices for self-measurement of blood pressure: according to the revised British Hypertension Society Protocol: the Omron HEM-705CP, Phillips HP5332, and Nissei DS-175. Blood Pressure Monitoring 1996;1:55-61 |
46 |
O'Brien E, Mee F, Atkins N, O'Malley K. Inaccuracy of the Hawksley random-zero sphygmomanometer. Lancet 1990;336:1465-1468. |
47 |
Brown WCB, Kennedy S, Inglis GC, Murray LS, Lever AF. Mechanisms by which the Hawksley random zero sphygmomanometer underestimates blood pressure and produces a non-random distribution of RZ values. J Hum Hypertens 1997;11:75-93 |
48 |
Golara M,Benedict A, Jones C. Randhawa M, Poston L, Shennan AH. Inflationary oscillometry provides accurate measurement of blood pressure in pre-eclampsia Br J Obstet Gynaecol 2002;109:1143-1147 |
49 |
Cuckson AC, Jones CR, Randhawa M, Vincent T, Poston L, Shennan AH. Oscillometry detected on inflation can accurately reflect mercury sphygmomanometry J Hypertension 2002;20 (suppl 4):S234 (Abstract) |
50 |
Germanó G, Muscolo M, Angotti S, Bravo S, Codispoti P, Federico L, Damiani S. Evaluation of a New Ambulatory Blood Pressure Device. Am J Hypertens 1998;11:1486-1491 |
51 |
Anwar YA, Giacco S, McCabe EJ, Tendler BE, White WB. Evaluation of the efficacy of the Omron HEM-737 Intellisense device for use on adults according to the recommendations of the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1998;3:261-26 |
52 |
Mee F, Atkins N, O’Brien E. Evaluation of the Shiller BR-102 ambulatory blood pressure system according to the protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1999;4:35-43 |
53 |
Mee F, Atkins N, O’Brien E. Evaluation of the Shiller BR-102 ambulatory blood pressure system according to the protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1999;4:35-43 |
55 |
Walma EP, van Dooren C, van der Does E, Prins A, Mulder P, Hoes AW. Accuracy of an oscillometric automatic blood pressure device: the Omron HEM403C. J Human Hypertens 1995;9:169-174. |
56 |
Naschitz JE, Loewenstein L, Lewis R, Keren D, Gaitini L, Tamir A, Yeshurun D. Accuracy of the OMRON M4 automatic blood pressure measuring device. J Hum Hypertens 2000;14:423-427 |
57 |
Artigao M, Llavador J, Rubio M, Torres C, Lopez J, Sanchis C, Vidal A, Puras A. Evaluation of two devices for self-measurement of blood pressure according to the British Hypertension Society protocol: The Omron M4 and MX2. Am J Hypertens 1998;16:S276 (Abstract) |
58 |
Rithalia SVS, Edwards D. Comparison of oscillometric and intra-arterial blood pressure and pulse measurement. J Med Engin Tech 1994;18:179-181. |
59 |
O'Brien E, Mee F, Atkins N. An accurate automated device for home blood pressure measurement at last! The Omron HEM-705CP. J Hypertens 1994;12:1317-1318 |
60 |
Artigao LM, Llavador JJ, Puras A, Lopex-Abril J, Rubio MM, Torres C, Vidal A, Sanchi C, Division JA, Naharro F, Caldevilla D, Fuentes G. Evaluation and validation of Omron Hem 705 CP and Hem 706/711 monitors for self-measurement of blood pressure. 2000.[dabl® Educational Trust regrets that full details are missing] |
61 |
Mufunda J, Sparks B, Chifomba J, Dakwa C, Matenga JA, Adams JA, et al. Comparison of the Omron HEM-713C automated blood pressure monitor with a standard auscultatory method using a mercury manometer. Central Afr J Med 1996;42:230-232 |
62 |
Furusawa EA, Ruiz MFO, Saito MI, Koch VH Evaluation of the Omron 705-CP blood pressure monitor for use in children and adolescents J Hypertens 2004;22 (suppl 1):S140 (Abstract P510) |
63 |
Foster C, McKinlay JM, Cruickshank JM, Coats AJS. Accuracy of the Omron HEM 706 portable monitor for home measurement of blood pressure. J Human Hypertens 1994;8:661-664. |
64 |
Córdoba R, Fuertes MI, Alvarez A, Molina L, Solans R, Melero I. Validation of a self-monitoring blood pressure monitor: The Omron-HM 722C. Aten Primaria 1997;20:247-250 |
65 |
Bortolotto LA, Henry O, Hanon O, Sikias P, Mourad J-J, Girerd X. Validation of two devices for self-measurement of blood pressure by elderly patients according to the revised British Hypertension Society protocol: the Omron HEM-722C and HEM-735C. Blood Press Monit 1999;4:21-25 |
66 |
Beaubien ER, Card CM, Card SE, Biem HJ, Wilson TW. Accuracy of the Dinamap 1846 XT automated blood pressure monitor J Hum Hypertens 2002;16:647-652 |
67 |
Altunkan S, Iliman N, Altunkan E. Validation of the Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure measuring devices in adults according to the International Protocol Blood Pressure Monitoring 2007;12:119-125 |
68 |
Dieterle T, Battegay E, Bucheli B, Martina B. Accuracy and 'range of uncertainty' of oscillometric blood pressure monitors around the upper arm and wrist. Blood Press Moni 1998;3:339-346 |
69 |
de Greeff A, Beg Z, Gangji Z, Dorney E, Shennan AH. Accuracy of inflationary versus deflationary oscillometry in pregnancy and preeclampsia: OMRON-MIT versus OMRON-M7. Blood Press Monit 2009;14:37-40 |
70 |
O'Brien E, Atkins N, Murphy A, Lyons S. Validation of the Rossmax blood pressure measuring monitor according to the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults Blood Press Monit 2003;8:261-265 |
71 |
Jones CR, Taylor K, Chowienczyk P, Poston L, Shennan A. A validation of the Mobil O Graph (version 12) ambulatory blood pressure monitor. Blood Press Monit 2000;5:233-238 |
72 |
Octavio JA, Leal N, Contreras J, Octavio B, Cressa L, Valecillos I, Castellanos D, Fabiano D. Accuracy of Jotatec®-P24B device for teletransmitted ambulatory blood pressure monitoring J Hypertens 2008;26 (suppl. 1):S493 (Abstract PS34/MON/60) |
73 |
Bonso E, Saladini F, Zanier A, Benetti E, Dorigatti F, Palatini P. Accuracy of a single rigid conical cuff with standard-size bladder coupled to an automatic oscillometric device over a wide range of arm circumferences. Hypertens Res. 2010;33(11):1186-91. |
74 |
Nolly H, Romero M, Nolly A, Silva M, Nolly M. Clinical evaluation of the Braun BP 3000 Easy Click according to the ANSI/AAMI SP 10:2002 Blood Pressure Monitoring 2006;11:91-95 |
75 |
Nolly H, Romero M, Nolly A, Silva M, Nolly M. Clinical evaluation of the Braun BP 3000 Easy Click according to the ANSI/AAMI SP 10:2002. Blood Press Monit 2006;11:91-95 |
76 |
Iliman N, Altunkan S, Kayaturk N, Altunkan E. Validation of the Braun BP 3550 wrist blood pressure measuring device with a position sensor and an EasyClick cuff according to the International Protocol in adults. Blood Press Monit 2007;12:45-49 |
77 |
Wessig K, Hollinger S, Schmalzhaf I, Lenz T. Clinical evaluation of the efficacy of the Braun PrecisionSensor oscillometric wrist
blood pressure monitor for use on adults versus auscultation as defined by ANSI/AAMI SP 10-1992 Blood Pres Monit 2000;5:239-245
|
78 |
Nolly H, Osso P, Nolly M. Home blood pressure monitoring Intercontinental Cardiology 2001;10:111-114 |
79 |
Nolly H, Osso A, Romero M, Nolly M. The reproducibility of home- self measurements using the Braun PrecisionSensor BP 2550 Monitor J Hypertens 2001;19 (suppl 2):S26 |
80 |
Sandevik E, Dyrdal A. Evaluation of the Braun BP2550 PrecisionSensor wrist blood pressure monitor in clinical practice [dabl® Educational Trust regrets that full details are missing] (Abstract) |
81 |
Nolly H, Romero M, Nolly A, Osso P, Reinoso O, Nolly M. Home blood pressure measurement: validation of the Braun BP 2550 (UG) monitor according to the ESH International Protocol Blood Press Monit 2004;9:53-58 |
82 |
Nolly H, Romero M, Nolly A, Nolly M. Home blood pressure (BP) measurement using B.P. Braun Vitalscan monitor. J Hypertens 200;18(suppl 4):S119 (Abstract) |