General Papers on SBPM
|
O'Brien E.
What's
new in blood pressure measurement? A workshop of the European
Society of Hypertension.
Blood Pressure Monitoring 2004, 9:281–282 (pdf)
Proceedings of the European Society of Hypertension
Working Group on Blood Pressure Monitoring Scientific Workshop:
What's New in Blood Pressure Measurement? Paris, France on
17th June 2004.
|
|
George Stergiou, Thomas Mengden, Paul L Padfield,
Gianfranco Parati, Eoin O'Brien and working group on blood
pressure monitoring of the European Society of Hypertension
Self
monitoring of blood pressure at home.
BMJ 2004, 329:870-871(pdf)
|
|
Ohkubo T, Asayama K, Kikuya M, Metoki H, Obara
T, Saito S, Hoshi H, Hashimoto J, Totsune K, Satoh H, Imai
Y. Prediction of ischaemic and haemorrhagic stroke by self-measured
blood pressure at home: the Ohasama study. Blood Press Monit 2004;9:315-320
|
|
Mengden T, Ewald S, Kaufmann S, vor dem Esche
J, Uen S, Vetter H. Telemonitoring of blood pressure self measurement
in the OLMETEL study. Blood Press Monit 2004;9:321-325
|
|
Chiolero A, Gervasoni JP, Rwebogora A, Mkamba
M, Waeber B, Paccaud F, Burnier M, Bovet P. Discordant prevalence
of hypertension using two different automated blood pressure
measurement devices: a population-based study in Dar es Salaam
(Tanzania) Blood Press Monit 2004;9:59-64
|
|
O'Brien E, Atkins N, Mee F, O'Malley K.� Evaluation
of blood pressure measuring devices.� Clin & Exper Hypertension
1993;15(6):1087-1097
|
|
Ng K-G, Small CF. Survey of Automated Noninvasive
Blood Pressure Monitors. J Clin Eng 1994;19:452-475
|
|
Rithalia SVS, Edwards D. Comparison of oscillometric
and intra-arterial blood pressure and pulse measurement.
J
Med Engin Tech
1994;18:179-181.
|
|
Ng K-G. Automated Noninvasive blood pressure measurement.
Intensive Care World 1995;12:89-102
|
|
Livi R, Teghini L, Cagnoni S, Scarpelli PT. Simultaneous
and Sequential Same-Arm Measurements in the Validation Studies
of Automated Blood Pressure Measuring Devices. Am J Hypertens
1996;9:1228-1231
|
|
Dieterle T, Battegay E, Bucheli B, Martina B.
Accuracy and 'range of uncertainity' of oscillometric blood
pressure monitors around the upper arm and wrist.
Blood Pressure
Monitoring
1998;3:339-346
|
|
Rogers P, Burke V, Stroud P, Puddey IB. Comparison
of oscillometric blood pressure measurements at the wrist with
an upper-arm auscultatory mercury sphygmomanometer.
Clin
& Exper Pharmacol & Pharmacol
1999;26:477-481
|
|
Natarajan P, Shennan AH, Penny
J, Halligan AW, de Swiet M, Anthony J. Comparison of auscultatory
and oscillometric automated blood pressure monitors in the setting
of preeclampsia. Am J Obstet� Gynecol 1999;181:1203-1210
|
|
Kikuya M, Chonan K, Imai Y, Goto E, Ishii M on
behalf of the Research Group. Accuracy and reliability of wrist-cuff
devices for self-measurement of blood pressure.
J Hypertens
2002;20:629-638
|
|
Longo D, Toffanin G, Bertolo O, Frezza P, Palatini
P. Blood pressure measured at the wrist overestimates blood
pressure measured at the arm. J Hypertens 2002;
20
(suppl 4)
:S230
|
|
Types of Sphygmomanometers for Self Measurement |
Wrist devices |
Mourad A, Gillies A, Carney S. Inaccuracy of wrist-cuff oscillometric blood pressure devices: an arm position artefact? Blood Press Monit 2005; 10 :67-71 |
|
Palatini P, Longo D, Toffanin G, Bertolo O, Zaetta V, Pessina AC, Wrist blood pressure overestimates blood pressure measured at the upper arm. Blood Press Monit 2004;9:77-81 |
|
Specific SBPM Sphygmomanometers
|
Upper Arm Devices for Self Measurement
|
A & D Takeda |
|
|
UA-631 (UA-779) |
|
Longo D, Bertolo O, Toffanin G, Frezza P, Palatini P. Validation of the A & D UA-631 (UA-779 Life Source) device for self-measurement of blood pressure and relationship between its performance and large artery compliance. Blood Press Monit 2002;7:243-248 |
|
|
UA-704 |
|
Koblava ZD, Kotovskaya YV, Rodionov E Validation of semi-automatic device UA-704 for self-measurement of blood pressure Blood Press Monit 2005; 10 :223-225 |
|
|
|
UA-705 |
|
Verdecchia P, Angeli F, Gattobigio R, Angeli E, Pede S, Pittavini L. Validation of the A&D UA-705 device for self-measurement of blood pressure according to the British Hypertension Society protocol Blood Press Monit 2006; 11 :223-227 |
|
|
UA-767 |
|
Davis
P D, Dennis J L and Railton R.
Evaluation
of the A&D UA-767
and Welch Allyn Spot Vital Signs noninvasive blood pressure
monitors using a blood pressure simulator.
Journal
of Human Hypertension
2005;3:197-203
|
|
|
|
Rogoza AN, Pavlova TS, Sergeeva MV. Validation of A & D UA-767 device for the self-measurement of blood pressure. Blood Press Monit 2000;5:227-231 |
|
|
Yarrows SA, Brook RD. Measurement Variation Among
12 Electronic Home Blood Pressure Monitors. Am J Hypertens
2000;13:276-282
|
|
|
UA-774 (UA-767 Plus) |
|
Verdecchia P, Angeli F, Poeta F, Reboldi GP, Borgioni C, Pittavini L, Porcellati C. Validation of the A&D UA774 (UA-767Plus) device for self measurement of blood pressure. Blood Press Monit 2004;9:225-229 |
|
|
|
Kobalava ZD, Kotovskaia IV, Rusakova OS, Babaeva LA. Validation of UA-767 Plus device for self measurement of blood pressure Clin. Pharmacol. Ther. 2003;12 (2):70-72 |
|
|
|
UA-787 |
|
Longo D, Toffanin G, Garbelotto R, Zaetta V, Businaro L, Palatini P. Performance of the UA-787 oscillometric blood pressure monitor according to the European Society of Hypertension Protocol. Blood Press Monit 2003;8:91-95 |
|
Artsana 610 |
Pini C, Pastori M, Baccheschi J, Omboni S, Parati G. Validation of the Artsana CSI 610 automated blood pressure monitor in adults according to the International Protocol of the European Society of Hypertension Blood Pressure Monitoring 2007;12:179–184 |
|
Assure 605
|
Yarrows SA, Brook RD. Measurement Variation Among
12 Electronic Home Blood Pressure Monitors. Am J Hypertens
2000;13:276-282
|
|
Bard Medical Sentron
|
Zambanini A, Cox A, McIntosh C, Feher M. A comparison
of semi-automated BP monitors in diabetic patients.
J Hypertens
2000;18(suppl 4):S121 (Abstract)
|
|
Boso carat
|
Eckbert S, Kemper B, Horstkotte D. How accurate
are devices for blood pressure self-measurement at the upper
arm, wrist and finger in young normotensive patients and elder
hypertensive patients? J Hypertens 2001;19(suppl 2):S24
(Abstract)
|
|
Braun Vital-Scan (BVS) BP Monitor
|
Nolly H, Romero M, Nolly A, Nolly M. Home blood pressure (BP) measurement using B.P. Braun Vitalscan monitor. J Hypertens 200;18(supp 4):S119 |
|
Colson MAM BP3AA1-2
|
Pereira T, Maldonado J. Valida çã o de aparelho de medi çã o autom á tica da press ã o arterial Colson MAM BP3AA1-2 segundo o Protocolo da Sociedade Europeia de Hipertens ã o Rev Port Cardiol 2005;24(11):1341-1351 |
|
Health & Life |
|
HL-868BA |
|
Chen HE, Cui Y, Sheng CS, Li LH, Li Y, Wang JG. Validation of the Health & Life HL868BA blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol. Blood Press Monit 2008;13:305-308 |
|
|
Chen HE, Cui Y, Sheng CS, Li LH, Li Y, Wang JG. Validation of the Health & Life BSP-4007 blood pressure monitor in adult Chinese according to the European Society of Hypertension International Protocol J Hypertens 2008;26 (suppl. 1):S496 (Abstract PS35/TUE/07) |
|
|
HL-888HA |
|
de Greeff A, Shennan AH. The Health and Life device (HL888HA): accuracy assessment in an adult population according to the British Hypertension Society protocol Blood Pressure Monitoring 2007;12:107-111 |
|
|
Tholl U, Assouline E. Validation of the automatic arm type non-invasive blood pressure monitor HL-888HA J Hypertens 2006; 24 (suppl 4):S285 Abstract P14.496 |
|
IEM Stabil-O-Graph |
Westhoff T, Schmidt S, Zidek W, van der Giet M. Validation of the Stabil-O-Graph® Blood Pressure Self-Measurement Device According to BHS and ESH Criteria Journal of Hypertension 2006; 25 (suppl 2) S161 |
|
Oever GHM ten, Bour J, Cortez-Campeao, van de Kerkhof HM, Wolf A. Validation of Stabil-O-Graph device for self-measurement of blood pressure monitoring according to the British Hypertension Society (BHS) standards and protocol [dabl® Educational Trust regrets that full details are missing] (Abstract) |
|
Healthcheck CX-5
|
O'Brien E, Mee F, Atkins N, O'Malley
K.� Inaccuracy of seven popular sphygmomanometers for home-measurement
of blood pressure.� J Hypertens 1990;8:621-634
|
|
Lumoscope1085M
|
Yarrows SA, Brook RD. Measurement Variation Among
12 Electronic Home Blood Pressure Monitors. Am J Hypertens
2000;13:276-282
|
|
Microlife |
|
BP 3AC1-1 |
|
de Greeff A, Reggiori F, Anthony J, Shennan A. The Microlife 3AC1: An accurate blood pressure measurement device in pregnancy and pre-eclampsia J Hypertens 2006; 24 (suppl 4):S279 Abstract P14.475 |
|
|
|
Topouchian JA, El Assad MA, Orobinskaia LV, El Feghali RN, Asmar RG. Validation of two devices for self-measurement of brachial blood pressure according to the International Protocol of the European Society of Hypertension: the Seinex SE-9400 and the Microlife BP 3AC1-1 Blood Press Monit 2005;10:325-331 |
|
|
|
El Assad M, Topouchian J, Labaki G, Asmar R. Validation of
the Microlife BP 3AC1-1 blood pressure measuring device according
to the International Protocol of the European Society of Hypertension
|
|
|
El Assad M, Topouchian J, Labaki G, Asmar R. Validation of
the Microlife BP 3AC1-1 device for blood pressure measurement
according to the International Validation Protocol J Hypertens 2003;[dabl® Educational Trust regrets that full details are missing]
|
|
|
El Assad M, Topouchian J, Labaki G, Asmar R. Validation of
the Microlife BP 3AC1-1 device for blood pressure measurement
according to the International Validation Protocol American Journal of Hypertension 2003;16:47A
|
|
|
|
BP 3AC1-1 PC |
|
Thompson AM, Eguchi K, Reznik ME, Shah SS, Pickering TG. Validation of an oscillometric home blood pressure monitor in an end-stage renal disease population and the effect of arterial stiffness on its accuracy Blood Pressure Monitoring 2007;12:227-232 |
|
|
BP 3BTO-A |
|
Reinders
A, Cuckson A C, Lee J T M, Shennan A H.
An
accurate automated blood pressure device for use in pregnancy
and pre-eclampsia: the Microlife 3BTO-A.
BJOG 2005;112:1-6
|
|
|
|
Cuckson A, Reinders A, Shennan A. Validation of the Microlife BP 3BTO-A oscillometric blood pressure monitoring device according to a modified British Hypertension Society protocol |
|
|
|
BP A100 Plus |
|
Belghazi J, El Feghali RN, Moussalem T, Rejdych M, Asmar RG. Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension Vascular Health and Risk Management 2007;3(4):389-400 |
|
Nais EW284
|
De Luisignam S, Thiru K. Meredith K, Majeed A,
Johnson P. Measuring blood pressure at the wrist: more comfortable
for patients and more convenient for doctors? Public Health
2000;114:165-168
|
|
Nissei
|
|
Analogue |
|
O'Brien E, Mee F, Atkins N, O'Malley K.� Inaccuracy of seven popular sphygmomanometers for home-measurement of blood pressure.� J Hypertens 1990;8:621-634 |
|
|
|
DS-175 |
|
O'Brien E, Mee F, Atkins N, Thomas M. Evaluation of three devices for self-measurement of blood pressure: according to the revised British Hypertension Society Protocol: the Omron HEM-705CP, Phillips HP5332, and Nissei DS-175. Blood Press Monit �1996;1:55-61 |
|
NuMed Stabil-O-Graph |
Oever GHM ten, Bour J, Cortez-Campeao, van de Kerkhof HM, Wolf A. Validation of Stabil-O-Graph device for self-measurement of blood pressure monitoring according to the British Hypertension Society (BHS) standards and protocol [dabl® Educational Trust regrets that full details are missing] (Abstract) |
|
Omron |
|
705IT (HEM 759-E) |
|
Stergiou GS, Yiannes NG, Rarra VC. Validation of the Omron 705 IT oscillometric device for home blood pressure measurement in children and adolescents: The Arsakion School Study Blood Pressure Monitoring 2006,11:229-234 |
|
|
|
Coleman A, Freeman P, Steel S, Shennan A. Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol Blood Pressure Monitoring 2006;11:27-32 |
|
|
|
El Assad M, Topouchian J, Asmar R. Evaluation of two devices for self-measurement of blood pressure according to the international protocol: the Omron M5-I and the Omron 705IT Blood Press Monit 2003;8(3):127-133 |
|
|
|
907 (HEM-907/HEM-907-UK) |
|
Amoore JN, Vacher E, Murray IC, Mieke S, King ST, Smith FE, Murray A. Can a
simulator that regenerates physiological waveforms evaluate oscillometric
non-invasive blood pressure devices? Blood Pressure Monitoring 2006;11:63-67
|
|
|
|
El Assaad MA, Topouchian JA, Darne BM, Asmar RG Validation of the Omron HEM-907 device for blood pressure measurement. Blood Press Monit 2002;7:237-241 |
|
|
Arita M, Satani Y, Ueno Y, Nishio I. Comparison with casual and automatically measured blood pressure in hypertensive patients - using new developed HEM 907. J Hypertens ?(Abstract) |
|
|
HEM-703CP |
|
Rithalia SVS, Edwards D. Comparison of oscillometric and intra-arterial blood pressure and pulse measurement. J Med Engin Tech 1994;18:179-181. |
|
|
HEM-705CP |
|
Graves JW, Grossardt BR, Gullerud RE, Bailey KR, Feldstein J. The trained observer better predicts daytime ABPM diastolic blood pressure in hypertensive patients than does an automated (Omron) device Blood Pressure Monitoring 2006;11:53-58 |
|
|
|
O'Brien E, Mee F, Atkins N. An accurate automated device for home blood pressure measurement at last!� The Omron HEM-705CP. J Hypertens 1994;12:1317-1318 |
|
|
O'Brien E, Mee F, Atkins N, Thomas M. Evaluation of three devices for self-measurement of blood pressure: according to the revised British Hypertension Society Protocol: the Omron HEM-705CP, Phillips HP5332, and Nissei DS-175. Blood Pressure Monitoring �1996;1:55-61 |
|
|
Artigao LM, Llavador JJ, Puras A, Lopex-Abril J, Rubio MM, Torres C, Vidal A, Sanchi C, Divison JA, Naharro F, Caldevilla D, Fuentes G. Evaluation and validation of Omron Hem 705 CP and Hem 706/711 monitors for self-measurement of blood pressure. 2000.[dabl® Educational Trust regrets that full details are missing.] |
|
|
Markandu ND, Duneclift S, Carney C, Arnold A. Further evaluation of a semi-automatic blood pressure machine (OMRON HEM-705CP) with an assessment of its use in a hypertension unit and preliminary findings in a general medical ward. J Hum Hypertens 2001;15:357-358 (Abstract) |
|
|
Czarkowski M, Malgorzata M, Zebrowski M, Baran A, Rozanowski K. Venous blood redistribution alters the accuracy of Omron HEM-705CP. Blood Pressure Monitoring 2004;9:9-12 |
|
|
Furusawa EA, Ruiz MFO, Saito MI, Koch VH Evaluation of the Omron 705-CP blood pressure monitor for use in children and adolescents J Hypertens 2004;22 (suppl 1):S140 (Abstract P510) |
|
|
Czarkowski M, Mikulska M, Zebrowski M, Baran A, Rozanowski K. Venous blood redistribution alters the accuracy of the Omron HEM-705CP Blood Press Monit 2004;9:9-12 |
|
|
Czarkowski M, Mikulska M, Zebrowski M, Baran A, Rozanowski K. Tilt test alters accuracy of the oscillometric device Omron HEM-705CP J Human Hypertens 2004;18:293-294 (Letter) |
|
|
El Assad M, Topouchian J, Labaki G, Asmar R. Validation of the Omron 705 IT device for blood pressure measurement according to the international validation protocol American Journal of Hypertension 2003;16:47A |
|
|
|
HEM-706 |
|
Foster C, McKinlay JM, Cruickshank JM, Coats AJS. Accuracy of the Omron HEM 706 portable monitor for home measurement of blood pressure. J Human Hypertens 1994;8:661-664. |
|
|
|
Artigao LM, Llavador JJ, Puras A, Lopex-Abril J, Rubio MM, Torres C, Vidal A, Sanchi C, Divison JA, Naharro F, Caldevilla D, Fuentes G. Evaluation and validation of Omron Hem 705 CP and Hem 706/711 monitors for self-measurement of blood pressure. 2000. [dabl® Educational Trust regrets that full details are missing.] |
|
|
HEM-711 |
|
Yarrows SA, Brook RD. Measurement Variation Among 12 Electronic Home Blood Pressure Monitors. Am J Hypertens 2000;13:276-282 |
|
|
HEM-713C |
|
Mufunda J, Sparks B, Chifomba J, Dakwa C, Matenga
JA, Adams JA, et al. Comparison of the Omron HEM-713C
automated blood pressure monitor with a standard auscultatory
method using a mercury manometer. Central Afr J Med 1996;42:230-232
|
|
|
Afzali B, Abbs IC, Goldsmith D. Automated oscillometric
BP devices can significantly overestimate true BP in renal transplant
patients. [dabl® Educational Trust regrets that
full details are missing.] (Abstract)
|
|
|
HEM-722C |
|
C�rdoba R, Fuertes MI, Alvarez A, Molina L, Solans
R, Melero I. Validation of a self-monitoring blood pressure
monitor: The Omron-HM 722C. Aten Primaria 1997;20:247-250
|
|
|
Bortolotto LA, Henry O, Hanon O, Sikias P, Mourad
J-J, Girerd X. Validation of two devices for self-measurement
of blood pressure by elderly patients according to the revised
British Hypertension Society protocol: the Omron HEM-722C and
HEM-735C. Blood Press Monit 1999;4:21-25
|
|
|
Yarrows SA, Brook RD. Measurement Variation Among
12 Electronic Home Blood Pressure Monitors.
Am J Hypertens
2000;13:276-282
|
|
|
HEM-725 CIC |
|
De Luisignam S, Thiru K. Meredith
K, Majeed A, Johnson P. Measuring blood pressure at the wrist:
more comfortable for patients and more convenient for doctors?
Public Health 2000;114:165-168
|
|
|
Yarrows SA, Brook RD. Measurement Variation Among 12 Electronic Home Blood Pressure Monitors. Am J Hypertens 2000;13:276-282 |
|
|
HEM-735C |
|
Bortolotto LA, Henry O, Hanon O, Sikias P, Mourad
J-J, Girerd X. Validation of two devices for self-measurement
of blood pressure by elderly patients according to the revised
British Hypertension Society protocol: the Omron HEM-722C and
HEM-735C. Blood Press Monit 1999;4:21-25
|
|
|
HEM-737 Intellisense |
|
Anwar YA, Giacco S, McCabe EJ, Tendler BE, White WB. Evaluation of the efficacy of the Omron HEM-737 Intellisense device for use on adults according to the recommendations of the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1998;3:261-26 |
|
|
HEM-747-IC-N |
|
Imai Y, Kikuya M, Matsubara M, Araki T, Michimata M, Ohkubo T, Houzawa A, Yamamoto N, Miyawaki Y. Device for self-blood pressure measurements which can monitor blood pressure during sleep. J Hypertens [dabl® Educational Trust regrets that full details are missing.](suppl ?):S25 (Abstract) |
|
|
HEM 780REL |
|
Viera AJ, Hinderliter AL. Validation of the HEM-780REL with easy wrap cuff for self-measurement of blood pressure according to the European Society of Hypertension International Protocol Blood Pressure Monitoring 2007;12:335-338 |
|
|
|
M1 Plus (HEM-4011C-E) |
|
Belghazi J, El Feghali RN, Moussalem T, Rejdych M, Asmar RG. Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension Vascular Health and Risk Management 2007;3(4):389-400 |
|
|
|
M3 Intellisense (HEM-7051-E) |
|
Topouchian J, El Feghali R, El Assaad H, Pannier B, Asmar R. Validation of the Omron HEM-7051-E Device for Blood Pressure Measurement According to the International Validation Protocol Journal of Hypertension 2006; 24 (suppl 4) S278 |
|
|
|
M4 |
|
Artigao M, Llavador J, Rubio M, Torres C, Lopez
J, Sanchis C, Vidal A, Puras A. Evaluation of two devices for
self-measurement of blood pressure according to the British
Hypertension Society protocol: The Omron M4 and MX2.
Am J
Hypertens
1998;16:S276 (Abstract only)
|
|
|
Naschitz JE, Loewenstein L, Lewis R, Keren D, Gaitini L, Tamir A, Yeshurun D. Accuracy of the OMRON M4 automatic blood pressure measuring device. J Hum Hypertens 2000;14:423-427 |
|
|
M5-I (HEM-757-E) |
|
Omboni S, Riva I, Giglio I, Caldara G, Groppelli A, Parati G. Validation of the Omron M5-I, R5-I and HEM-907 automated blood pressure monitors in elderly individuals according to the International Protocol of the European Society of Hypertension Blood Pressure Monitoring 2007;12:233-242 |
|
|
|
El Assad M, Topouchian J, Labaki G, Asmar R. Validation of the Omron M5-I device for blood pressure measurement according to the international validation protocol |
|
|
El Assad M, Topouchian J, Asmar R. Evaluation of two devices for self-measurement of blood pressure according to the international protocol: the Omron M5-I and the Omron 705IT Blood Press Monit 2003;8(3):127-133 |
|
|
M6 (HEM-7001-E) |
|
Altunkan S, Iliman N, Kayaturk N, Altunkan E. Validation of the Omron M6 (HEM-7001-E) upper-arm blood pressure measuring device according to the International Protocol in adults and obese adults Blood Pressure Monitoring 2007;12:219-225 |
|
|
|
Altunkan S, Iliman N, Kayaturk N, Altunkan E. Validation of the Omron M6 upper arm blood pressure measuring device according to the International Protocol in adults. J Hypertens 2006; xx (suppl x):213 (Abstract PO1-43) |
|
|
|
Topouchian JA, El Assaad MA, Orobinskaia LV, El Feghali RN, Asmar RG. Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT) Blood Press Monit 2006; 11 :165-171 |
|
|
|
M6 Comfort (HEM-7000-E) |
|
Belghazi J, El Feghali RN, Moussalem T, Rejdych M, Asmar RG. Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension Vascular Health and Risk Management 2007;3(4):389-400 |
|
|
|
M7 (HEM-780E) |
|
El Feghali RN, Topouchian JA, Pannier BM, El Assaad HA, Asmar RG. Validation of the Omron M7 (HEM-780-E) blood pressure measuring device in a population requiring large cuff use according to the International Protocol of the European Society of Hypertension Blood Pressure Monitoring 2007;12:173-178 |
|
|
|
Coleman A, Steel S, Freeman P, de Greeff A, Shennan A. Validation of the Omron M7 (HEM-780-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol. Blood Pressure Monitoring 2008;13:49-54 |
|
|
|
MX2 |
|
Artigao M, Llavador J, Rubio M, Torres C, Lopez
J, Sanchis C, Vidal A, Puras A. Evaluation of two devices for
self-measurement of blood pressure according to the British
Hypertension Society protocol: The Omron M4 and MX2.
Am J
Hypertens
1998;16:S276 (Abstract only)
|
|
|
MX3 Plus |
|
Coleman A, Freeman P, Steel S, Shennan A Validation of the Omron MX3 Plus oscillometric blood pressure monitoring device according to the European Society of Hypertension international protocol Blood Press Monit 2005; 10 :165-168 |
|
Oregon Scientific BPW810 |
Li L, Wang Hui W, Jin Lan W. Validation of the Oregon Scientific BPW810 blood pressure monitor according to the European Society of Hypertension Protocol. Blood Press Monit 2006;11:343-347 |
|
Phillips HP5308, HP5306/B, HP5332 |
O'Brien E, Mee F, Atkins N, O'Malley K.� Inaccuracy
of seven popular sphygmomanometers for home-measurement of blood
pressure.� J Hypertens 1990;8:621-634
|
|
O'Brien E, Mee F, Atkins N, Thomas M. Evaluation
of three devices for self-measurement of blood pressure: according
to the revised British Hypertension Society Protocol: the Omron
HEM-705CP, Phillips HP5332, and Nissei DS-175.
Blood Pressure
Monitoring
�1996;1:55-61
|
|
Rossmax |
O'Brien E, Atkins N, Murphy A, Lyons S. Validation of the Rossmax blood pressure measuring monitor according to the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults Blood Press Monit 2003;8:261-265 |
|
Rossmax ME 701 Series |
de Greeff A, Shennan AH. The Rossmax (ME 701 series) upper arm device: accuracy assessment in an adult population according to the International and the British Hypertension Society protocols Blood Pressure Monitoring 2008;13:43-48 |
|
Seinex SE-9400 |
Topouchian JA, El Assad MA, Orobinskaia LV, El Feghali RN, Asmar RG. Validation of two devices for self-measurement of brachial blood pressure according to the International Protocol of the European Society of Hypertension: the Seinex SE-9400 and the Microlife BP 3AC1-1 Blood Press Monit 2005;10:325-331 |
|
Sensacare SAA-102 |
Zaetta V, Daniele L, Perkovic D, Prattico F, Barisa M, Perfetti P, Gabrieli A, Buonocore F, Winnicki M. Validation of the SAA-102 home blood pressure monitor according to the protocols of the European Society of Hypertension, the Association for the Advancement of Medical Instrumentation and the British Society of Hypertension Blood Pressure Monitoring 2007;12:363-368 |
|
Spengler KP7500D |
Belghazi J, El Feghali RN, Moussalem T, Rejdych M, Asmar RG. Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension Vascular Health and Risk Management 2007;3(4):389-400 |
|
Sunbeam7623
|
Yarrows SA, Brook RD. Measurement Variation Among
12 Electronic Home Blood Pressure Monitors. Am J Hypertens
2000;13:276-282
|
|
Systema Dr MI-150
|
O'Brien E, Mee F, Atkins N, O'Malley K.� Inaccuracy
of seven popular sphygmomanometers for home-measurement of blood
pressure.� J Hypertens 1990;8:621-634
|
|
Terumo ES-H51
|
Yarrows SA, Brook RD. Measurement Variation Among
12 Electronic Home Blood Pressure Monitors. Am J Hypertens
2000;13:276-282
|
|
Visomat OZ2 |
Dieterle T, Battegay E, Bucheli B, Martina B. Accuracy and ‘range of uncertainity’ of oscillometric blood pressure monitors around the upper arm and wrist. Blood Press Moni 1998;3:339-346 |
|
Welch Allyn Transtelephonic Home Blood Pressure Monitor
|
Herbst T, Anis-Anwar Y, Giacco S, White W. Evaluation
of the overall system precision of the Welch-Allyn transtelephonic
home blood pressure monitor in adults with Parkinson's disease.
Blood Press Monit 2002;7:285-288
|
|
Wrist Devices for Self Measurement
|
A&D UB-511 |
Angeli F, Sardone M, Angeli E, Repaci S, Gattobigio R, Verdecchia P. Validation of the A&D wrist-cuff UB-511 (UB-512) device for self-measurement of blood pressure. Blood Press Monit 2006; 11 :349-354 |
|
Assure A30/BD-W 20
|
Yarrows SA, Brook RD. Measurement Variation Among
12 Electronic Home Blood Pressure Monitors. Am J Hypertens
2000;13:276-282
|
|
Boso-Mediwatch |
(See Specific Ambulatory Sphygmomanometers for Continuous Measurement for later references) |
Rogers P, Burke V, Stroud P, Puddey IB. Comparison of oscillometric blood pressure measurements at the wrist with an upper-arm auscultatory mercury sphygmomanometer Clin & Exper Pharmacol & Pharmacol 1999;26:477-481 |
|
Braun |
|
Braun BP 3550 |
|
Iliman N, Altunkan S, Kayaturk N, Altunkan E. Validation of the Braun BP 3550 wrist blood pressure measuring device with a position sensor and an EasyClick cuff according to the International Protocol in adults. Blood Pressure Monitoring 2007;12:45-49 |
|
|
|
BP 3000 |
|
Nolly H, Romero M, Nolly A, Silva M, Nolly M. Clinical evaluation of the Braun BP 3000 Easy Click according to the ANSI/AAMI SP 10:2002 Blood Pressure Monitoring 2006;11:91-95 |
|
|
|
PrecisionSensor 2000 |
|
Wessig K, Hollinger S, Schmalzhaf I, Lenz T. Clinical evaluation of the efficacy of the Braun PrecisionSensor oscillometric wrist blood pressure monitor for use on adults versus auscultation as deficned by ANSI/AAMI SP 10-1992. Blood Pres Monit 2000;5:239-245 |
|
|
|
Grassi G, Foglia G, Dell'Oro R, Seravalle G, Boari G, Mancia G. Reproducility of home blood pressure monitoring by a new oscillometric wrist automatic device. J Hypertens 2001; 19 (suppl 2):S22 |
|
|
|
Nolly H, Osso A, Romero M, Nolly M. The reproducibility of home-self measurements using the Braun PrecisionSensor BP 2550 Monitor. J Hypertens 2001;19 (suppl 2):S26 (Abstract) |
|
|
|
Nolly H, Osso A, Romero M, Nolly A. Home Blood Pressure Monitoring: Validation of the Braun BP 2550 Equipment. AJH �2002;15:83A (Abstract) |
|
|
|
Nolly H, Osso A, Nolly A, Romero M, Nolly M. Usefullness of home blood pressure determination: validation of� the Braun BP 2550 Equipment. J Hypertens 2002;20 (suppl 4):S238 |
|
|
|
Uen S, Weisser B, Wieneke P, Vetter H, Mengden T. Evaluation of the Performance of a Wrist Blood Pressure Measuring Device With a Position Sensor Compared to Ambualatory 24-Hour Blood Pressure Measurements. Am J Hypertens 2002;15:787-792 |
|
|
PrecisionSensor BP2550 and BP2550 (UG) |
|
Sandevik E, Dyrdal A. Evaluation of the Braun BP2550 PrecisionSensor wrist blood pressure monitor in clinical practice[dabl® Educational Trust regrets that full details are missing] |
|
|
Nolly H, Raposo I, Nolly A. Accuracy of the Braun BP2550 (UG) device in home blood pressure measurement[dabl® Educational Trust regrets that full details are missing] |
|
|
Nolly H, Romero M, Nolly A, Osso P, Reinoso O, Nolly M. Home blood pressure measurement: validation of the Braun BP 2550 (UG) monitor according to the ESH International Protocol Blood Press Monit 2004;9:53-58 |
|
|
Nolly H, Osso P, Nolly M. Home blood pressure monitoring Intercontinental Cardiology 2001;10:111-114 |
|
|
|
Nolly H, Osso A, Romero M, Nolly M. The reproducibility of home- self measurements using the Braun PrecisionSensor BP 2550 Monitor J Hypertens 2001;19 (suppl 2):S26 |
|
|
Vital Scan |
|
Nolly H, Romero M, Nolly A, Nolly M. Home blood pressure (BP) measurement using B.P. Braun Vitalscan monitor. J Hypertens 200;18(supp 4):S119 (Abstract) |
|
Klock
|
Zweiker R, Schumacher M, Fruhwald
FM, Watzinger N, Klein W. Comparison of wrist blood pressure
measurement with conventional sphygmomanometry at a cardiology
clinic. J Hypertens 2000;18:1013-1018
|
|
Microlife |
|
|
BP 3BU1-5 |
|
Altunkan S, Iliman N, Altunkan E. Validation of the Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure measuring devices in adults according to the International Protocol Blood Pressure Monitoring 2007;12:119-125 |
|
|
BP W200-1 |
|
Palatini P, Dorigatti F, Bonso E, Ragazzo F. Validation of the Microlife BP W200-1 wrist device for blood pressure measurement. Blood Press Monit 2008;13:295-298 |
|
Nais
|
Caridi G, Zoccali C, Enia G. The oscillometric
wrist devices: an estimate of the 'supine' error.
J Human
Hypertens
2002;16:525-526
|
|
Nissei WS-310
|
Altunkan Ş, Yildiz SA. Wrist blood pressure measuring
devices: a comparitive study of the accuracy with a standard
auscultatory method using a mercury sphygmomanometer.
J Hypertens
[dabl® Educational Trust regrets that full
details are missing.](suppl ?):S25 (Abstract)
|
|
Omron
|
|
HEM-601 |
|
Yarrows SA, Brook RD. Measurement Variation Among 12 Electronic Home Blood Pressure Monitors. Am J Hypertens 2000;13:276-282 |
|
|
HEM-605 |
|
Yarrows SA, Brook RD. Measurement Variation Among 12 Electronic Home Blood Pressure Monitors. Am J Hypertens 2000;13:276-282 |
|
|
HEM-608 |
|
Shennan AH, Rushbrook J, Power J, Wright J, Poston L An accurate oscillometric wrist blood pressure monitor: validation of the Omron Rx [HEM-608]. J Human Hypertens 1998;12:794 |
|
|
HEM-637IT |
|
Altunkan Ş, Öztas K, Altunkan E. Validation of the Omron 637IT wrist blood pressure measuring device with a position sensor according to the International Protocol in adults and obese adults Blood Press Monit 2006;11:79-85 |
|
|
|
Altunkan Ş, Altunkan E. Validation of the Omron 637IT wrist blood pressure measuring device with a position sensor according to the International Protocol in the elderly Blood Press Monit 2006;11:97-102 |
|
|
|
Altunkan Ş, Iliman N, Altunkan E Accuracy of the new wrist blood pressure monitor (Omron 637 IT), for blood pressure measurement J Hypertens June 2003;21(Suppl.4):S22 |
|
|
Yarrows SA. Comparison of the Omron HEM-637 wrist monitor to the auscultation method with the wrist position sensor on or disabled Americam Journal of Hypertension 2004;17(No 1) 54:58 |
|
|
Yarrows SA. Comparison of the Omron HEM-637 wrist monitor to the ausculatation method with the wrist position sensor on or disabled Am J Hypertens 2004;17:54-58 |
|
|
HEM-804F |
|
Yarrows SA, Brook RD. Measurement Variation Among 12 Electronic Home Blood Pressure Monitors. Am J Hypertens 2000;13:276-282 |
|
|
IC HEM-725CIC |
|
De Luisignam S, Thiru K. Meredith K, Majeed A, Johnson P. Measuring blood pressure at the wrist: more comfortable for patients and more convenient for doctors? Public Health 2000;114:165-168 |
|
|
R3 |
|
Eckert S, Gleichmann U, Zagorski O, Klapp A. Validation of the Omron R3� blood pressure self-measuring device through simultaneous comparative invasive measurements according to protocol 58130 of the German Institute for Validation. Blood Pressure Monitoring 1997;2:189-192 |
|
|
Yarrows SA, Brook RD. Measurement Variation Among 12 Electronic Home Blood Pressure Monitors. Am J Hypertens 2000;13:276-282 |
|
|
Eckbert S, Kemper B, Horstkotte D. How accurate are devices for blood pressure self-measurement at the upper arm, wrist and finger in young normotensive patients and elder hypertensive patients? J Hypertens 2001;19(suppl 2):S24 (Abstract) |
|
|
R5-I |
|
Omboni S, Riva I, Giglio I, Caldara G, Groppelli A, Parati G. Validation of the Omron M5-I, R5-I and HEM-907 automated blood pressure monitors in elderly individuals according to the International Protocol of the European Society of Hypertension Blood Pressure Monitoring 2007;12:233-242 |
|
|
|
R7 |
|
Topouchian JA, El Assaad MA, Orobinskaia LV, El Feghali RN, Asmar RG. Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT) Blood Press Monit 2006;11:165-171 |
|
|
|
RX, RX-M, RX2 |
|
Zambanini A, Cox A, McIntosh C, Feher M. A comparison of semi-automated BP monitors in diabetic patients. J Hypertens 2000;18(suppl 4):S121 (Abstract) |
|
|
Altunkan Ş, Yildiz SA. Wrist blood pressure measuring devices: a comparitive study of the accuracy with a standard auscultatory method using a mercury sphygmomanometer. J Hypertens [dabl® Educational Trust regrets that full details are missing.](suppl ?):S25 (Abstract) |
|
|
Longo D, Bertolo O, Toffanin G, Frezza P, Palatini P. Validation of the A & D UA-631 (UA-779 Life Source) device for self-measurement of blood pressure and relationship between its performance and large artery compliance. Blood Press Monit 2002;7:243-248 |
|
|
Artigao LM, Llavador JJ, Puras A, Lopex-Abril J, Rubio MM, Torres C, Vidal A, Sanchi C, Divison JA, Naharro F, Caldevilla D, Fuentes G. Evaluation and validation of Omron Hem 705 CP and Hem 706/711 monitors for self-measurement of blood pressure. 2000.[dabl® Educational Trust regrets that full details are missing] |
|
|
|
Shennan AH, Rushbrook J, Power J, Wright J, Poston L. An accurate oscillometric wrist blood pressure monitor: validation of the Omron Rx [HEM-608] J Human Hypertens 1998;12:794 (Abstract) |
|
|
|
Braam R, Aslan B, Thein T. Accuracy of the Omron RX-M, an automated wrist blood pressure measuring device J Hypertens 2003;[dabl® Educational Trust regrets that full details are missing] (Abstract) |
|
|
|
Braam R, Aslan B, Thein T. Accuracy of the Omron RX-M, an automated blood pressure measuring device, measureing blood pressure at the wrist, according to a modified British Hypertension Society protocol Blood Press Monit 2004;9:25-30 |
|
|
|
Cuckson AC, Moran P, Seed P, Reinders A, Shennan AH. Clinical evaluation of an automated blood pressure wrist device Blood Press Monit 2004;9:31-37 |
|
|
|
RX3 |
|
El-Assaad MA, Orobinskaya LV, Topouchian JA, Asmar RG. Validation of the Omron® RX-3 Blood Pressure Measuring Device at the wrist level according to the International Protocol of the European Society of Hypertension AJH 2005;18(No. 5, Part 2):29A (P58) |
|
Samsung SBM-100A |
Altunkan S, Iliman N, Altunkan E. Validation of the Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure measuring devices in adults according to the International Protocol Blood Pressure Monitoring 2007;12:119-125 |
|
Sensacare SAW-102 |
Zaetta V, Longo D, Perkovic D, Prattico F, Barisa M, Perfetti P, Gabrieli A, Buonocore F, Winnicki M. Performance of the SAA-I02 and of the SAW-102, Arm and Wrist Blood Pressure Monitors, According to the Protocols of the European Society of Hypertension, Association for the Advancement of Medical Instrumentation and British Society of Hypertension Journal of Hypertension 2006;25 (suppl 2) |
|
Sunbeam7654
|
Yarrows SA, Brook RD. Measurement Variation Among
12 Electronic Home Blood Pressure Monitors. Am J Hypertens 2000;13:276-282
|
|
Finger Devices for Self-measurement
|
Lumoscope 1085M
|
Yarrows SA, Brook RD. Measurement Variation Among
12 Electronic Home Blood Pressure Monitors. Am J Hypertens
2000;13:276-282
|
|
Marshall F-88
|
Yarrows SA, Brook RD. Measurement Variation Among
12 Electronic Home Blood Pressure Monitors. Am J Hypertens
2000;13:276-282
|
|
National Ltd. EW276H
|
Yarrows SA, Brook RD. Measurement Variation Among
12 Electronic Home Blood Pressure Monitors. Am J Hypertens
2000;13:276-282
|
|
Omron |
|
B15 |
|
Yarrows SA, Brook RD. Measurement Variation Among 12 Electronic Home Blood Pressure Monitors. Am J Hypertens 2000;13:276-282 |
|
|
F3 |
|
Eckbert S, Kemper B, Horstkotte D. How accurate are devices for blood pressure self-measurement at the upper arm, wrist and finger in young normotensive patients and elder hypertensive patients? J Hypertens 2001;19(suppl 2):S24 (Abstract) |
|
|
HEM-812F |
|
Yarrows SA, Brook RD. Measurement Variation Among 12 Electronic Home Blood Pressure Monitors. Am J Hypertens 2000;13:276-282 |
|
Devices for Measuring Blood Pressure in the Community
|
A&D TM-2655 |
Kobalava ZD, Kotovskaya YV, Babaeva LA, Moiseev VS. Validation of oscillometric device for blood pressure measurement Blood Pressure Monitoring 2006;11:87-90 |
|
Keito |
Staal EM, Nygard OK, Omvik P, Gerdts E. Blood pressure measurements by the Keito machine. Evaluation versus office blood pressure by physicians Blood Press Monit 2004;9:167-172 |
|
Pharma-Smart PS-2000 |
Alpert BS. Validation of the Pharma-Smart PS-2000 public use blood pressure monitor Blood Press Monit 2004;9:19-23 |
|
Vita-Stat
|
Nara AR. Performance Review of a Noninvasive Blood
Pressure Monitor. Medical Electronics 1996;27:63-67
|
|